In-Tube Solid-Phase Microextraction and Liquid Chromatography Using a Monolithic Column for the Selective Determination of Residual Ethylenediamine in Industrial Cationic Polymers

被引:15
|
作者
Prieto-Blanco, M. C. [2 ]
Lopez-Mahia, P. [2 ]
Campins-Falco, P. [1 ]
机构
[1] Univ Valencia, Fac Quim, Dept Quim Analit, E-46100 Valencia, Spain
[2] Univ A Coruna, Fac Ciencias, Dept Quim Analit, E-15071 La Coruna, Spain
关键词
PRECOLUMN DERIVATIZATION; WATER SAMPLES; AMINES; POLYAMINES; STABILITY; HPLC;
D O I
10.1021/ac900796j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The selective determination of the diamine ethylenediamine (EDA) in the presence of a higher amount of residual dimethylamine in cationic polymers has been developed. The strategy uses both a solution derivatization with a selective agent of primary amines such as o-phthaldialdehyde-N-acetyl-L-cysteine (OPA-NAC) and an in-tube solid-phase microextraction (IT-SPME) coupled to liquid chromatography (LC). A 70 cm long, 0.32 mm internal diameter, and 3 mu m thick commercially available capillary column coated with 95% polydimethylsiloxane and 5% polydiphenylsiloxane was employed to replace the injection loop of a Rheodyne injection valve. A volume of 1 mL of derivatized sample was passed through the capillary, and 100 mu L of water was later used for cleaning and filling the capillary. After, the injection was effected and the desorption of the derivative from the capillary was carried out in the dynamic mode using the mobile phase. Chromatographic separation was performed in less than 2 min using a monolithic silica column Onyx (100 mm x 3.0 mm i.d.) under isocratic conditions. The effect of several parameters affecting derivatization and IT-SPME was investigated. The quantification of EDA was realized over the range of 0.07-2 mu g/mL with adequate linearity, accuracy, and reproducibility, and the limit of detection was 20 ng/mL The method is rapid and low in cost, and sample handling and organic solvent consumption have been minimized. Its application to polymeric cationic surfactants used in water treatment allowed the selective quantification of residual EDA at low microgram per milliliter levels of concentration without off-line preconcentration.
引用
收藏
页码:5827 / 5832
页数:6
相关论文
共 50 条
  • [21] Determination of fluoroquinolones in environmental waters by in-tube solid-phase microextraction coupled with liquid chromatography-tandem mass spectrometry
    Mitani, K
    Kataoka, H
    ANALYTICA CHIMICA ACTA, 2006, 562 (01) : 16 - 22
  • [22] Automated in-tube solid-phase microextraction-liquid chromatography-electrospray ionization mass spectrometry for the determination of ranitidine
    Kataoka, H
    Lord, HL
    Pawliszyn, J
    JOURNAL OF CHROMATOGRAPHY B, 1999, 731 (02): : 353 - 359
  • [23] Determination of cortisol in human saliva by automated in-tube solid-phase microextraction coupled with liquid chromatography-mass spectrometry
    Kataoka, Hiroyuki
    Matsuura, Eriko
    Mitani, Kurie
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 44 (01) : 160 - 165
  • [24] Determination of ochratoxins in nuts and grain samples by in-tube solid-phase microextraction coupled with liquid chromatography-mass spectrometry
    Saito, Keita
    Ikeuchi, Risa
    Kataoka, Hiroyuki
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1220 : 1 - 6
  • [25] Ion-pair in-tube solid-phase microextraction and capillary liquid chromatography using a titania-based column: Application to the specific lauralkonium chloride determination in water
    Prieto-Blanco, M. C.
    Moliner-Martinez, Y.
    Lopez-Mahia, P.
    Campins-Falco, P.
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1248 : 55 - 59
  • [26] Advantages of monolithic over particulate columns for multiresidue analysis of organic pollutants by in-tube solid-phase microextraction coupled to capillary liquid chromatography
    Moliner-Martinez, Y.
    Molins-Legua, C.
    Verdu-Andres, J.
    Herraez-Hernandez, R.
    Campins-Falco, P.
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (37) : 6256 - 6262
  • [27] On-fibre solid-phase microextraction coupled to conventional liquid chromatography versus in-tube solid-phase microextraction coupled to capillary liquid chromatography for the screening analysis of triazines in water samples
    Chafer-Pericas, C.
    Herraez-Hernandez, R.
    Campins-Falco, P.
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1125 (02) : 159 - 171
  • [28] Silver nanoparticles-coated monolithic column for in-tube solid-phase microextraction of monounsaturated fatty acid methyl esters
    Jiang, Nan
    Wang, Jiabin
    Li, Wenbang
    Xiao, Jianhua
    Li, Jianhua
    Lin, Xucong
    Xie, Zenghong
    You, Lijun
    Zhang, Qiqing
    JOURNAL OF CHROMATOGRAPHY A, 2019, 1585 : 19 - 26
  • [29] Innovations in Extractive Phases for In-Tube Solid-Phase Microextraction Coupled to Miniaturized Liquid Chromatography: A Critical Review
    Ponce-Rodriguez, Henry Daniel
    Verdu-Andres, Jorge
    Herraez-Hernandez, Rosa
    Campins-Falco, Pilar
    MOLECULES, 2020, 25 (10):
  • [30] Sodium hyaluronate-functionalized urea-formaldehyde monolithic column for hydrophilic in-tube solid-phase microextraction of melamine
    Wang, Jiabin
    Jiang, Nan
    Cai, Zhengmiao
    Li, Wenbang
    Li, Jianhua
    Lin, Xucong
    Xie, Zenghong
    You, Lijun
    Zhang, Qiqing
    JOURNAL OF CHROMATOGRAPHY A, 2017, 1515 : 54 - 61