Coded excitation using biphase-coded pulse with mismatched filters for high-frequency ultrasound imaging

被引:25
|
作者
Hu, Chang-Hong [1 ]
Liu, Ruibin
Zhou, Qifa
Yen, Jesse
Shung, K. Kirk
机构
[1] Univ So Calif, Biomed Engn Dept, Los Angeles, CA 90089 USA
[2] Univ So Calif, NIH Transducer Resource Ctr, Los Angeles, CA 90089 USA
关键词
coded excitation; mismatched filter; high-frequency ultrasound;
D O I
10.1016/j.ultras.2006.04.002
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A scheme of using phase-coded excitation and mismatched filter compression for high-frequency ultrasound imaging is presented in this paper. Biphase-coded pulses were constructed to excite the transducer. Received signals were compressed with mismatched filters optimized by minimizing peak-sidelobe-level (PSL). Both simulation and experiments were carried out to demonstrate the advantage of this technique. The simulation results demonstrated a possible sidelobe reduction (<-90 dB) with a slightly decrease of the signal-to-noise ratio of less than 1 dB compared with the compression using matched filters alone. The experimental results showed about 14 dB SNR improvement as well as -40 dB sidelobe level when the Barker-13 code excitation with 3-cycle sinusoidal wave carrier was used. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:330 / 336
页数:7
相关论文
共 50 条
  • [41] A Full Digital Design of TCD Ultrasound System Using Normal Pulse and Coded Excitation
    Li, Jie
    Diao, Xianfen
    Zhan, Kai
    Qin, Zhengdi
    [J]. 1ST GLOBAL CONFERENCE ON BIOMEDICAL ENGINEERING & 9TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 47 : 136 - +
  • [42] In vitro volume imaging of articular cartilage using chirp-coded high frequency ultrasound
    Habib, Anowarul
    Vierinen, Juha
    Islam, Ashraful
    Martinez, Inigo Zubiavrre
    Melandso, Frank
    [J]. 2018 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2018,
  • [43] Photoacoustic Coded Excitation using Pulse Position Modulation
    Beckmann, Martin F.
    Schmitz, Georg
    [J]. 2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 1853 - 1856
  • [44] Weighted Capon beamformer combined with coded excitation in ultrasound imaging
    Seyed Amin Izadi
    Ali Mahloojifar
    Babak Mohammadzadeh Asl
    [J]. Journal of Medical Ultrasonics, 2015, 42 : 477 - 488
  • [45] Weighted Capon beamformer combined with coded excitation in ultrasound imaging
    Izadi, Seyed Amin
    Mahloojifar, Ali
    Asl, Babak Mohammadzadeh
    [J]. JOURNAL OF MEDICAL ULTRASONICS, 2015, 42 (04) : 477 - 488
  • [46] Coded excitation with spectrum inversion (CEXSI) for ultrasound array imaging
    Wang, Y
    Metzger, K
    Stephens, DN
    Williams, G
    Brownlie, S
    O'Donnell, M
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (07) : 805 - 823
  • [47] The use of the fractional Fourier transform with coded excitation in ultrasound imaging
    Bennett, MJ
    McLaughlin, S
    Anderson, T
    McDicken, N
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (04) : 754 - 756
  • [48] Coded excitation and nonlinear pulse compression in pulse-inversion fundamental imaging
    Shen, Che-Chou
    Cheng, Yun-Chien
    Li, Pai-Chi
    [J]. 2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 1744 - 1747
  • [49] Coded excitation waveform engineering for high frame rate synthetic aperture ultrasound imaging
    Lashkari, Bahman
    Zhang, Kaicheng
    Dovlo, Edem
    Mandelis, Andreas
    [J]. ULTRASONICS, 2017, 77 : 121 - 132
  • [50] Resolution Enhanced Non-Contact Thermoacoustic Imaging using Coded Pulse Excitation
    Singhvi, Ajay
    Fitzpatrick, Aidan
    Arbabian, Amin