Assessment of Prosthesis Alignment after Revision Total Knee Arthroplasty Using EOS 2D and 3D Imaging: A Reliability Study

被引:26
|
作者
Meijer, Marrigje F. [1 ]
Boerboom, Alexander L. [1 ]
Stevens, Martin [1 ]
Bulstra, Sjoerd K. [1 ]
Reininga, Inge H. F. [2 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Dept Orthopaed, Groningen, Netherlands
[2] Univ Groningen, Univ Med Ctr Groningen, Dept Trauma Surg, Groningen, Netherlands
来源
PLOS ONE | 2014年 / 9卷 / 09期
关键词
LOWER-EXTREMITY; REPRODUCIBILITY; REHABILITATION; AGREEMENT; POSITION; ROTATION; SYSTEM;
D O I
10.1371/journal.pone.0104613
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Introduction: A new low-dose X-ray device, called EOS, has been introduced for determining lower-limb alignment in 2D and 3D. Reliability has not yet been assessed when using EOS on lower limbs containing a knee prosthesis. Therefore purpose of this study was to determine intraobserver and interobserver reliability of EOS 2D and 3D knee prosthesis alignment measurements after revision total knee arthroplasty (rTKA). Methods: Forty anteroposterior and lateral images of 37 rTKA patients were included. Two observers independently performed measurements on these images twice. Varus/valgus angles were measured in 2D (VV2D) and 3D (VV3D). Intraclass correlation coefficients and the Bland and Altman method were used to determine reliability. T-tests were used to test potential differences. Results: Intraobserver and interobserver reliability were excellent for VV2D and VV3D. No significant difference or bias between the first and second measurements or the two observers was found. A significant mean and absolute difference of respectively 1.00 degrees and 1.61 degrees existed between 2D and 3D measurements. Conclusions: EOS provides reliable varus/valgus measurements in 2D and 3D for the alignment of the knee joint with a knee prosthesis. However, significant differences exist between varus/valgus measurements in 2D and 3D.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] 2D versus 3D templating in total knee arthroplasty
    Ettinger, Max
    Claassen, Leif
    Paes, Peter
    Calliess, Tilman
    KNEE, 2016, 23 (01): : 149 - 151
  • [2] Evaluation of implant alignment in navigated unicompartmental knee arthroplasty : a comparison of 2D and 3D imaging
    Ishida, Kazunari
    Toda, Akihiko
    Shibanuma, Nao
    Matsumoto, Tomoyuki
    Kuroda, Ryosuke
    Kurosaka, Masahiro
    ACTA ORTHOPAEDICA BELGICA, 2015, 81 (04): : 654 - 661
  • [3] Pelvic tilt and functional acetabular position after total hip arthroplasty: an EOS 2D/3D radiographic study
    Loppini, Mattia
    Pisano, Alessandro
    Ruggeri, Riccardo
    Della Rocca, Antonello
    Grappiolo, Guido
    HIP INTERNATIONAL, 2023, 33 (03) : 365 - 370
  • [4] Assessment of pedicle size in patients with scoliosis using EOS 2D imaging: a validity and reliability study
    C. M. M. Peeters
    L. van Houten
    D. H. R. Kempen
    F. H. Wapstra
    P. C. Jutte
    I. van den Akker-Scheek
    C. Faber
    European Spine Journal, 2021, 30 : 3473 - 3481
  • [5] Assessment of pedicle size in patients with scoliosis using EOS 2D imaging: a validity and reliability study
    Peeters, C. M. M.
    van Houten, L.
    Kempen, D. H. R.
    Wapstra, F. H.
    Jutte, P. C.
    van den Akker-Scheek, I
    Faber, C.
    EUROPEAN SPINE JOURNAL, 2021, 30 (12) : 3473 - 3481
  • [6] Revision total knee arthroplasty using a novel 3D printed titanium augment: A biomechanical cadaveric study
    Dion, C.
    Yamomo, G.
    Howard, J.
    Teeter, M.
    Willing, R.
    Lanting, B.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2020, 110
  • [7] Imaging in revision total knee arthroplasty: A novel 3D classification system for tibial bone defects
    Brenneis, Marco
    Flevas, Dimitrios A.
    Braun, Sebastian
    Sculco, Peter K.
    Boettner, Friedrich
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2024, 32 (02) : 323 - 333
  • [8] Validating a Novel 2D to 3D Knee Reconstruction Method on Preoperative Total Knee Arthroplasty Patient Anatomies
    Factor, Shai
    Gurel, Ron
    Dan, Dor
    Benkovich, Guy
    Sagi, Amit
    Abialevich, Artsiom
    Benkovich, Vadim
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (05)
  • [9] 2D/3D registration for measurement of implant alignment after total hip replacement
    Jaramaz, Branislav
    Eckman, Kort
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2006, PT 2, 2006, 4191 : 653 - 661
  • [10] Assessment Accuracy of 2D vs. 3D Imaging for Custom-Made Acetabular Implants in Revision Hip Arthroplasty
    Nees, Timo Albert
    Mueller, Christian Thomas
    Innmann, Moritz Maximilian
    Spranz, David Maximilian
    Westhauser, Fabian
    Renkawitz, Tobias
    Reiner, Tobias
    Walker, Tilman
    JOURNAL OF PERSONALIZED MEDICINE, 2024, 14 (08):