Proton Energy Measurement Using Si(Li) and Scintillation Crystal Detector in the MC-50 Cyclotron

被引:5
|
作者
Kim, Kye Ryung [1 ]
Cho, Yong Sub [1 ]
Kim, Hong Ju [2 ]
So, Jung Ho [2 ]
Lee, Min Yong [3 ]
机构
[1] Korea Atom Energy Res Inst, Proton Engn Frontier Project, Taejon, South Korea
[2] Kyungpook Natl Univ, Taegu 702701, South Korea
[3] Korea Inst Radiol & Med Sci, Seoul, South Korea
关键词
proton beam energy measurement; Si(Li) detector; Monte Carlo simulation; BGO scintillation detector; absolute energy measurement;
D O I
10.1080/00223131.2008.10875910
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Absolute proton beam energy measurement with Si(Li) detector has been performed in the MC-50 cyclotron. It is known that the measurement of the absolute energy of the medium energy proton with high accuracy is very important for some experiments concerned with nuclear physics, space device test, etc. For the absolute particle energy measurement, the semiconductor detector has been recognized as a very useful tool because of their total particle energy absorbing characteristics. In the case of a scintillation detector, a part of particle energy can be lost through quenching process during penetrating the detector material and it could be a reason to increase the energy measurement error. The thicknesses of Si(Li) detector and BGO scintillation detector were 5mm and 30mm, and the sizes of active area are 75mm(2) and 49mm(2). Even though, the Si(Li) detector was originally designed and fabricated proper to the operation at the atmosphere, Si(Li) detector was installed in the vacuum chamber with the 50um thickness Al beam entrance window for more accurate measurement. The energy calibration of Si(Li) detector was done with Am-241 alpha sources. The proton beam energies from the cyclotron are 35MeV and 45MeV, and the beam current is below 0.5nA. The proton beam energy extracted through the 2mm thickness Al exit window was changed and controlled by Al degrader and air thickness between exit window and detector. The change of proton beam energy property such as mean energy and energy spread has been measured using Si(Li). The results are compared to the calculated results by Monte-Carlo code simulation and measured results by the BGO scintillation detector.
引用
收藏
页码:538 / 541
页数:4
相关论文
共 36 条
  • [21] Si(Li) detector efficiency and peak shape calibration in the low-energy range using synchrotron radiation
    Lepy, MC
    Plagnard, J
    Stemmler, P
    Ban, G
    Beck, L
    Dhez, P
    X-RAY SPECTROMETRY, 1997, 26 (04) : 195 - 202
  • [22] Energy Calibration and Inverse-Square Law of Radiation in Gamma-Ray Measurement Using a Scintillation Detector for Laboratory Experiments
    Iida, Shiori
    Jin, Wei-Guo
    JOURNAL OF ENGINEERING, 2023, 2023
  • [23] Study of Recoil-Proton-Detector System Using Organic and Inorganic Scintillators for High Energy Neutron Measurement
    Kajimoto, Tsuyoshi
    Arakawa, Hiroyuki
    Noda, Shusaku
    Watanabe, Takehito
    Shigyo, Nobuhiro
    Ishibashi, Kenji
    Haight, Robert C.
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2008, : 526 - 529
  • [24] MEASUREMENT OF THE SPECTRAL EFFICIENCY OF ENERGY DISPERSIVE SI(LI) DETECTORS BELOW 5 KEV PHOTON ENERGY USING BESSY AS A STANDARD SOURCE
    MULLER, P
    RIEHLE, F
    TEGELER, E
    WENDE, B
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3): : 569 - 571
  • [25] Optimization of the signal-to-background ratio in energy-dispersive X-ray fluorescence analysis using a Si(Li) detector
    Portnoi, A.Yu.
    Pavlinskii, G.V.
    Gorbunov, M.S.
    Baranov, E.O.
    Zuzaan, P.
    Journal of Analytical Chemistry, 2009, 64 (05): : 495 - 504
  • [26] Optimization of the signal-to-background ratio in energy-dispersive X-ray fluorescence analysis using a Si(Li) detector
    A. Yu. Portnoi
    G. V. Pavlinskii
    M. S. Gorbunov
    E. O. Baranov
    P. Zuzaan
    Journal of Analytical Chemistry, 2009, 64 : 495 - 504
  • [27] Optimization of the signal-to-background ratio in energy-dispersive X-ray fluorescence analysis using a Si(Li) detector
    Portnoi, A. Yu.
    Pavlinskii, G. V.
    Gorbunov, M. S.
    Baranov, E. O.
    Zuzaan, P.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2009, 64 (05) : 495 - 504
  • [28] APPARATUS FOR LOW-ENERGY X-RAY-FLUORESCENCE USING AN AL K X-RAY SOURCE AND AN SI(LI) DETECTOR
    MUSKET, RG
    HOLMES, SJ
    X-RAY SPECTROMETRY, 1990, 19 (04) : 177 - 184
  • [29] PRECISE MEASUREMENT OF THE PRIMARY PROTON BEAM ENERGY OF AN ISOCHRONOUS 18 MEV CYCLOTRON USING A STACKED FOILS TECHNIQUE: BEFORE AND FOLLOWING A MAJOR ENGINEERING UPGRADE
    Asad, Ali
    Asad, Ali
    Chan, Sun
    Cryer, David
    Price, Roger
    Gibbons, Peter
    INTERNAL MEDICINE JOURNAL, 2013, 43 : 33 - 34
  • [30] Measurement of the neutron capture cross-sections for Dy and Hf between 0.001 eV and 50 keV using total energy absorption detector
    Cho, HJ
    Kobayashi, K
    Yamamoto, S
    Fujita, Y
    Kim, GN
    Chang, JH
    Ko, SK
    REACTOR DOSIMETRY: RADIATION METROLOGY AND ASSESSMENT, 2001, 1398 : 462 - 469