A 256 x 256 2-D Array Transducer with Row-Column Addressing for 3-D Rectilinear Imaging

被引:90
|
作者
Seo, Chi Hyung [1 ]
Yen, Jesse T. [2 ]
机构
[1] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[2] Univ So Calif, Dept Biomed Engn, Los Angeles, CA 90089 USA
关键词
3-DIMENSIONAL ULTRASOUND; BREAST;
D O I
10.1109/TUFFC.2009.1107
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
We present simulation and experimental results from a 5-MHz, 256 x 256 2-D (65536 elements, 38.4 x 38.4 mm) 2-D array transducer with row-column addressing. The main benefits of this design are a reduced number of interconnects, a modified transmit/receive switching scheme with a simple diode circuit, and an ability to perform volumetric imaging of targets near the transducer with transmit beamforming in azimuth and receive beamforming in elevation. The final dimensions of the transducer were 38.4 mm x 38.4 rum x 300 mu m. After a row-column transducer was prototyped, the series resonance impedance was 104 Omega at 5.4 MHz. The measured -6 dB fractional bandwidth was 53% with a center frequency of 5.3 MHz. The SNR at the transmit focus was measured to be 30 dB. At 5 MHz, the average nearest neighbor crosstalk was -25 dB. In this paper, we present 3-D images of both 5 pairs of nylon wires embedded in a clear gelatin phantom and an 8 mm diameter cylindrical anechoic cyst phantom acquired from a 256 x 256 2-D array transducer made from a 1-3 composite. We display the azimuth and elevation B-scans as well as the C-scan for each image. The cross-section of the wires is visible in the azimuth B-scan, and the long axes can be seen in the elevation B-scan and C-scans. The pair of wires with 1-mm axial separation is discernible in the elevational B-scan. When a single wire from the wire target phantom was used, the measured lateral beamwidth was 0.68 mm and 0.70 mm at 30 mm depth in transmit beamforming and receive beamforming, respectively, compared with the simulated beamwidth of 0.55 mm. The cross-section of the cyst is visible in the azimuth B-scan whereas the long axes can be seen as a rectangle in the elevation B-scan and C-scans.
引用
收藏
页码:837 / 847
页数:11
相关论文
共 50 条
  • [21] Development of a 2-D Array Ultrasonic Transducer for 3-D Imaging of Objects Immersed in Water
    Rodrigues, Estevio Patricio
    De Oliveira, Timoteo Francisco
    Matuda, Marcelo Yassunori
    Buiochi, Flavio
    SENSORS, 2021, 21 (10)
  • [22] 3-D Imaging using Row-Column-Addressed 2-D Arrays with a Diverging Lens
    Bouzari, Hamed
    Engholm, Mathias
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [23] 7.5 MHz Dual-Layer Transducer Array for 3-D Rectilinear Imaging
    Chen, Yuling
    Man Nguyen
    Yen, Jesse T.
    ULTRASONIC IMAGING, 2011, 33 (03) : 205 - 216
  • [24] 3-D strain imaging using a rectilinear 2D array
    Awad, Samer I.
    Yen, Jesse T.
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 228 - 231
  • [25] Row-Column Addressed 2-D CMUT Arrays with Integrated Apodization
    Christiansen, Thomas Lehrmann
    Rasmussen, Morten Fischer
    Jensen, Jorgen Arendt
    Thomsen, Erik Vilain
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 600 - 603
  • [26] Three-Dimensional Shear Wave Elastography Using Acoustic Radiation Force and a 2-D Row-Column Addressing (RCA) Array
    Dong, Zhijie
    Lok, U-Wai
    Lowerison, Matthew R.
    Huang, Chengwu
    Chen, Shigao
    Song, Pengfei
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2024, 71 (04) : 448 - 458
  • [27] 3-D Vector Velocity Estimation with Row-Column Addressed Arrays
    Holbek, Simon
    Christiansen, Thomas Lehrmann
    Rasmussen, Morten Fischer
    Stuart, Matthias Bo
    Thomasen, Erik Vilain
    Jensen, Jorgen Arendt
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [28] Real-time 3-D image reconstruction of a 24x24 row-column addressing array-from raw data to image
    Li, Chunyu
    Yang, Jiali
    Li, Xu
    Zhong, Xiaoli
    Song, Junjie
    Ding, Mingyue
    Ming Yuchi
    MEDICAL IMAGING 2016: ULTRASONIC IMAGING AND TOMOGRAPHY, 2016, 9790
  • [29] Improved Focusing Method for 3-D Imaging using Row-Column-Addressed 2-D Arrays
    Bouzari, Hamed
    Engholm, Mathias
    Stuart, Matthias Bo
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [30] Ultrafast 3-D Ultrasound Imaging Using Row-Column Array-Specific Frame-Multiply-and-Sum Beamforming
    Hansen-Shearer, Joseph
    Lerendegui, Marcelo
    Toulemonde, Matthieu
    Tang, Meng-Xing
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2022, 69 (02) : 480 - 488