Evaluation of a clinical TOF-PET detector design that achieves ≤100 ps coincidence time resolution

被引:54
|
作者
Cates, Joshua W. [1 ]
Levin, Craig S. [1 ,2 ,3 ,4 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Phys, Stanford, CA 94305 USA
[4] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 11期
关键词
time-of-flight positron emission tomography; silicon photomultipliers; fast timing; TIMING RESOLUTION; SENSITIVITY; PERFORMANCE; CRYSTALS; PHOTONS; DEPTH;
D O I
10.1088/1361-6560/aac504
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Commercially available clinical positron emission tomography (PET) detectors employ scintillation crystals that are long (>= 20 mm length) and narrow (4-5 mm width) optically coupled on their narrow end to a photosensor. The aspect ratio of this traditional crystal rod configuration and 511 keV photon attenuation properties yield significant variances in scintillation light collection efficiency and transit time to the photodetector, due to variations in the 511 keV photon interaction depth in the crystal. These variances contribute significant to coincidence time resolution degradation. If instead, crystals are coupled to a photosensor on their long side, near-complete light collection efficiency can be achieved, and scintillation photon transit time jitter is reduced. In this work, we compare the achievable coincidence time resolution (CTR) of LGSO:Ce(0.025 mol%) crystals 3-20 mm in length when optically coupled to silicon photomultipliers (SiPMs) on either their short end or long side face. In this 'side readout' configuration, a CTR of 102 +/- 2 ps FWHM was measured with 2.9 x .2.9 x 20 mm(3) crystals coupled to rows of 3 x 3 mm(2) SensL-J SiPMs using leading edge time pickoff and a single timing channel. This is in contrast to a CTR of 137 +/- 3 ps FWHM when the same crystals were coupled to single 3x3 mm(2) SiPMs on their narrow ends. We further study the statistical limit on CTR using side readout via the Cramer-Rao lower bound (CRLB), with consideration given to ongoing work to further improve photosensor technologies and exploit fast phenomena to ultimately achieve 10 ps FWHM CTR. Potential design aspects of scalable front-end signal processing readout electronics using this side readout configuration are discussed. Altogether, we demonstrate that the side readout configuration offers an immediate solution for 100 ps CTR clinical PET detectors and mitigates factors prohibiting future efforts to achieve 10 ps FWHM CTR.
引用
收藏
页数:15
相关论文
共 50 条
  • [41] Recovery of Inter-crystal Compton Scattering Events for Sensitivity Improvement of Sub-250 ps TOF-PET Detector
    Fu, Geng
    Ivan, Adrian
    Qian, Iiua
    [J]. 2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [42] A method to enhance coincidence time resolution with applications for medical imaging systems (TOF/PET)
    Ermis, E. E.
    Celiktas, C.
    Pilicer, E.
    [J]. RADIATION MEASUREMENTS, 2014, 62 : 52 - 59
  • [43] The new opportunities for high time resolution clinical TOF PET
    Conti, Maurizio
    Bendriem, Bernard
    [J]. CLINICAL AND TRANSLATIONAL IMAGING, 2019, 7 (02) : 139 - 147
  • [44] The new opportunities for high time resolution clinical TOF PET
    Maurizio Conti
    Bernard Bendriem
    [J]. Clinical and Translational Imaging, 2019, 7 : 139 - 147
  • [45] Design rules for time of flight Positron Emission Tomography (ToF-PET) heterostructure radiation detectors
    Krause, Philip
    Rogers, Edith
    Birowosuto, Muhammad Danang
    Pei, Qibing
    Auffray, Etiennette
    Vasil'ev, Andrey N.
    Bizarri, Gregory
    [J]. HELIYON, 2022, 8 (06)
  • [46] Comparative timing measurements of LYSO and LFS-3 to achieve the best time resolution for TOF-PET
    Doroud, K.
    Williams, M. C. S.
    Zichichi, A.
    Zuyeuski, R.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 793 : 57 - 61
  • [47] PROSPECTS FOR ACHIEVING < 100 PS FWHM COINCIDENCE RESOLVING TIME IN TIME-OF-FLIGHT PET
    Schaart, D. R.
    Seifert, S.
    van Dam, H. T.
    Borghi, G.
    ter Weele, D. N.
    Dorenbos, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S86 - S87
  • [48] Time resolution below 100 ps for the SciTil detector of PANDA employing SiPM
    Brunner, S. E.
    Gruber, L.
    Marton, J.
    Orth, H.
    Suzuki, K.
    [J]. JOURNAL OF INSTRUMENTATION, 2014, 9
  • [49] Study of "A method to enhance coincidence time resolution with applications for medical imaging systems (TOF/PET)"
    Bikit, I.
    Mrdja, D.
    Bikit, K.
    Slivka, J.
    Forkapic, S.
    [J]. RADIATION MEASUREMENTS, 2015, 72 : 95 - 99
  • [50] Evaluation and clinical quantification of neoplastic lesions and physiological structures in TOF-PET/MRI and non-TOF/MRI - a pilot study
    Barbosa, Felipe
    Delso, Gaspar
    Zeimpekis, Konstantinos
    ter Voert, Edwin
    Huellner, Martin
    Stolzmann, Paul
    Veit-Haibach, Patrick
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)