Contralateral pneumothorax after endocardial dual-chamber pacemaker implantation resulting from atrial lead perforation
被引:7
|作者:
Van Herendael, Hugo
[1
]
Willems, Rik
[2
]
机构:
[1] Univ Toronto, St Michaels Hosp, Div Cardiol, Toronto, ON M5B 1W8, Canada
[2] Univ Leuven, Univ Hosp Leuven, Div Cardiol, Louvain, Belgium
关键词:
Contralateral pneumothorax;
pacemaker;
atrial lead perforation;
ACTIVE-FIXATION;
PACING LEAD;
PERICARDITIS;
BIPOLAR;
D O I:
10.2143/AC.64.2.2036150
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
We describe the occurrence of a right-sided pneumothorax following a left-sided dual chamber pacemaker implantation, due to a perforation of the screw of the J-shaped active-fixation lead through the right atrial wall. A review of the literature regarding complications of different atrial lead types (passive vs. active and J-shape vs. straight) is provided. Current data suggest that passive-fixation leads could have an advantage, provided they can be positioned satisfactorily because of the lower risk of pericardial complications. If active-fixation leads are used, straight leads might be preferable above J-shaped leads because of the lower rate of perforation.
引用
收藏
页码:271 / 273
页数:3
相关论文