On-chip integrated hydrolysis, fluorescent labeling, and electrophoretic separation utilized for acetylcholinesterase assay

被引:13
|
作者
Heleg-Shabtai, Vered [1 ]
Gratziany, Natzach [1 ]
Liron, Zvi [1 ]
机构
[1] Israel Inst Biol Res, Dept Phys Chem, IL-74100 Ness Ziona, Israel
关键词
acetylcholinesterase electrophoretic separation; laser-induced fluorescence; microchip; paraoxon; O-ethyl S-[2-(diisopropylamino)ethyl] methylphosphonothiolate;
D O I
10.1016/j.aca.2006.04.087
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive on-chip acetylcholinesterase (AChE) assay that serves as a basis for the development of a fully integrated on-chip AChE-inhibitor detection assay is presented. The sequential steps required for the on-chip analysis process were integrated into a microchip. Transport and mixing of the reagents occurred by a combination of electroosmosis and electrophoresis using computer-controlled electrokinetic transport. AChE-catalyzed hydrolysis of acetylthiocholine to thiocholine was determined by on-chip reaction of thiocholine with eosinmaleimide, and the resulting thioether was electrophoretically separated and detected by laser-induced fluorescence (LIF). Enzyme-substrate mixing and reaction by confluent flow of reagents was compared with electrophoretically mediated microanalysis (EMMA), based on injection of an enzyme plug, and the utilization of differences in electrophoretic mobility as a driving force for efficient mixing and reaction. Both methods yielded similar results, however the EMMA-plug technique is preferable. The EMMA-plug technique was optimized for length and pushing time of enzyme plug, length of dyes mixture plug, acetylthiocholine concentration, and detector location. Detection of O-ethyl S-[2-(diisopropylamino)ethyl] methylphosphonothiolate (VX) and paraoxon, two AChE inhibitors, was demonstrated by off-chip mixing of the inhibitor and AChE, followed by the on-chip AChE assay. Limit of detection of VX for 5.5 min incubation and of paraoxon for 8 min incubation was 4 x 10(-10) and 4 x 10(-7) M, respectively. Utilization of the AChE microchip assay for inhibition kinetics was demonstrated also by evaluation of the inhibitor-enzyme bimolecular reaction constant (k(i)). The evaluated k(i) values for VX and paraoxon for AChE from the electric eel were 3.5 x 10(7) and 1.7 x 10(5) M-1 min(-1), respectively, conforming well to reported values obtained by bulk methods. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:228 / 234
页数:7
相关论文
共 30 条
  • [1] An integrated microdevice for on-chip preconcentration, separation and labeling of proteins
    Petersen, D
    Foote, RS
    Geschke, O
    Ramsey, JM
    Micro Total Analysis Systems 2004, Vol 1, 2005, (296): : 117 - 119
  • [2] An integrated on-chip sirtuin assay
    Ohla, Stefan
    Beyreiss, Reinhild
    Scriba, Gerhard K. E.
    Fan, Yi
    Belder, Detlev
    ELECTROPHORESIS, 2010, 31 (19) : 3263 - 3267
  • [3] On-chip fluorescent labeling using reversed-phase monoliths and microchip electrophoretic separations of selected preterm birth biomarkers
    Sonker, Mukul
    Yang, Rui
    Sahore, Vishal
    Kumar, Suresh
    Woolley, Adam T.
    ANALYTICAL METHODS, 2016, 8 (43) : 7739 - 7746
  • [4] Electrophoretic mobility measurements of fluorescent dyes using on-chip capillary electrophoresis
    Milanova, Denitsa
    Chambers, Robert D.
    Bahga, Supreet S.
    Santiago, Juan G.
    ELECTROPHORESIS, 2011, 32 (22) : 3286 - 3294
  • [5] Tunable thick polymer coatings for on-chip electrophoretic protein and peptide separation
    He, Mei
    Zeng, Yong
    Jemere, Abebaw B.
    Harrison, D. Jed
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1241 : 112 - 116
  • [6] High-resolution electrophoretic separation and integrated-waveguide excitation of fluorescent DNA molecules in a lab on a chip
    Dongre, Chaitanya
    van Weerd, Jasper
    Besselink, Geert A. J.
    van Weeghel, Rob
    Vazquez, Rebeca Martinez
    Osellame, Roberto
    Cerullo, Giulio
    Cretich, Marina
    Chiari, Marcella
    Hoekstra, Hugo J. W. M.
    Pollnau, Markus
    ELECTROPHORESIS, 2010, 31 (15) : 2584 - 2588
  • [7] On-chip magnetic separation of superparamagnetic beads for integrated molecular analysis
    Florescu, Octavian
    Wang, Kevan
    Au, Patrick
    Tang, Jimmy
    Harris, Eva
    Beatty, P. Robert
    Boser, Bernhard E.
    JOURNAL OF APPLIED PHYSICS, 2010, 107 (05)
  • [8] On-chip pressure injection for integration of infrared-mediated DNA amplification with electrophoretic separation
    Easley, CJ
    Karlinsey, JM
    Landers, JP
    LAB ON A CHIP, 2006, 6 (05) : 601 - 610
  • [9] Time and length scales in governing equations and boundary conditions for on-chip electrophoretic sample separation
    Pezeshkpour, Pegah
    Schneider, Gerry
    Ren, Carolyn
    NUMERICAL HEAT TRANSFER PART A-APPLICATIONS, 2019, 76 (01) : 1 - 18
  • [10] Ultrashort Separation Length Homogeneous Electrophoretic Immunoassays Using On-Chip Discontinuous Polyacrylamide Gels
    Hou, Chenlu
    Herr, Amy E.
    ANALYTICAL CHEMISTRY, 2010, 82 (08) : 3343 - 3351