Impact on Image Noise of Incorporating Detector Blurring Into Image Reconstruction for a Small Animal PET Scanner

被引:6
|
作者
Lee, Kisung [1 ]
Miyaoka, Robert S. [2 ]
Lewellen, Tom K. [2 ]
Alessio, Adam M. [2 ]
Kinahan, Paul E. [2 ]
机构
[1] Korea Univ, Dept Radiol Sci, Seoul 136703, South Korea
[2] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
基金
美国国家卫生研究院;
关键词
Detector blurring; Fourier rebinning; noise property; ordered subsets expectation maximization (OSEM); positron emission tomography; RESOLUTION; ALGORITHMS; EM;
D O I
10.1109/TNS.2009.2021610
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We study the noise characteristics of an image reconstruction algorithm that incorporates a model of the non-stationary detector blurring (DB) for a mouse-imaging positron emission tomography (PET) scanner. The algorithm uses ordered subsets expectation maximization (OSEM) image reconstruction., which is used to suppress statistical noise. Including the non-stationary detector blurring in the reconstruction process [OSEM(DB)] has been shown to increase contrast in images reconstructed from measured data acquired on the fully-3D MiCES PET scanner developed at the University of Washington. As an extension, this study uses simulation studies with a fully-3D acquisition mode and our proposed FORE+OSEM(DB) reconstruction process to evaluate the volumetric contrast versus noise trade-offs of this approach. Multiple realizations were simulated to estimate the true noise properties of the algorithm. The results show that incorporation of detector blurring FORE+OSEM(DB) into the reconstruction process improves the contrast/noise trade-offs compared to FORE+OSEM in a radially dependent manner. Adding post reconstruction 3D Gaussian smoothing to FORE+OSENT and FORE+OSEM(DB) reduces the contrast versus noise advantages of FORE+OSEM(DB).
引用
收藏
页码:2769 / 2776
页数:8
相关论文
共 50 条
  • [1] Contribution of normalization to image noise for the Siemens Inveon small-animal PET scanner
    Visser, Eric P.
    Disselhorst, Jonathan A.
    Laverman, Peter
    Gotthardt, Martin
    Oyen, Wim J. G.
    Boerman, Otto C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 605 (03): : 433 - 435
  • [2] Trade-Off between Contrast Recovery, Image Noise and Edge Artifacts in PET Image Reconstruction Using Detector Blurring Models
    Ahn, Sangtae
    Asma, Evren
    Thielemans, Kris
    Deller, Timothy W.
    Ross, Steven G.
    Stearns, Charles W.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4110 - 4114
  • [3] Efficient image reconstruction for a small animal PET system with dual-layer-offset detector design
    Cheng, Li
    Lyu, Zhenlei
    Liu, Hui
    Wu, Jing
    Jia, Chao
    Wu, Yuanguang
    Ji, Yingcai
    Jiang, Nianming
    Ma, Tianyu
    Liu, Yaqiang
    MEDICAL PHYSICS, 2024, 51 (04) : 2772 - 2787
  • [4] Theoretical Analysis of Resolution and Noise Properties of PET Image Reconstruction with and without Sinogram Blurring Modeling
    Ahn, Sangtae
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [5] Scanner Dependent Noise Properties of the Q.Clear PET Image Reconstruction Tool
    Lantos, Judit
    Iagaru, Andrei
    Levin, Craig S.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [6] TOF Benefits and Trade-offs on Image Contrast-to-Noise Ratio Performance for a Small Animal PET Scanner
    Yussefian, Nikta Zarif
    Toussaint, Maxime
    Gaudin, Emilie
    Lecomte, Roger
    Fontaine, Rejean
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (05) : 687 - 693
  • [7] Evaluation of 3D image reconstruction methods for a dual-head small-animal PET scanner
    Kao, Chien-Min
    Dong, Yun
    Xie, Qingguo
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3046 - +
  • [8] Improving PET image spatial resolution by experimental measurement of scanner blurring properties
    De Bernardi, E
    Zito, F
    Michelutti, L
    Mainardi, L
    Gerundini, P
    Baselli, G
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 975 - 977
  • [9] Scanning Multiple Mice in a Small-Animal PET Scanner: Influence on Image Quality
    Disselhorst, J. A.
    Siepel, F. J.
    van Lier, M. G. J. T.
    Chen, M.
    Meeuwis, A. P. W.
    Oyen, W. J. G.
    Boerman, O. C.
    Visser, E. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S294 - S294
  • [10] Scanning multiple mice in a small-animal PET scanner: Influence on image quality
    Siepel, Francoise J.
    van Lier, Monique G. J. T. B.
    Chen, Mu
    Disselhorst, Jonathan A.
    Meeuwis, Antoi P. W.
    Oyen, Wim J. G.
    Boerman, Otto C.
    Visser, Eric P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 605 - 610