Wear testing of crosslinked polyethylene: Wear rate variability and microbial contamination

被引:12
|
作者
Brandt, J. -M. [1 ,2 ,4 ]
Vecherya, A. [3 ]
Guenther, L. E. [3 ]
Koval, S. F. [4 ]
Petrak, M. J. [1 ,3 ]
Bohm, E. R. [1 ]
Wyss, U. P. [1 ,2 ]
机构
[1] Concordia Joint Replacement Grp, Winnipeg, MB R2K 2M9, Canada
[2] Univ Manitoba, Dept Mech & Mfg Engn, Winnipeg, MB R3T 2N2, Canada
[3] Orthopaed Innovat Ctr Inc, Winnipeg, MB R2K 2M9, Canada
[4] Univ Western Ontario, Dept Microbiol & Immunol, London, ON N6A 5C1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Polyethylene; Wear testing; Pin-on-disc testing; Hip simulator; Knee simulator; FEMORAL KNEE COMPONENTS; ACETABULAR CUPS; SIMULATOR; HIP; ARTHROPLASTY;
D O I
10.1016/j.jmbbm.2014.02.016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The wear performance of two types of crosslinked polyethylene (Marathon (TM) and XLIC (TM), DePuy Synthes Inc., Warsaw, IN) was evaluated in a pin-on-disc wear tester, a hip wear simulator, and a knee wear simulator. Sodium azide was used as the microbial inhibitor in the calf serum-based lubricant. In the pin-on-disc wear tester, the Marathon wear rate of 5.33 +/- 0.54 mm(3)/Mc was significantly lower (p=0.002) than the wear rate of 6.43 +/- 0.60 mm(3)/Mc for XLK. Inversely, the Marathon wear rate of 15.07 +/- 1.03 mm(3)/Mc from the hip wear simulator was 2.2-times greater than the XLK wear rate of 6.71 +/- 1.03 mm(3)/Mc from the knee wear simulator. Differences in implant design, conformity, GUR type, and kinematic test conditions were suggested to account for the difference between the wear rates generated in the different types of wear testing apparati. In all wear tests, sodium azide was ineffective at inhibiting microbial growth in the lubricant. Eight different organisms were identified in the lubricant samples from the wear tests, which suggested the necessity of using an alternative, more effective microbial inhibitor. Careful sample preparation and thorough cleaning has shown to improve the consistency of the wear results. The wear rates generated in the hip and knee wear simulators closely reflected the wear behaviour of Marathon and XLK reported in published data that were tested under similar conditions. Crown Copyright (C) 2014 Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:208 / 216
页数:9
相关论文
共 50 条
  • [1] Wear of highly crosslinked polyethylene acetabular components
    Callary, Stuart A.
    Solomon, Lucian B.
    Holubowycz, Oksana T.
    Campbell, David G.
    Munn, Zachary
    Howie, Donald W.
    ACTA ORTHOPAEDICA, 2015, 86 (02) : 159 - 168
  • [2] Influence of crosslinked polyethylene structure on wear of joint replacements
    Kilgour, Alastair
    Elfick, Alistair
    TRIBOLOGY INTERNATIONAL, 2009, 42 (11-12) : 1582 - 1594
  • [3] Prediction of Wear in Crosslinked Polyethylene Unicompartmental Knee Arthroplasty
    Netter, Jonathan
    Hermida, Juan
    Flores-Hernandez, Cesar
    Steklov, Nikolai
    Kester, Mark
    D'Lima, Darryl D.
    LUBRICANTS, 2015, 3 (02): : 381 - 393
  • [4] Wear of crosslinked polyethylene under different tribological conditions
    Galvin, A
    Kang, L
    Tipper, J
    Stone, M
    Ingham, E
    Jin, ZM
    Fisher, J
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2006, 17 (03) : 235 - 243
  • [5] Wear of crosslinked polyethylene under different tribological conditions
    Alison Galvin
    Lu Kang
    Joanne Tipper
    Martin Stone
    Eileen Ingham
    Zhongmin Jin
    John Fisher
    Journal of Materials Science: Materials in Medicine, 2006, 17 : 235 - 243
  • [6] Wear rate of polyethylene bearing components
    Morrey, BF
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (06): : 899 - 899
  • [7] Wear Analysis in THA Utilizing Oxidized Zirconium and Crosslinked Polyethylene
    Garvin, Kevin L.
    Hartman, Curtis W.
    Mangla, Jimmi
    Murdoch, Nathan
    Martell, John M.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (01) : 141 - 145
  • [8] Screening of stabilized crosslinked polyethylene using a novel wear tester
    Hoffmann, A
    Gonzalez-Mora, V
    Chiesa, R
    Cigada, A
    Stroosnijder, MF
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2002, 12 (04) : 387 - 395
  • [9] Wear testing of hard coatings: More than wear rate?
    Celis, JP
    Vancoille, E
    Mohrbacher, H
    Stals, L
    SURFACE MODIFICATION TECHNOLOGIES XI, 1998, : 860 - 871
  • [10] Wear testing of hard coatings: More than wear rate?
    Celis, J.-P.
    Stals, L.
    Vancoille, E.
    Mohrbacher, H.
    Surface Engineering, 1998, 14 (03): : 205 - 210