Comparison of additional pseudophakic multifocal lenses and multifocal intraocular lens in the capsular bag

被引:9
|
作者
Schrecker, J. [1 ]
Feith, A. [1 ]
Langenbucher, A. [2 ]
机构
[1] Rudolf Virchow Klinikum Glauchau, Dept Ophthalmol, D-08371 Glauchau, Germany
[2] Univ Saarland, Chair Expt Ophthalmol, Homburg, Germany
关键词
CONTRAST SENSITIVITY; IMPLANTATION;
D O I
10.1136/bjophthalmol-2013-304591
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims To compare the clinical results and the intraoperative and postoperative performance of an additional sulcus-fixated multifocal intraocular lens (IOL) with a conventional multifocal IOL implanted in the capsular bag. Methods In this prospective, randomised, monocentric study, the intraoperative and postoperative outcomes after implantation of an additional sulcus-fixated multifocal IOL (group A; 29 eyes) and a conventional, multifocal posterior chamber IOL (group B; 25 eyes) are compared. 1-year postoperatively, visual acuity, contrast sensitivity and defocus curve were determined. Patients implanted bilaterally with the same IOL filled in a questionnaire to assess patient satisfaction. Results Between the groups, only slight differences were found in visual acuity at all distances. All patients achieved an uncorrected distance visual acuity of 0.1 LogMAR or better and an uncorrected intermediate/near visual acuity of 0.3 LogMAR or better. Slightly better results in contrast sensitivity were achieved in group A at high spatial frequencies under almost all lighting conditions. Patient survey showed a higher degree of satisfaction of patients in group B, while a smaller proportion of group A stated to have disturbing light phenomena. Conclusions Only small functional differences between the examined methods were found. Both types of multifocal IOL performed well at all distances. There were no intraoperative and postoperative complications in the add-on IOL group which affirms the high safety level such as with conventional multifocal IOL in the capsular bag.
引用
收藏
页码:915 / 919
页数:5
相关论文
共 50 条
  • [1] Additional multifocal sulcus-based intraocular lens: Alternative to multifocal intraocular lens in the capsular bag
    Schrecker, Jens
    Kroeber, Sandra
    Eppig, Timo
    Langenbucher, Achim
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (04): : 548 - 555
  • [2] Clinical results of a multifocal pseudophakic additional lens
    Schrecker, J.
    Langenbucher, A.
    OPHTHALMOLOGE, 2015, 112 (02): : 148 - 154
  • [3] Assessment of dysphotopsia in pseudophakic subjects with multifocal intraocular lenses
    Buckhurst, Phillip J.
    Naroo, Shehzad A.
    Davies, Leon N.
    Shah, Sunil
    Drew, Tom
    Wolffsohn, James S.
    BMJ OPEN OPHTHALMOLOGY, 2016, 1 (01):
  • [4] Dual intraocular lens implantation: Monofocal lens in the bag and additional diffractive multifocal lens in the sulcus
    Gerten, Georg
    Kermani, Omid
    Schmiedt, Karl
    Farvili, Elham
    Foerster, Andreas
    Oberheide, Uwe
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (12): : 2136 - 2143
  • [5] Lens Surgery: Use of multifocal Intraocular Lenses
    Shajari, Mehdi
    Mayer, Wolfgang J.
    Priglinger, Siegfried
    AUGENHEILKUNDE UP2DATE, 2019, 9 (04) : 347 - 357
  • [6] Lens Surgery: Implantation of Multifocal Intraocular Lenses
    Shajari, Mehdi
    Mayer, Wolfgang J.
    Priglinger, Siegfried
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2019, 236 (12) : 1461 - 1469
  • [8] INTRAOCULAR-LENS POWER CALCULATIONS FOR MULTIFOCAL INTRAOCULAR LENSES
    HOLLADAY, JT
    HOFFER, KJ
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1992, 114 (04) : 405 - 408
  • [9] Different multifocal intraocular lenses - Effect on lens epithelial cells in vitro on human capsular bags
    Liekfeld, A
    Hartmann, C
    Kojetinsky, C
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U83 - U83
  • [10] Multifocal intraocular lenses
    Shajari, Mehdi
    Mayer, Wolfgang J.
    Priglinger, Siegfried
    Kreutzer, Thomas
    SPEKTRUM DER AUGENHEILKUNDE, 2019, 33 (06) : 165 - 168