Grip force monitoring on the hand: Manugraphy system versus Jamar dynamometer

被引:24
|
作者
Muehldorfer-Fodor, Marion [1 ]
Ziegler, Steffen [1 ]
Harms, Christoph [2 ]
Neumann, Julia [2 ]
Cristalli, Alessandra [3 ]
Kalpen, Axel [3 ]
Kundt, Guenther [4 ]
Mittlmeier, Thomas [2 ]
Prommersberger, K. J. [1 ]
机构
[1] Rhon Klinikum AG, Clin Hand Surg, D-97616 Bad Neustadt an der Saale, Germany
[2] Univ Rostock, Dept Trauma & Reconstruct Surg, D-18057 Rostock, Germany
[3] Novel Biomech Lab, D-81675 Munich, Germany
[4] Univ Rostock, Inst Biostat & Informat Med & Ageing Res, D-18057 Rostock, Germany
关键词
Dynamic force measurement; Grip force; Grip strength; Jamar dynamometer; Manugraphy system; Sensor matrix; Pressure sensor mat; PINCH STRENGTH; FINGER FORCES; SIZE; RELIABILITY; VALIDITY; OBJECT;
D O I
10.1007/s00402-014-2027-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
For clinical grip force assessment, the Jamar dynamometer is a wide accepted tool. Users have to be aware that this method does not represent all grip efforts applied. The Manugraphy system is a tool that measure total grip force as well as identify load distribution patterns of the hand while gripping cylinders wrapped with calibrated capacitive matrix sensor mats. The aim of this study was to validate an assessment setting of the Manugraphy system for clinical use. Further, the relationship and difference between the Manugraphy system and the Jamar dynamometer were investigated. At two study centers, 152 healthy volunteers performed grip force tests with a digital Jamar dynamometer using handle positions 3 and 4 and the novel(A (R)) Manugraphy system using two cylinders with circumferences of 150 and 200 mm. The subjects performed grip force testing with both devices on three different days. The intra- and inter-day variability for both methods was evaluated. To compare the values of both systems, the Spearman correlation coefficient was calculated. The force values, as measured by the sensor matrix, were higher than those of the Jamar dynamometer. Analyses showed significant positive correlations between values obtained by the two measurement methods (p < 0.001). There was no significant inter-day variation for the 200-mm cylinder of the Manugraphy system. For the 150-mm cylinder, a significant variation was observed at center B, but not at A. Nevertheless, the fluctuation of the grip force values obtained with the Manugraphy system was equal or better than those obtained with the Jamar dynamometer. The force values, obtained using the two systems, have a high correlation but are not directly comparable. Both systems allow valid and constant grip force measurement. As the sensor mat detects all forces applied perpendicularly to the cylinder surface, it characterizes grip force better than the Jamar dynamometer. In addition, information about load distribution of the hand is gained.
引用
收藏
页码:1179 / 1188
页数:10
相关论文
共 50 条
  • [41] Dynamometer measurements of hand-grip strength predict multiple outcomes
    Bohannon, RW
    [J]. PERCEPTUAL AND MOTOR SKILLS, 2001, 93 (02) : 323 - 328
  • [42] COMPARISON OF STOELTING HAND GRIP DYNAMOMETER AND LINEAR VOLTAGE DIFFERENTIAL TRANSFORMER FOR MEASURING MAXIMAL GRIP STRENGTH
    HEYWARD, V
    MCKEOWN, B
    GEESEMAN, R
    [J]. RESEARCH QUARTERLY, 1975, 46 (02): : 262 - 266
  • [43] Development and application of a multi-axis dynamometer for measuring grip force
    Irwin, C. B.
    Towles, J. D.
    Radwin, R. G.
    [J]. ERGONOMICS, 2013, 56 (12) : 1841 - 1849
  • [44] A Surface Electromyography (sEMG) System Applied for Grip Force Monitoring
    Wu, Dantong
    Tian, Peng
    Zhang, Shuai
    Wang, Qihang
    Yu, Kang
    Wang, Yunfeng
    Gao, Zhixing
    Huang, Lin
    Li, Xiangyu
    Zhai, Xingchen
    Tian, Meng
    Huang, Chengjun
    Zhang, Haiying
    Zhang, Jun
    [J]. SENSORS, 2024, 24 (12)
  • [45] Reference values for Jamar+ digital dynamometer hand grip strength in healthy adults and in adults with non-communicable diseases or osteoarthritis: the Norwegian Tromsø study 2015–2016
    Odd-Einar Svinøy
    Gunvor Hilde
    Astrid Bergland
    Bjørn Heine Strand
    [J]. European Journal of Ageing, 2023, 20
  • [46] Grip Force Control in Individuals with Hand Osteoarthritis
    de Oliveira, Diana Guimaraes
    Nunes, Paula Martins
    Aruin, Alexander S.
    dos Santos, Marcio Jose
    [J]. JOURNAL OF HAND THERAPY, 2011, 24 (04) : 345 - 355
  • [47] A video game based hand grip system for measuring muscle force in children
    Mark Gotthelf
    DeWayne Townsend
    William Durfee
    [J]. Journal of NeuroEngineering and Rehabilitation, 18
  • [48] A video game based hand grip system for measuring muscle force in children
    Gotthelf, Mark
    Townsend, DeWayne
    Durfee, William
    [J]. JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2021, 18 (01)
  • [49] Perception of Hand Force in Power Grip for Females
    Li, Kai Way
    Yu, Ruifeng
    Zhang, Wei
    [J]. HUMAN FACTORS AND ERGONOMICS IN MANUFACTURING & SERVICE INDUSTRIES, 2013, 23 (02) : 77 - 84
  • [50] Reference Values of Grip Strength Measured with a Jamar Dynamometer in 1526 Adults with Intellectual Disabilities and Compared to Adults without Intellectual Disability
    Cuesta-Vargas, Antonio
    Hilgenkamp, Thessa
    [J]. PLOS ONE, 2015, 10 (06):