The Clinical Rationale for the Sentry Bioconvertible Inferior Vena Cava Filter for the Prevention of Pulmonary Embolism

被引:6
|
作者
Dake, Michael D. [1 ,2 ]
Ansel, Gary M. [3 ]
Johnson, Matthew S. [4 ]
Mendes, Robert [5 ]
Smouse, H. Bob [6 ]
机构
[1] Univ Arizona Hlth Sci, Dept Med Imaging, Tucson, AZ 85721 USA
[2] Univ Arizona Hlth Sci, Dept Surg, Tucson, AZ 85721 USA
[3] OhioHlth Riverside Methodist Hosp, Dept Cardiol & Vasc Med, Columbus, OH USA
[4] Indiana Univ Sch Med, Dept Vasc & Intervent Radiol, Indianapolis, IN 46202 USA
[5] UNC Rex Hosp, NC Heart & Vasc Res, Dept Vasc Surg, Raleigh, NC USA
[6] OSF St Francis Med Ctr, Dept Intervent Radiol, Peoria, IL USA
关键词
VENOUS THROMBOEMBOLISM;
D O I
10.1155/2019/5795148
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The Sentry inferior vena cava (IVC) filter is designed to provide temporary protection against pulmonary embolism (PE) during transient high-risk periods and then to bioconvert after 60 days after implantation. At the time of bioconversion, the device's nitinol arms retract from the filtering position into the caval wall. Subsequently, the stable stent-like nitinol frame is endothelialized. The Sentry bioconvertible IVC filter has been evaluated in a multicenter investigational-device-exemption pivotal trial (NCT01975090) of 129 patients with documented deep vein thrombosis (DVT) or PE, or at temporary risk of developing DVT or PE, and with contraindications to anticoagulation. Successful filter conversion was observed in 95.7% of patients at 6 months (110/115) and 96.4% at 12 months (106/110). Through 12 months, there were no cases of symptomatic PE. The rationale for development of the Sentry bioconvertible device includes the following considerations: (1) the period of highest risk of PE for the vast majority of patients occurs within the first 60 days after an index event, with most of the PEs occurring in the first 30 days; (2) the design of retrievable IVC filters to support their removal after a transitory high-PE-risk period has, in practice, been associated with insecure filter dynamics and time-dependent complications including tilting, fracture, embolization, migration, and IVC perforation; (3) most retrievable IVC filters are placed for temporary protection, but for a variety of reasons they are not removed in any more than half of implanted patients, and when removal is attempted, the procedure is not always successful even with advanced techniques; and (4) analysis of Medicare hospital data suggests that payment for the retrieval procedure does not routinely compensate for expense. The Sentry device is not intended for removal after bioconversion. In initial clinical use, complications have been limited. Long-term results for the Sentry bioconvertible IVC filter are anticipated soon.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Bioconvertible inferior vena cava filter
    Arroyo Lopez, Marta
    Villar Esnal, Roberto
    Sanchez Guerrero, Angel
    Gonzalez-Fajardo, Jose A.
    [J]. ANGIOLOGIA, 2021, 73 (02): : 103 - 106
  • [2] Final Two-Year Outcomes for the Sentry Bioconvertible Inferior Vena Cava Filter in Patients Requiring Temporary Protection from Pulmonary Embolism
    Dake, Michael D.
    Murphy, Timothy P.
    Kramer, Albrecht H.
    Darcy, Michael D.
    Sewall, Luke E.
    Curi, Michael A.
    Johnson, Matthew S.
    Arena, Frank
    Swischuk, James L.
    Ansel, Gary M.
    Silver, Mitchell J.
    Saddekni, Souheil
    Brower, Jayson S.
    Mendes, Robert
    Feezor, Robert
    Kalva, Sanjeeva
    Kies, Darren
    Bosiers, Marc
    Ziegler, Werner
    Farber, Mark
    Paolini, David
    Spillane, Robert
    Jones, Steven
    Peeters, Patrick
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 31 (02) : 221 - 230
  • [3] Sentry Bioconvertible Inferior Vena Cava Filter: Study of Stages of Incorporation in an Experimental Ovine Model
    Gaines, Peter A.
    Kolodgie, Frank D.
    Crowley, Gordon
    Horan, Steven
    MacDonagh, Megan
    McLucas, Emily
    Rosenthal, David
    Strong, Ashley
    Sweet, Michael
    Panchal, Deepal K.
    [J]. INTERNATIONAL JOURNAL OF VASCULAR MEDICINE, 2018, 2018
  • [4] One-Year Analysis of the Prospective Multicenter SENTRY Clinical Trial: Safety and Effectiveness of the Novate Sentry Bioconvertible Inferior Vena Cava Filter
    Dake, Michael D.
    Murphy, Timothy P.
    Kramer, Albrecht H.
    Darcy, Michael D.
    Sewall, Luke E.
    Curi, Michael A.
    Johnson, Matthew S.
    Arena, Frank
    Swischuk, James L.
    Ansel, Gary M.
    Silver, Mitchell J.
    Saddekni, Souheil
    Brower, Jayson S.
    Mendes, Robert
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 29 (10) : 1350 - 1361
  • [5] Clinical outcomes of inferior vena cava filter in complicated pulmonary embolism
    Gul, Muhammad H.
    Htun, Zin M.
    Rigdon, Joseph
    Rivera-Lebron, Belinda
    Perez, Vinicio de Jesus
    [J]. PULMONARY CIRCULATION, 2019, 9 (04)
  • [6] Prevention of pulmonary embolism with a vena cava filter
    Obergassel, L
    Miketic, S
    Carlsson, J
    Tebbe, U
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1996, 121 (34-35) : 1060 - 1065
  • [7] Inferior Vena Cava Filter and Recurrent Pulmonary Embolism
    Hoffer, Eric K.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2015, 314 (07): : 728 - 729
  • [8] Inferior Vena Cava Filter Improves Mortality in Pulmonary Embolism
    Htun, Z.
    Gul, M. H.
    Aye, H. H.
    Jain, P.
    Lwin, A. K.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2018, 197
  • [9] RECURRENT PULMONARY EMBOLISM WITH INFERIOR VENA CAVA FILTER PLACEMENT
    Jiwa, Nasheena
    Mutneja, Rahul
    [J]. CHEST, 2020, 158 (04) : 2069A - 2069A
  • [10] Inferior Vena Cava Filter and Recurrent Pulmonary Embolism Reply
    Mismetti, Patrick
    Laporte, Silvy
    Meyer, Guy
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2015, 314 (07): : 729 - 730