Normalized glandular dose (DgN) coefficients from experimental mammographic x-ray spectra

被引:5
|
作者
Santos, Josilene C. [1 ]
Tomal, Alessandra [2 ]
de Barros, Nestor [3 ]
Costa, Paulo R. [1 ]
机构
[1] Univ Sao Paulo, Inst Fis, BR-05508090 Sao Paulo, SP, Brazil
[2] IFGW UNICAMP, Inst Fis Gleb Wataghin, BR-13083859 Campinas, SP, Brazil
[3] Univ Sao Paulo, Fac Med, BR-05403900 Sao Paulo, SP, Brazil
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 10期
基金
巴西圣保罗研究基金会;
关键词
mammography; normalized glandular dose coefficients; x-ray spectrometry; dosimetry; MONTE-CARLO; BREAST DOSIMETRY;
D O I
10.1088/1361-6560/ab171a
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mean glandular dose is the quantity used for dosimetry in mammography and depends on breast-related characteristics, such as thickness and density, and on the x-ray spectrum used for breast imaging. This work aims to present an experimentally-based method to derive polyenergetic normalized glandular dose coefficients (DgN(p)) from the spectral difference between x-ray spectra incident and transmitted through breast phantoms with glandular/adipose proportions of 30/70 and 50/50 and thicknesses up to 4.5 cm. The spectra were produced by a Mammomat 3000 Nova system using radiographic techniques commonly applied for imaging compressed breast thickness lower than 6 cm (Mo/Mo, Mo/Rh and W/Rh spectra at 26 and 28 kVp). DgN(p) coefficients were compared with values estimated using Boones' method and data from breast images (DICOM Organ Dose and VolparaDose calculations). The DgN(p) were also evaluated in layers into the phantoms (depth-DgN(p)) using both x-ray spectra and thermoluminescent dosimeters (TLD-100). Maximum differences between DgN(p) from the method presented in this study and results using Boone's method was 11%, with larger differences for Mo/Rh spectra in relation to the Mo/Mo. The DgN(p) maximum differences to the coefficients obtained using patient images were 8.0%, for the DgN calculated using Volpara and 6.4% for the DgN from DICOM Organ Dose, for a 4.5 cm breast phantom with 30% glandularity. The DgN(p) estimated from the depth-DgN(p) distributions differ up to 5.2% to the coefficients obtained using the pair incident-transmitted spectra to calculate the DgN(p) directly in the whole phantom. The depth-DgN(p) distributions estimated with TLDs were consistent with the results observed using the experimental spectra, with maximum difference of 3.9%. In conclusion, polyenergetic x-ray spectrometry proved to be an applicable tool for research in dosimetry in mammography allowing spectral characterization. This approach can also be useful for investigation of the influence of x-ray spectra on glandular dose.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Mean glandular dose coefficients (DgN) for x-ray spectra used in contemporary breast imaging systems
    Nosratieh, Anita
    Hernandez, Andrew
    Shen, Sam Z.
    Yaffe, Martin J.
    Seibert, J. Anthony
    Boone, John M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (18): : 7179 - 7190
  • [2] Normalized glandular dose (DgN) coefficients for arbitrary x-ray spectra in mammography: Computer-fit values of Monte Carlo derived data
    Boone, JM
    MEDICAL PHYSICS, 2002, 29 (05) : 869 - 875
  • [3] A Monte Carlo study of monoenergetic and polyenergetic normalized glandular dose (DgN) coefficients in mammography
    Sarno, Antonio
    Mettivier, Giovanni
    Di Lillo, Francesca
    Russo, Paolo
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01): : 306 - 325
  • [4] Normalized glandular dose (DgN) coefficients for flat-panel CT breast imaging
    Thacker, SC
    Glick, SJ
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (24): : 5433 - 5444
  • [5] Conversion coefficients from kerma to ambient dose and personal dose for X-ray spectra
    Otto, T.
    JOURNAL OF INSTRUMENTATION, 2019, 14 (11)
  • [6] New digital Mammography systems may require different X-ray spectra and therefore, more general normalized glandular dose values
    Kimme-Smith, C
    RADIOLOGY, 1999, 213 (01) : 7 - 10
  • [7] Three-layer heterogeneous mammographic phantoms for Monte Carlo simulation of normalized glandular dose coefficients in mammography
    Tien-Yu Chang
    Kuan-Jen Lai
    Chun-Yuan Tu
    Jay Wu
    Scientific Reports, 10
  • [8] Three-layer heterogeneous mammographic phantoms for Monte Carlo simulation of normalized glandular dose coefficients in mammography
    Chang, Tien-Yu
    Lai, Kuan-Jen
    Tu, Chun-Yuan
    Wu, Jay
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [9] Monte Carlo Evaluation of Normalized Glandular Dose Coefficients in Mammography
    Sarno, Antonio
    Mettivier, Giovanni
    Di Lillo, Francesca
    Russo, Paolo
    BREAST IMAGING, IWDM 2016, 2016, 9699 : 190 - 196
  • [10] Effect of X-ray field gradient on mean glandular dose
    Aus, RJ
    Ng, KH
    DeWerd, LA
    RADIOLOGY, 1996, 201 : 119 - 119