Electromembrane Surrounded Solid Phase Microextraction Followed by Injection Port Derivatization and Gas Chromatography-Flame Ionization Detector Analysis for Determination of Acidic Herbicides in Plant Tissue

被引:32
|
作者
Rezazadeh, Maryam [1 ]
Yamini, Yadollah [1 ]
Seidi, Shahram [2 ]
Tahmasebi, Elham [1 ]
Rezaei, Fatemeh [1 ]
机构
[1] Tarbiat Modares Univ, Dept Chem, Tehran, Iran
[2] KN Toosi Univ Technol, Dept Analyt Chem, Fac Chem, Tehran 163151355, Iran
关键词
electromembrane surrounded solid phase microextraction; gas chromatography; acidic compounds; chlorophenoxy acid herbicides; plant tissue; ELECTRO MEMBRANE EXTRACTION; BASIC DRUGS; HOLLOW-FIBER; MASS SPECTROMETRY; WATER SAMPLES; COMBINATION; PYROLYSIS; MIGRATION; PLASMA; URINE;
D O I
10.1021/jf500017r
中图分类号
S [农业科学];
学科分类号
09 ;
摘要
Electromembrane surrounded solid phase microextraction (EM-SPME) of acidic herbicides was studied for the first time. In order to investigate the capability of this new microextraction technique to analyze acidic targets, chlorophenory acid (CPA) herbicides were quantified in plant tissue. 1-Octanol, was sustained in the pores of the wall of a hollow fiber and served as supported liquid membrane (SLM). Other EM-SPME related parameters, including extraction time, applied voltage, and pHs of the sample solution and the acceptor phase, were optimized using experimental design. A 20 min time frame was needed to reach the highest extraction efficiency of the analytes from a 24 mL alkaline sample solution across the organic liquid membrane and into the aqueous acceptor phase through a 50 V electrical field, and to their final adsorption on a carbonaceous anode. In addition to high sample cleanup, which made the proposed method appropriate for analysis of acidic compounds in a complicated media (plant tissue), 4.8% of 2-methyl-4-chlorophenoxyacetic acid (MCPA) and 0.6% of 2,4-dichlorophenoxyacetic acid (2,4-D) were adsorbed on the anode, resulting in suitable detection limits (less than 5 ng mL(-1)), and admissible repeatability and reproducibility (intra- and interassay precision were in the ranges of 5.2-8.5% and 8.8-12.096, respectively). Linearity of the method was scrutinized within the ranges of 1.0-500.0 and 10.0-500.0 ng mL(-1) for MCPA and 2,4-D, respectively, and coefficients of determination greater than 0.9958 were obtained. Optimal conditions of EM-SPME of the herbicides were employed for analysis of CPAs in whole wheat tissue.
引用
收藏
页码:3134 / 3142
页数:9
相关论文
共 50 条