Accounting for pharmacokinetic differences in dual-tracer receptor density imaging

被引:24
|
作者
Tichauer, K. M. [1 ]
Diop, M. [2 ]
Elliott, J. T. [3 ]
Samkoe, K. S. [3 ]
Hasan, T. [4 ]
Lawrence, K. St. [2 ]
Pogue, B. W. [3 ,4 ]
机构
[1] IIT, Chicago, IL 60616 USA
[2] Univ Western Ontario, London, ON N6A 4V2, Canada
[3] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[4] Massachusetts Gen Hosp, Wellman Ctr Photomed, Boston, MA 02114 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 10期
关键词
fluorescence imaging; tracer kinetics; cancer; xenograft mouse model; epidermal growth factor receptor; affibodies; DRUG-BINDING; MODEL; TISSUES; VOLUME;
D O I
10.1088/0031-9155/59/10/2341
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dual-tracer molecular imaging is a powerful approach to quantify receptor expression in a wide range of tissues by using an untargeted tracer to account for any nonspecific uptake of a molecular-targeted tracer. This approach has previously required the pharmacokinetics of the receptor-targeted and untargeted tracers to be identical, requiring careful selection of an ideal untargeted tracer for any given targeted tracer. In this study, methodology capable of correcting for tracer differences in arterial input functions, as well as binding-independent delivery and retention, is derived and evaluated in a mouse U251 glioma xenograft model using an Affibody tracer targeted to epidermal growth factor receptor (EGFR), a cell membrane receptor overexpressed in many cancers. Simulations demonstrated that blood, and to a lesser extent vascular-permeability, pharmacokinetic differences between targeted and untargeted tracers could be quantified by deconvolving the uptakes of the two tracers in a region of interest devoid of targeted tracer binding, and therefore corrected for, by convolving the uptake of the untargeted tracer in all regions of interest by the product of the deconvolution. Using fluorescently labeled, EGFR-targeted and untargeted Affibodies (known to have different blood clearance rates), the average tumor concentration of EGFR in four mice was estimated using dual-tracer kinetic modeling to be 3.9 +/- 2.4 nM compared to an expected concentration of 2.0 +/- 0.4 nM. However, with deconvolution correction a more equivalent EGFR concentration of 2.0 +/- 0.4 nM was measured.
引用
收藏
页码:2341 / 2351
页数:11
相关论文
共 50 条
  • [1] Improving Efficiency of Simultaneous Dual-Tracer PET Imaging
    Li, Z.
    Shah, V.
    Viol, S. L. M.
    Van Snick, P.
    Luurtsema, G.
    Slart, R. H. J. A.
    Lammertsma, A.
    Tsoumpas, C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S705 - S705
  • [2] DYNAMIC OR EARLY IMAGING IN DUAL-TRACER PARATHYROID SCINTIGRAPHY
    GREENBERG, SB
    PARK, CH
    INTENZO, C
    CLINICAL NUCLEAR MEDICINE, 1986, 11 (09) : 627 - 628
  • [3] Dual-tracer receptor concentration imaging using tracers with different tissue delivery kinetics
    Tichauer, Kenneth M.
    Diop, Mamadou
    Elliott, Jonathan T.
    Samkoe, Kimberley S.
    Hasan, Tayyaba
    St Lawrence, Keith
    Pogue, Brian W.
    REPORTERS, MARKERS, DYES, NANOPARTICLES, AND MOLECULAR PROBES FOR BIOMEDICAL APPLICATIONS VI, 2014, 8956
  • [4] Rapid dual-tracer PET imaging using deep image prior powered by hierarchical pharmacokinetic modeling
    Ding, Wenxiang
    Wang, Hanzhong
    Feng, Dagan
    Shi, Kuangyu
    Qiu, Huang
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [5] Dynamic dual-tracer MRI-guided fluorescence tomography to quantify receptor density in vivo
    Davis, Scott C.
    Samkoe, Kimberley S.
    Tichauer, Kenneth M.
    Sexton, Kristian J.
    Gunn, Jason R.
    Deharvengt, Sophie J.
    Hasan, Tayyaba
    Pogue, Brian W.
    PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2013, 110 (22) : 9025 - 9030
  • [6] Joint Direct Dynamic Analysis in Dual-Tracer PET Imaging
    Zhu, Wentao
    Feng, Tao
    Chen, Mu
    Dong, Yun
    Bao, Jun
    Li, Hongdi
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [7] Pixel-based absorption correction for dual-tracer fluorescence imaging of receptor binding potential
    Kanick, Stephen C.
    Tichauer, Kenneth M.
    Gunn, Jason
    Samkoe, Kimberley S.
    Pogue, Brian W.
    BIOMEDICAL OPTICS EXPRESS, 2014, 5 (10): : 3280 - +
  • [8] Dual-Tracer Parathyroid Imaging Using Joint SPECT Reconstruction
    Jaruwan Onwanna
    Maythinee Chantadisai
    Tawatchai Chaiwatanarat
    Yothin Rakvongthai
    Nuclear Medicine and Molecular Imaging, 2023, 57 : 126 - 136
  • [9] Predicting the outcome of peptide receptor radionuclide therapy in neuroendocrine tumors: the importance of dual-tracer imaging
    Zanotti-Fregonara, Paolo
    Morgat, Clement
    Champion, Christophe
    Hindie, Elif
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 (06) : 1095 - 1096
  • [10] Initial studies of PET-SPECT dual-tracer imaging
    Shao, Yiping
    Yao, Rutao
    Ma, Tianyu
    Manchiraju, Phani
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4198 - +