Comparison of tungsten coil electrothermal vaporization and thermospray sample introduction methods for flame furnace atomic absorption spectrometry

被引:24
|
作者
Wu, Peng [1 ]
Zhang, Yunchang [2 ]
Liu, Rui [2 ]
Lv, Yi [2 ]
Hou, Xiandeng [1 ,2 ]
机构
[1] Sichuan Univ, Analyt & Testing Ctr, Chengdu 610064, Sichuan, Peoples R China
[2] Sichuan Univ, Coll Chem, Chengdu 610064, Sichuan, Peoples R China
基金
中国国家自然科学基金;
关键词
Flame furnace; Atomic absorption spectrometry; Tungsten coil; Electrothermal vaporization; Thermospray; TS-FF-AAS; EMISSION-SPECTROMETRY; CADMIUM; PRECONCENTRATION; URINE; PB; CD; EXTRACTION; GENERATION; SYSTEM;
D O I
10.1016/j.talanta.2008.10.017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Flame furnace atomic absorption spectrometry (FF-AAS) is a newly developed flame atomic absorption spectrometric technique based on arranging a flame furnace onto the top of the flame burner head. In this fundamental investigation, 25 elements were carefully tested by using either thermospray FF-AAS or tungsten coil electrothermal vaporization FF-AAS, of which 15 volatile and semi-volatile elements (Cd, Tl, Ag, Pb, Zn, Hg, Cu, Sb, Bi, Te, In, As. Se, Sn and Au) exhibited better limits of detection compared to those by conventional FAAS: however, non-volatile or refractory elements (Fe. Co, Ni, Cr, Mn, Pd, Pt, Al, Be and V) showed inferior sensitivities by the proposed methods. (C) 2008 Elsevier B.V. All rights reserved.