摘要
This paper examines the dissolution apparatus referred to as the flow-through cell from an engineering fluid mechanics viewpoint. The analysis demonstrates that laminar now predominantly occurs in the standard operation of this apparatus. It is argued that fully turbulent conditions are unlikely. Consequently. the phrases 'open column' and 'packed column' are suggested as technically more accurate terms for its operational characteristics than the conventionally referenced 'turbulent mode' and 'laminar mode'. Examples of flow profiles are given to show that the criterion of a "sinusoidal" input flow profile required by USP is not a sufficiently accurate characterization at the inlet as numerous profiles can be conceived which have the same average flow rate. The rationale of pulsating flows Versus constant, steady flows is discussed. Examples of how references to turbulence in the dissolution-related literature can lead to ambiguities and/or inconsistencies are highlighted. Published by Elsevier B.V.
引用
收藏
页码:531 / 544
页数:14
相关论文
共 50 条
  • [41] RED-BLOOD-CELL FLOW-THROUGH A MICROFABRICATED STRUCTURE
    BRODY, J
    AUSTIN, B
    BITENSKY, M
    BIOPHYSICAL JOURNAL, 1994, 66 (02) : A85 - A85
  • [42] ELECTRODEPOSITION AND STRIPPING AT GRAPHITE CLOTH ELECTRODES IN A FLOW-THROUGH CELL
    YANIV, D
    ARIEL, M
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1981, 129 (1-2): : 301 - 313
  • [43] APPLICATION OF AN IR FLOW-THROUGH CELL TO EQUILIBRIUM AND KINETICS EXPERIMENTS
    BRABSON, GD
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1993, 206 : 108 - CHED
  • [44] A miniaturized flow-through cell to evaluate skin permeation of endoxifen
    Mah, Choon Siong
    Kochhar, Jaspreet Singh
    Ong, Pei Shi
    Kang, Lifeng
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2013, 441 (1-2) : 433 - 440
  • [45] VARIABLE PATH-LENGTH FLOW-THROUGH CELL FOR SPECTROPHOTOMETRY
    CHOAT, T
    CRUYWAGEN, JJ
    HEYNS, JBB
    ANALYTICAL CHEMISTRY, 1986, 58 (12) : 2570 - 2571
  • [46] A sediment flow-through system to study the impact of shifting fluid and methane flow regimes on the efficiency of the benthic methane filter
    Steeb, Philip
    Linke, Peter
    Treude, Tina
    LIMNOLOGY AND OCEANOGRAPHY-METHODS, 2014, 12 : 25 - 45
  • [47] Flow-through bioreactors for cell-free protein synthesis
    Proceedings of the Frontiers in Bioprocessing, 1992,
  • [48] Continuous, flow-through immunomagnetic cell sorting in a quadrupole field
    Sun, LP
    Zborowski, M
    Moore, LR
    Chalmers, JJ
    CYTOMETRY, 1998, 33 (04): : 469 - 475
  • [49] DYNAMIC GROUP SCHEDULING HEURISTICS IN A FLOW-THROUGH CELL ENVIRONMENT
    MAHMOODI, F
    TIERNEY, EJ
    MOSIER, CT
    DECISION SCIENCES, 1992, 23 (01) : 61 - 85
  • [50] Carbon Nanotube Based Flow-Through Electrochemical Cell for Electroanalysis
    Buffa, Andrea
    Erel, Yigal
    Mandler, Daniel
    ANALYTICAL CHEMISTRY, 2016, 88 (22) : 11007 - 11015