Transit and non-transit 3D EPID dosimetry versus detector arrays for patient specific QA

被引:28
|
作者
Olaciregui-Ruiz, Igor [1 ]
Vivas-Maiques, Begona [1 ]
Kaas, Jochem [1 ]
Perik, Thijs [1 ]
Wittkamper, Frits [1 ]
Mijnheer, Ben [1 ]
Mans, Anton [1 ]
机构
[1] Netherlands Canc Inst Antoni van Leeuwenhoek, Dept Radiat Oncol, Amsterdam, Netherlands
来源
关键词
detector arrays; EPID dosimetry; IMRT; patient specific QA; VMAT; POLYMER GEL DOSIMETRY; IN-VIVO DOSIMETRY; DOSE RECONSTRUCTION; CLINICAL-EXPERIENCE; DIODE-ARRAY; IMRT; VERIFICATION; VALIDATION; ALGORITHM; SOFTWARE;
D O I
10.1002/acm2.12610
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Despite their availability and simplicity of use, Electronic Portal Imaging Devices (EPIDs) have not yet replaced detector arrays for patient specific QA in 3D. The purpose of this study is to perform a large scale dosimetric evaluation of transit and non-transit EPID dosimetry against absolute dose measurements in 3D. Methods After evaluating basic dosimetric characteristics of the EPID and two detector arrays (Octavius 1500 and Octavius 1000(SRS)), 3D dose distributions for 68 VMAT arcs, and 10 IMRT plans were reconstructed within the same phantom geometry using transit EPID dosimetry, non-transit EPID dosimetry, and the Octavius 4D system. The reconstructed 3D dose distributions were directly compared by gamma-analysis (2L2 = 2% local/2 mm and 3G2 = 3% global/2 mm, 50% isodose) and by the percentage difference in median dose to the high dose volume (% increment HDVD50). Results Regarding dose rate dependency, dose linearity, and field size dependence, the agreement between EPID dosimetry and the two detector arrays was found to be within 1.0%. In the 2L2 gamma-comparison with Octavius 4D dose distributions, the average gamma-pass rate value was 92.2 +/- 5.2%(1SD) and 94.1 +/- 4.3%(1SD) for transit and non-transit EPID dosimetry, respectively. 3G2 gamma-pass rate values were higher than 95% in 150/156 cases. % increment HDVD50 values were within 2% in 134/156 cases and within 3% in 155/156 cases. With regard to the clinical classification of alerts, 97.5% of the treatments were equally classified by EPID dosimetry and Octavius 4D. Conclusion Transit and non-transit EPID dosimetry are equivalent in dosimetric terms to conventional detector arrays for patient specific QA. Non-transit 3D EPID dosimetry can be readily used for pre-treatment patient specific QA of IMRT and VMAT, eliminating the need of phantom positioning.
引用
收藏
页码:79 / 90
页数:12
相关论文
共 50 条
  • [1] A convolutional neural network model for EPID-based non-transit dosimetry
    Bosco, Lucas Dal
    Franceries, Xavier
    Romain, Blandine
    Smekens, Francois
    Husson, Francois
    Le Lann, Marie-Veronique
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (06):
  • [2] ROC Analysis of 3D EPID Transit Dosimetry to Patient Related Variations
    Ruiz, I. Olaciregui
    Rozendaal, R.
    Mijnheer, B.
    Mans, A.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E182 - E182
  • [3] Site-specific alert criteria to detect patient-related errors with 3D EPID transit dosimetry
    Olaciregui-Ruiz, Igor
    Rozendaal, Roel
    Mijnheer, Ben
    Mans, Anton
    [J]. MEDICAL PHYSICS, 2019, 46 (01) : 45 - 55
  • [4] Comparison of deep learning models for building two-dimensional non-transit EPID Dosimetry on Varian Halcyon
    Ramadhan, Muhammad Mahdi
    Wibowo, Wahyu Edy
    Prajitno, Prawito
    Pawiro, Supriyanto Ardjo
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2023, 28 (06) : 737 - 745
  • [5] A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry A step to QA with in vivo EPID dosimetry
    Moustakis, Christos
    Ebrahimi Tazehmahalleh, Fatemeh
    Elsayad, Khaled
    Fezeu, Francis
    Scobioala, Sergiu
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2020, 196 (02) : 182 - 192
  • [6] Error detection during VMAT delivery using EPID-based 3D transit dosimetry
    Mijnheer, B.
    Jomehzadeh, A.
    Gonzalez, P.
    Olaciregui-Ruiz, I
    Rozendaal, R.
    Shokrani, P.
    Spreeuw, H.
    Tielenburg, R.
    Mans, A.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 54 : 137 - 145
  • [7] Patient-Specific EPID Based High Resolution 3D VMAT QA
    Nicol, S.
    Furstoss, C.
    Munger, P.
    Wierzbicki, W.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Implementing an EPID Based 3D Absolute Dosimetry System for IMRT QA
    Patel, J. B.
    Bismack, B. J.
    Parsai, E. I.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S746 - S746
  • [9] A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetryA step to QA with in vivo EPID dosimetry
    Christos Moustakis
    Fatemeh Ebrahimi Tazehmahalleh
    Khaled Elsayad
    Francis Fezeu
    Sergiu Scobioala
    [J]. Strahlentherapie und Onkologie, 2020, 196 : 182 - 192
  • [10] Quasi 3D dosimetry (EPID, conventional 2D/3D detector matrices)
    Back, A.
    [J]. 8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573