ECG-based gating in cardiac MR imaging requires additional patient preparation time, is susceptible to RF and magnetic interference, and is ineffective in a significant percentage of patients. "Wireless" or "self-gating" techniques have been described using either interleaved central k-space lines or projection reconstruction to obtain MR signals synchronous with the cardiac cycle. However, the interleaved, central line method results in a doubling of the acquisition time, while radial streak artifacts are encountered with the projection reconstruction method. In this work, a new self-gating technique is presented to overcome these limitations. A retrospectively gated TrueFISP cine sequence was modified to acquire a short second echo after the readout and phase gradients are rewound. The information obtained from this second echo was used to derive a gating signal. This technique was compared to ECG-based gating in 10 healthy volunteers and shown to have no significant difference in image quality. The results indicate that this method could serve as an alternative gating strategy without the need for external physiological signal detection. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:782 / 788
页数:7
相关论文
共 50 条
  • [1] Self-Gated Cardiac Cine Imaging Using Phase Information
    Seo, Hyunseok
    Kim, Dongchan
    Oh, Changheun
    Park, HyunWook
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (03) : 1216 - 1222
  • [2] Self-gated cardiac cine MRI
    Larson, AC
    White, RD
    Laub, G
    McVeigh, ER
    Li, DB
    Simonetti, OP
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (01) : 93 - 102
  • [3] Self-gated PROPELLER-encoded cine cardiac imaging
    Wang, Chi-Chung
    Huang, Teng-Yi
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (06): : 1477 - 1485
  • [4] Self-gated PROPELLER-encoded cine cardiac imaging
    Chi-Chung Wang
    Teng-Yi Huang
    [J]. The International Journal of Cardiovascular Imaging, 2012, 28 : 1477 - 1485
  • [5] Self-Gated Fat-Suppressed Cardiac Cine MRI
    Ingle, R. Reeve
    Santos, Juan M.
    Overall, William R.
    McConnell, Michael V.
    Hu, Bob S.
    Nishimura, Dwight G.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (05) : 1764 - 1774
  • [6] AUTOMATED CARDIAC SELF-GATED RADIAL CMRI
    Yang, Xiahan
    Zheng, Yahong Rosa
    Ma, Lixin
    [J]. 2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS, 2015, : 365 - 368
  • [7] Cardiac and respiratory self-gated cine MRI in the mouse: Comparison between radial and rectilinear techniques at 7T
    Hiba, Bassem
    Richard, Nathalie
    Thibault, Helene
    Janier, Marc
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2007, 58 (04) : 745 - 753
  • [8] Comparison of self-gated cine MRI retrospective cardiac synchronization algorithms
    Nijm, Grace M.
    Sahakian, Alan V.
    Swiryn, Steven
    Carr, James C.
    Sheehan, John J.
    Larson, Andrew C.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (03) : 767 - 772
  • [9] Wireless self-gated multiple-mouse cardiac cine MRI
    Esparza-Coss, Emilio
    Ramirez, Marc S.
    Bankson, James A.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) : 1203 - 1206
  • [10] Cardiac and respiratory double self-gated cine MRI in the mouse at 7 T
    Hiba, B
    Richard, N
    Janier, M
    Croisille, P
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (03) : 506 - 513