Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure

被引:22
|
作者
Fujimura, Fusako [1 ,2 ]
Kamiya, Kazutaka [2 ,3 ]
Fujiwara, Kazuko [2 ]
Shoji, Nobuyuki [1 ,2 ]
Shimizu, Kimiya [2 ,3 ]
机构
[1] Kitasato Univ, Sch Allied Hlth Sci, Orthopt & Visual Sci Course, Dept Rehabil,Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[2] Kitasato Univ Hosp, Dept Ophthalmol, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[3] Kitasato Univ, Sch Med, Dept Ophthalmol, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
关键词
APPLANATION TONOMETRY;
D O I
10.1155/2013/370592
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Purpose. To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP. Methods. Forty-six eyes of healthy volunteersweremeasured by two examiners. Three consecutivemeasurements per eyewere performed. Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P. Results. The coefficient of variation for IOP was 6.4% and for CCT was 0.4%. The 95% limits of agreement between examiners were -0.17 +/- 1.42 mmHg (range: -2.95 to 2.61mmHg) for IOP, -0.93 +/- 4.37.. m (range: -9.50 to 7.64.. m) for CCT. The corrected IOP was significantly higher than the noncorrected IOP (p= 0.010.3). The noncorrected IOP significantly correlated with CCT (r= -0.4883,.. = 0.0006). The corrected IOP showed no significant correlation with CCT (r= -0.0285, p= 0.8509). Conclusions. NT-530P offered repeatability and reproducibility in both IOP and CCT measurements. The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Short-term variation in central corneal thickness and in raocular pressure using the Tono-Pachymeter NT530P (Tonopachy (TM))
    Mashigea, K. P.
    Rampersad, N.
    Jhetam, S.
    Govender, P.
    AFRICAN VISION AND EYE HEALTH JOURNAL, 2012, 71 (01): : 12 - 21
  • [2] Short-term variation of central corneal thickness measurements in keratoconus using the Tono-Pachymeter NT530P (Tonopachy (TM))
    Mashige, K. P.
    Oduntan, O. A.
    Gcabashe, N. M.
    AFRICAN VISION AND EYE HEALTH JOURNAL, 2012, 71 (03):
  • [3] Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes
    Garcia-Resua, Carlos
    Pena-Verdeal, Hugo
    Minones, Mercedes
    Jesus Giraldez, M.
    Yebra-Pimentel, Eva
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2013, 96 (03) : 286 - 294
  • [4] Accuracy and Repeatability of a New Tono-Pachymeter for Measuring Central Corneal Thickness
    Garcia-Resua, Carlos
    Blanco, Antia
    Minones, Mercedes
    Yebra-Pimentel, Eva
    Giraldez, Maria Jesus
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2012, 38 (03): : 158 - 163
  • [5] Positive correlation between Tono-Pen intraocular pressure and central corneal thickness
    Dohadwala, AA
    Munger, R
    Damji, KF
    OPHTHALMOLOGY, 1998, 105 (10) : 1849 - 1854
  • [6] Comparison of central corneal thickness and intraocular pressure measured with two different tono/pachymeter devices in non-glaucomatouse children
    Teberik, Kuddusi
    Eski, Mehmet Tahir
    Ankarali, Handan
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (06) : 3010 - 3017
  • [7] A comparison of intraocular pressure values obtained with the Tono-Pachymeter NT530P iCare (R) rebound tonometer and Goldmann applanation tonometer
    Rampersad, N.
    Mashige, K. P.
    Jhetamt, S.
    AFRICAN VISION AND EYE HEALTH JOURNAL, 2011, 70 (03):
  • [8] Effect of central corneal thickness and radius of the corneal curvature on intraocular pressure measured with the tono-pen and noncontact tonometer in healthy schoolchildren
    Yildirim, Nigun
    Sahin, Afiun
    Basmak, Hikmet
    Bal, Cengiz
    JOURNAL OF PEDIATRIC OPHTHALMOLOGY & STRABISMUS, 2007, 44 (04) : 216 - 222
  • [9] Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter
    Gokcinar, Nesrin Buyuktortop
    Yumusak, Erhan
    Ornek, Nurgul
    Yorubulut, Serap
    Onaran, Zafer
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (07) : 1589 - 1598
  • [10] Repeatability and agreement of central corneal thickness measurements with a new handheld non-contact pachymeter
    Lawrenson, John G. G.
    Gill, Simranjit
    Masuid, Isra
    Rashid, Fardip
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2023, 43 (06) : 1524 - 1530