Validity and intra-rater reliability of an Android phone application to measure cervical range-of-motion

被引:54
|
作者
Quek, June [1 ,2 ]
Brauer, Sandra G. [1 ]
Treleaven, Julia [1 ]
Pua, Yong-Hao [2 ]
Mentiplay, Benjamin [3 ]
Clark, Ross Allan [3 ]
机构
[1] Univ Queensland, St Lucia, Qld 4072, Australia
[2] Singapore Gen Hosp, Dept Physiotherapy, Singapore 169608, Singapore
[3] Australian Catholic Univ, Fac Hlth Sci, Sch Exercise Sci, Fitzroy, Vic 3065, Australia
关键词
Spine; Whiplash; Inertial monitoring unit; Phone; Cervical range of motion; AGREEMENT;
D O I
10.1186/1743-0003-11-65
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Concurrent validity and intra-rater reliability using a customized Android phone application to measure cervical-spine range-of-motion (ROM) has not been previously validated against a gold-standard three-dimensional motion analysis (3DMA) system. Findings: Twenty-one healthy individuals (age: 31 +/- 9.1 years, male: 11) participated, with 16 re-examined for intra-rater reliability 1-7 days later. An Android phone was fixed on a helmet, which was then securely fastened on the participant's head. Cervical-spine ROM in flexion, extension, lateral flexion and rotation were performed in sitting with concurrent measurements obtained from both a 3DMA system and the phone. The phone demonstrated moderate to excellent (ICC = 0.53-0.98, Spearman. = 0.52-0.98) concurrent validity for ROM measurements in cervical flexion, extension, lateral-flexion and rotation. However, cervical rotation demonstrated both proportional and fixed bias. Excellent intra-rater reliability was demonstrated for cervical flexion, extension and lateral flexion (ICC = 0.82-0.90), but poor for right- and left-rotation (ICC = 0.05-0.33) using the phone. Possible reasons for the outcome are that flexion, extension and lateral-flexion measurements are detected by gravity-dependent accelerometers while rotation measurements are detected by the magnetometer which can be adversely affected by surrounding magnetic fields. Conclusion: The results of this study demonstrate that the tested Android phone application is valid and reliable to measure ROM of the cervical-spine in flexion, extension and lateral-flexion but not in rotation likely due to magnetic interference. The clinical implication of this study is that therapists should be mindful of the plane of measurement when using the Android phone to measure ROM of the cervical-spine.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Validity and intra-rater reliability of an Android phone application to measure cervical range-of-motion
    June Quek
    Sandra G Brauer
    Julia Treleaven
    Yong-Hao Pua
    Benjamin Mentiplay
    Ross Allan Clark
    [J]. Journal of NeuroEngineering and Rehabilitation, 11
  • [2] Test-Retest and Intra-rater Reliability of Using Inertial Sensors and Its Integration with Microsoft Kinect™ to Measure Shoulder Range-of-Motion
    Beshara, Peter
    Chen, Judy
    Lagadec, Pierre
    Walsh, W. R.
    [J]. INTERNET OF THINGS TECHNOLOGIES FOR HEALTHCARE, HEALTHYIOT 2016, 2016, 187 : 177 - 184
  • [3] Inter-rater, intra-rater and inter-instrument reliability of an electrogoniometer to measure wrist range of motion
    da Silva Camassuti, Patricia Aparecida
    Marcolino, Alexandre Marcio
    Tamanini, Guilherme
    Barbosa, Rafael Inacio
    Barbosa, Amanda Matias
    Registro Fonseca, Marisa de Cassia
    [J]. HAND THERAPY, 2015, 20 (01) : 3 - 10
  • [4] Inter-rater and intra-rater reliability of Kinovea software for measurement of shoulder range of motion
    Reham M. Abd Elrahim
    Eman A. Embaby
    Mohamed F. Ali
    Ragia M. Kamel
    [J]. Bulletin of Faculty of Physical Therapy, 2016, 21 (2) : 80 - 87
  • [5] Intra-rater reliability in measuring range of motion in infants with congenital muscular torticollis
    Klackenberg, Eva Perbeck
    Elfving, Britt
    Haglund-Akerlind, Yvonne
    Carlberg, Eva Brogren
    [J]. EUROPEAN JOURNAL OF PHYSIOTHERAPY, 2005, 7 (02) : 84 - 91
  • [6] VALIDITY AND INTRA-RATER RELIABILITY OF LASER GONIOMETER VERSUS ELECTRO-GONIOMETER IN MEASURING SHOULDER RANGE OF MOTION
    Elgendy, Mohammed H.
    Abd El-khalek, Wael O. A.
    [J]. INTERNATIONAL JOURNAL OF PHYSIOTHERAPY, 2019, 6 (05) : 169 - 176
  • [7] Intra-rater and Inter-rater Reliability of Mandibular Range of Motion Measures Considering a Neutral Craniocervical Position
    Beltran-Alacreu, Hector
    Lopez-de-Uralde-Villanueva, Ibai
    Paris-Alemany, Alba
    Angulo-Diaz-Parreno, Santiago
    La Touche, Roy
    [J]. JOURNAL OF PHYSICAL THERAPY SCIENCE, 2014, 26 (06) : 915 - 920
  • [8] Intra-rater reliability of goniometry to measure scapular protraction and retraction
    Short, Nathan
    Almonroeder, Thomas G.
    Fenker, Caroline A.
    Fisher, Olivia A.
    Francetic, Kailey E.
    Hodel, Amy E.
    Lange, Clayton A.
    Mathew, Manu M.
    [J]. JOURNAL OF HAND THERAPY, 2022, 35 (02) : 275 - 281
  • [9] Inter and intra-rater reliability of mobile device goniometer in measuring lumbar flexion range of motion
    Bedekar, Nilima
    Suryawanshi, Mayuri
    Rairikar, Savita
    Sancheti, Parag
    Shyam, Ashok
    [J]. JOURNAL OF BACK AND MUSCULOSKELETAL REHABILITATION, 2014, 27 (02) : 161 - 166
  • [10] Intra-rater and inter-rater reliability of cervical active range of movement in young asymptomatic adults using inertial sensors
    Anoro-Hervera, Alejandro
    Lafuente-Perez, Adrian
    Navarro-Fernandez, Gonzalo
    Munoz-Garcia, Daniel
    Lerma-Lara, Sergio
    Beltran-Alacreu, Hector