Technical note: A 3D-printed phantom for routine accuracy check of Gamma Knife Icon HDMM system

被引:4
|
作者
Wu, Chuan [1 ]
Radevic, Marlyn B. [2 ]
Glass, Jennifer S. [2 ]
Skubic, Stan E. [2 ]
机构
[1] Sutter Med Fdn, Radiat Oncol, Roseville, CA 95661 USA
[2] Sutter Med Fdn, Radiat Oncol, Sacramento, CA USA
来源
关键词
Gamma Knife; HDMM; Icon; phantom; QA;
D O I
10.1002/acm2.12339
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To report a novel 3D-printed device (SH phantom) that is designed for routine accuracy check of the Gamma Knife Icon High Definition Motion Management (HDMM) system. Methods: SH phantom was designed using tinkerCAD software and printed on a commercial 3D printer. We evaluated the SH phantom on our Gamma Knife Icon unit regarding its usability and accuracy for routine HDMM QA. Results: Single-axis and multiple-axis measurements validated the SH phantom design and implementation. An HDMM QA accuracy of 0.22 mm or better along single axis was found using SH phantom. Conclusions: The SH phantom proved to be a quick and simple tool to use to perform the HDMM system QA. The SH phantom was tested successfully and adopted by us as part of monthly QA for the Gamma Knife Icon.
引用
收藏
页码:299 / 301
页数:3
相关论文
共 50 条
  • [1] A New Physical Phantom for Routine Quantitative QA of Gamma Knife ICON HDMM System
    Wu, C.
    Radevic, M.
    Glass, J.
    Skubic, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2880 - 2880
  • [2] Technical note: 3D-printed phantom for dedicated cardiac protocols and geometries in nuclear medicine
    Green, Samantha
    Grice, Jared
    MEDICAL PHYSICS, 2022, 49 (02) : 943 - 951
  • [3] 3D-Printed Headrest for Frameless Gamma Knife Radiosurgery: Design and Validation
    Baltz, G.
    Briere, T.
    Luo, D.
    Howell, R.
    Krafft, S.
    Han, E.
    MEDICAL PHYSICS, 2020, 47 (06) : E675 - E675
  • [4] 3D-printed headrest for frameless Gamma Knife radiosurgery: Design and validation
    Baltz, Garrett C.
    Briere, Tina
    Luo, Dershan
    Howell, Rebecca M.
    Krafft, Shane
    Han, Eun Young
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (09): : 6 - 15
  • [5] Technical Note: Immunohistochemical evaluation of mouse brain irradiation targeting accuracy with 3D-printed immobilization device
    Zarghami, Niloufar
    Jensen, Michael D.
    Talluri, Srikanth
    Foster, Paula J.
    Chambers, Ann F.
    Dick, Frederick A.
    Wong, Eugene
    MEDICAL PHYSICS, 2015, 42 (11) : 6507 - 6513
  • [6] Soft 3D-Printed Phantom of the Human Kidney with Collecting System
    Fabian Adams
    Tian Qiu
    Andrew Mark
    Benjamin Fritz
    Lena Kramer
    Daniel Schlager
    Ulrich Wetterauer
    Arkadiusz Miernik
    Peer Fischer
    Annals of Biomedical Engineering, 2017, 45 : 963 - 972
  • [7] Soft 3D-Printed Phantom of the Human Kidney with Collecting System
    Adams, Fabian
    Qiu, Tian
    Mark, Andrew
    Fritz, Benjamin
    Kramer, Lena
    Schlager, Daniel
    Wetterauer, Ulrich
    Miernik, Arkadiusz
    Fischer, Peer
    ANNALS OF BIOMEDICAL ENGINEERING, 2017, 45 (04) : 963 - 972
  • [8] Technical note: 3D-printed MRI-compatible syringe pump
    Holland, Martin D.
    Lee, Seth
    Kim, Harrison
    MEDICAL PHYSICS, 2023, 50 (11) : 7071 - 7082
  • [9] Technical Note: A 3D-printed phantom for radiochromic film evaluation of moving lung tumor SBRT without dose convolution
    Retif, Paul
    Djibo Sidikou, Abdourahamane
    Letellier, Romain
    Verrecchia-Ramos, Emilie
    Quetin, Philippe
    MEDICAL PHYSICS, 2021, 48 (07) : 3453 - 3458
  • [10] Accuracy of the 3D Reference Coordinate Definition with G-Frame for Gamma Knife ICON
    Claps, L.
    Mathew, D.
    Watanabe, Y.
    MEDICAL PHYSICS, 2020, 47 (06) : E574 - E575