On-chip integrated hydrolysis, fluorescent labeling, and electrophoretic separation utilized for acetylcholinesterase assay

被引:13
|
作者
Heleg-Shabtai, Vered [1 ]
Gratziany, Natzach [1 ]
Liron, Zvi [1 ]
机构
[1] Israel Inst Biol Res, Dept Phys Chem, IL-74100 Ness Ziona, Israel
关键词
acetylcholinesterase electrophoretic separation; laser-induced fluorescence; microchip; paraoxon; O-ethyl S-[2-(diisopropylamino)ethyl] methylphosphonothiolate;
D O I
10.1016/j.aca.2006.04.087
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive on-chip acetylcholinesterase (AChE) assay that serves as a basis for the development of a fully integrated on-chip AChE-inhibitor detection assay is presented. The sequential steps required for the on-chip analysis process were integrated into a microchip. Transport and mixing of the reagents occurred by a combination of electroosmosis and electrophoresis using computer-controlled electrokinetic transport. AChE-catalyzed hydrolysis of acetylthiocholine to thiocholine was determined by on-chip reaction of thiocholine with eosinmaleimide, and the resulting thioether was electrophoretically separated and detected by laser-induced fluorescence (LIF). Enzyme-substrate mixing and reaction by confluent flow of reagents was compared with electrophoretically mediated microanalysis (EMMA), based on injection of an enzyme plug, and the utilization of differences in electrophoretic mobility as a driving force for efficient mixing and reaction. Both methods yielded similar results, however the EMMA-plug technique is preferable. The EMMA-plug technique was optimized for length and pushing time of enzyme plug, length of dyes mixture plug, acetylthiocholine concentration, and detector location. Detection of O-ethyl S-[2-(diisopropylamino)ethyl] methylphosphonothiolate (VX) and paraoxon, two AChE inhibitors, was demonstrated by off-chip mixing of the inhibitor and AChE, followed by the on-chip AChE assay. Limit of detection of VX for 5.5 min incubation and of paraoxon for 8 min incubation was 4 x 10(-10) and 4 x 10(-7) M, respectively. Utilization of the AChE microchip assay for inhibition kinetics was demonstrated also by evaluation of the inhibitor-enzyme bimolecular reaction constant (k(i)). The evaluated k(i) values for VX and paraoxon for AChE from the electric eel were 3.5 x 10(7) and 1.7 x 10(5) M-1 min(-1), respectively, conforming well to reported values obtained by bulk methods. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:228 / 234
页数:7
相关论文
共 30 条
  • [21] Silicon-micromachined separation columns coated with amino acid films for an integrated on-chip gas chromatograph
    Hannoe, S
    Sugimoto, I
    Katoh, T
    MICRO TOTAL ANALYSIS SYSTEMS '98, 2000, : 145 - 148
  • [22] Multilayer Polymer Microchip Capillary Array Electrophoresis Devices with Integrated On-Chip Labeling for High-Throughput Protein Analysis
    Yu, Ming
    Wang, Qingsong
    Patterson, James E.
    Woolley, Adam T.
    ANALYTICAL CHEMISTRY, 2011, 83 (09) : 3541 - 3547
  • [23] An All-Glass Microfluidic Network with Integrated Amorphous Silicon Photosensors for on-Chip Monitoring of Enzymatic Biochemical Assay
    Costantini, Francesca
    Tiggelaar, Roald M.
    Salvio, Riccardo
    Nardecchia, Marco
    Schlautmann, Stefan
    Manetti, Cesare
    Gardeniers, Han J. G. E.
    de Cesare, Giampiero
    Caputo, Domenico
    Nascetti, Augusto
    BIOSENSORS-BASEL, 2017, 7 (04):
  • [24] 3D Printed Microfluidic Devices for Solid-Phase Extraction and On-Chip Fluorescent Labeling of Preterm Birth Risk Biomarkers
    Bickham, Anna, V
    Pang, Chao
    George, Benjamin Q.
    Topham, David J.
    Nielsen, Jacob B.
    Nordin, Gregory P.
    Woolley, Adam T.
    ANALYTICAL CHEMISTRY, 2020, 92 (18) : 12322 - 12329
  • [25] A microfluidic flow injection system for DNA assay with fluids driven by an on-chip integrated pump based on capillary and evaporation effects
    Xu, Zhang-Run
    Zhong, Chong-Hui
    Guan, Yan-Xia
    Chen, Xu-Wei
    Wang, Jian-Hua
    Fang, Zhao-Lun
    LAB ON A CHIP, 2008, 8 (10) : 1658 - 1663
  • [26] Microfluidic Electrochemical Aptameric Assay Integrated On-Chip: A Potentially Convenient Sensing Platform for the Amplified and Multiplex Analysis of Small Molecules
    Du, Yan
    Chen, Chaogui
    Zhou, Ming
    Dong, Shaojun
    Wang, Erkang
    ANALYTICAL CHEMISTRY, 2011, 83 (05) : 1523 - 1529
  • [27] Tailoring of a TiO2 nanotube array-integrated portable microdevice for efficient on-chip enrichment and isotope labeling of serum phosphopeptides
    Min, Qianhao
    Chen, Xueqin
    Zhang, Xiaoxia
    Zhu, Jun-Jie
    LAB ON A CHIP, 2013, 13 (19) : 3853 - 3861
  • [28] Ultrasensitive optofluidic enzyme-linked immunosorbent assay by on-chip integrated polymer whispering-gallery-mode microlaser sensors
    Ouyang, Xia
    Liu, Tong
    Zhang, Yangxi
    He, Jijun
    He, Zijian
    Zhang, A. Ping
    Tam, Hwa-Yaw
    LAB ON A CHIP, 2020, 20 (14) : 2438 - 2446
  • [29] Monitoring On-Chip Pictet-Spengler Reactions by Integrated Analytical Separation and Label-Free Time-Resolved Fluorescence
    Ohla, Stefan
    Beyreiss, Reinhild
    Fritzsche, Stefanie
    Glaser, Petra
    Nagl, Stefan
    Stockhausen, Kai
    Schneider, Christoph
    Belder, Detlev
    CHEMISTRY-A EUROPEAN JOURNAL, 2012, 18 (04) : 1240 - 1246
  • [30] Development and evaluation of a real-time fluorogenic loop-mediated isothermal amplification assay integrated on a microfluidic disc chip (on-chip LAMP) for rapid and simultaneous detection of ten pathogenic bacteria in aquatic animals
    Zhou, Qian-Jin
    Wang, Lei
    Chen, Jiong
    Wang, Rui-Na
    Shi, Yu-Hong
    Li, Chang-Hong
    Zhang, De-Min
    Yan, Xiao-Jun
    Zhang, Yan-Jun
    JOURNAL OF MICROBIOLOGICAL METHODS, 2014, 104 : 26 - 35