Cooling gradients and formation of gaseous microemboli with cardiopulmonary bypass: An echocardiographic study

被引:22
|
作者
Geissler, HJ [1 ]
Allen, SJ [1 ]
Mehlhorn, U [1 ]
Davis, KL [1 ]
deVivie, ER [1 ]
Kurusz, M [1 ]
Butler, BD [1 ]
机构
[1] UNIV TEXAS,SCH MED,DEPT ANESTHESIOL,HOUSTON,TX 77030
来源
ANNALS OF THORACIC SURGERY | 1997年 / 64卷 / 01期
关键词
D O I
10.1016/S0003-4975(97)82824-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Previous studies demonstrated gas emboli formation during rewarming from hypothermia on cardiopulmonary bypass when the temperature gradient exceeded a critical threshold. It also has been suggested that formation of arterial gas emboli may occur during cooling on cardiopulmonary bypass when cooled oxygenated blood exiting the heat exchanger is warmed on mixture with the patient's blood. The purpose of this study was to determine under what circumstances gas emboli formation would occur during cooling on cardiopulmonary bypass. Methods. Eight anesthetized mongrel dogs were placed on cardiopulmonary bypass using a roller pump, membrane oxygenator, and arterial line filter. For emboli detection, we positioned a transesophageal echocardiographic probe at the aortic arch distal to the aortic cannula and Doppler probes at the common carotid artery and the arterial line. Cooling gradients between normothermic blood and cooled arterial perfusate of 5 degrees, 10 degrees, 15 degrees, 20 degrees, and 0 degrees C (isothermal controls) were investigated. In addition to preestablished temperature gradients, we investigated the effect of rapid cooling (maximal flow through the heat exchanger at a water bath temperature of 4 degrees C) after the initiation of normothermic cardiopulmonary bypass. Results. Minimal gas emboli were detected at the aortic arch at gradients of 10 degrees C or greater. The incidence of emboli was related directly to the magnitude of the temperature gradient (p < 0.01). No emboli were detected at the carotid artery. During rapid cooling, no emboli were observed either at the aorta or at the carotid artery. Conclusions. Cooling gradients of 10 degrees C or greater may be associated with gas emboli formation, but they may be of limited clinical significance because no emboli were detected distal to the aortic arch. During the application of rapid cooling, no emboli formation was observed. (C) 1997 by The Society of Thoracic Surgeons.
引用
收藏
页码:100 / 104
页数:5
相关论文
共 50 条
  • [31] Commentary on: An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli
    Resley, Justin
    Melchior, Richard
    Beavers, James
    Shearer, Ian
    Riley, Jeffrey
    Groom, Robert
    PERFUSION-UK, 2010, 25 (04): : 265 - 266
  • [32] BRAIN MICROEMBOLI ASSOCIATED WITH CARDIOPULMONARY BYPASS - A HISTOLOGIC AND MAGNETIC-RESONANCE-IMAGING STUDY
    MOODY, DM
    BROWN, WR
    CHALLA, VR
    STUMP, DA
    REBOUSSIN, DM
    LEGAULT, C
    ANNALS OF THORACIC SURGERY, 1995, 59 (05): : 1304 - 1307
  • [33] Axillary artery cannulation for cardiopulmonary bypass reduces cerebral microemboli
    Hedayati, N
    Sherwood, JT
    Schomisch, SJ
    Carino, JJ
    Markowitz, AH
    CIRCULATION, 2002, 106 (19) : 576 - 576
  • [34] CONTINUOUS DETECTION OF MICROEMBOLI DURING CARDIOPULMONARY BYPASS IN ANIMALS AND MAN
    CLARK, RE
    DIETZ, DR
    MILLER, JG
    CIRCULATION, 1976, 54 (06) : 74 - 78
  • [35] Impact of the intensity of microemboli on neurocognitive outcome following cardiopulmonary bypass
    Doganci, S.
    Gunaydin, S.
    Kocak, O. Murat
    Yilmaz, S.
    Demirkilic, U.
    PERFUSION-UK, 2013, 28 (03): : 256 - 262
  • [36] DISPOSABLE FILTER FOR MICROEMBOLI IN CARDIOPULMONARY BYPASS - CLINICAL AND HISTOLOGICAL FINDINGS
    SKAGSETH, E
    FROYSAKER, T
    REFSUM, SB
    JOURNAL OF CARDIOVASCULAR SURGERY, 1974, 15 (03): : 318 - 322
  • [37] DISPOSABLE FILTER FOR MICROEMBOLI - USE IN CARDIOPULMONARY BYPASS AND MASSIVE TRANSFUSION
    PATTERSON, RH
    TWICHELL, JB
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1971, 215 (01) : 76 - +
  • [38] Axillary artery cannulation for cardiopulmonary bypass reduces cerebral microemboli
    Hedayati, N
    Sherwood, JT
    Schomisch, SJ
    Carino, JL
    Markowitz, AH
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2004, 128 (03): : 386 - 390
  • [39] CONTINUOUS DETECTION OF MICROEMBOLI DURING CARDIOPULMONARY BYPASS IN ANIMALS AND MAN
    CLARK, RE
    DIETZ, DR
    MILLER, JG
    CIRCULATION, 1975, 52 (04) : 74 - 74
  • [40] In Vitro Comparison of the Delivery of Gaseous Microemboli and Hemodynamic Energy for a Diagonal and a Roller Pump During Simulated Infantile Cardiopulmonary Bypass Procedures
    Dhami, Ranjodh
    Wang, Shigang
    Kunselman, Allen R.
    Uendar, Akif