Clinical evaluation of a new biofeedback standing balance training device

被引:17
|
作者
Lee, MY [1 ]
Wong, MK [1 ]
Tang, FT [1 ]
机构
[1] CHANG GUNG MEM HOSP,DEPT REHABIL MED,TAIPEI 10591,TAIWAN
关键词
D O I
10.3109/03091909609008381
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For patients with neurological damage of the central nervous system, such as that due to cerebrovascular accident (CVA), standing balance training is a critical therapeutic procedure to be undertaken before walking and self-care training. The identification and characterization of neurological disorder in postural steadiness will enhance our understanding of the postural control system, and help identify patients at risk falls in the CVA population. This paper discusses the design and clinical evaluation of a new biofeedback training device for static (postural steadiness) performance of the standing balance system. The device includes a height adjustable standing table, an instrumented force sensing platform, an on-line weight bearing audio/visual biofeedback system, a postural correction mirror, and a belt suspension system for the upper extremities. A quantitative evaluation protocol of bilateral asymmetries in weight distribution and postural sway to characterize standing balance with the force sensing platform is discussed. Finally, the clinical evaluation results of sixty patients with hemiplegia from acute stroke for a period of four weeks are discussed. With this economic standing training device, the static standing steadiness can be trained effectively through weight bearing biofeedback and a postural correction mirror int he clinical and home caring environments.
引用
收藏
页码:60 / 66
页数:7
相关论文
共 50 条
  • [41] Biofeedback Based Home Balance Training can Improve Balance but Not Gait in People with Multiple Sclerosis
    Novotna, Klara
    Janatova, Marketa
    Hana, Karel
    Svestkova, Olga
    Lizrova, Jana Preiningerova
    Havrdova, Eva Kubala
    MULTIPLE SCLEROSIS INTERNATIONAL, 2019, 2019
  • [42] Standing balance evaluation using a triaxial accelerometer
    Mayagoitia, RE
    Lötters, JC
    Veltink, PH
    Hermens, H
    GAIT & POSTURE, 2002, 16 (01) : 55 - 59
  • [43] The Effect Pelvic Floor Training with an Non-Invasive Biofeedback Training Device and APP
    Soeder, S.
    Paasch, C.
    Hunger, R.
    Heisler, S.
    Goetze, M.
    Olthoff, J.
    Huenerbein, M.
    Mantke, R.
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2023, 34 : S134 - S135
  • [44] FEEDBACK STIMULATION TRAINING A NEW CONCEPT IN BIOFEEDBACK
    BOWMAN, BR
    BAKER, LL
    WATERS, R
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1978, 59 (11): : 537 - 537
  • [45] CLINICAL AND ELECTROMYOGRAPHIC EFFECTS OF BIOFEEDBACK TRAINING IN MANDIBULAR DYSFUNCTION
    DAHLSTROM, L
    CARLSSON, SG
    GALE, EN
    JANSSON, TG
    BIOFEEDBACK AND SELF-REGULATION, 1984, 9 (01): : 37 - 47
  • [46] Evaluation of MDI Inhaler Technique Using a New Training Device
    Welch, Michael J.
    Ostrom, Nancy K.
    Greiner, Alexander N.
    Laubach, Susan Stefanac
    Sanders, Mark
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 2015, 135 (02) : AB48 - AB48
  • [47] A Wearable Reaction Wheel based Kinesthetic Biofeedback Device for Delivery of Balance Cues
    Eizad, Amre
    Afzal, Muhammad Raheel
    Lee, Hosu
    Yoon, Jungwon
    Lyu, Sung-Ki
    2019 19TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS 2019), 2019, : 173 - 178
  • [48] Functional balance training using a domed device
    Ruiz, R
    Richardson, MT
    STRENGTH AND CONDITIONING JOURNAL, 2005, 27 (01) : 50 - 55
  • [49] Towards the enhancement of body standing balance recovery by means of a wireless audio-biofeedback system
    Costantini, Giovanni
    Casali, Daniele
    Paolizzo, Fabio
    Alessandrini, Marco
    Micarelli, Alessandro
    Viziano, Andrea
    Saggio, Giovanni
    MEDICAL ENGINEERING & PHYSICS, 2018, 54 : 74 - 81
  • [50] Force platform feedback for standing balance training after stroke
    Barclay-Goddard, R
    Stevenson, T
    Poluha, W
    Moffatt, MEK
    Taback, SP