Preliminary results of Magnetic Resonance Imaging (MRI) of cardiac function in mice in vivo, using the homebuilt gradient coils and specialized RF probehead are demonstrated. An unshielded gradient system with inner diameter of 60 mm was designed and constructed for the 4.7T/310 magnet with MARAN DRX console (Resonance Instruments). Dedicated probehead, constructed to fit the gradient system, consists of the RF birdcage coil and the animal handling system. ECG-triggered cine images of eight FVB (wild-type) and four transgenic mice with heart failure (Tg alpha q*44) were acquired at physiological temperature (37 degrees C) with a good quality of multislice images in different phases of the cardiac cycle, as well as a good contrast between myocardium and flowing blood. This technique allowed the calculation of the end-diastolic (EDV) and end-systolic (ESV) volumes of working heart that could be used to monitor the development of heart failure in vivo in Tg alpha q*44 mice.
机构:
King Saud Med City, Dept Med Imaging, Riyadh 12746, Saudi Arabia
Univ Queensland, Ctr Adv Imaging, St Lucia, Qld 4072, AustraliaKing Saud Med City, Dept Med Imaging, Riyadh 12746, Saudi Arabia
Alqarni, Faiz
论文数: 引用数:
h-index:
机构:
Alsaadi, Mohammed
Karem, Fayka
论文数: 0引用数: 0
h-index: 0
机构:
King Saud Med City, Dept Med Imaging, Riyadh 12746, Saudi Arabia
Al Azhar Univ, Fac Med Girls, Radiol & Med Imaging Dept, POB 1175, Cairo, EgyptKing Saud Med City, Dept Med Imaging, Riyadh 12746, Saudi Arabia