Metamaterial lens applicator for microwave hyperthermia of breast cancer

被引:19
|
作者
Wang, Gang [1 ]
Gong, Yu [1 ]
机构
[1] Jiangsu Univ, Dept Telecommun Engn, Zhenjiang 212013, Peoples R China
基金
中国国家自然科学基金;
关键词
Breast tumour; hyperthermia; microwave; left-handed metamaterial (LHM); lens applicator; TARGET DETECTION; HEATING-SYSTEM; ANTENNA; FUTURE; FOCUS; HEAD;
D O I
10.1080/02656730903061609
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Artificial left-handed metamaterial (LHM) provides a new perspective for microwave hyperthermia. Four flat LHM slab lenses can be used to form a focus-flexible applicator for breast tumour hyperthermia. By adjusting microwave sources behind the four flat LHM lenses, microwaves emitted from the sources can be focused tightly at different points in the breast tissue so that necessary heating depth in breast tissue can be achieved. Numerical simulations with a two-dimensional finite-difference time-domain method indicate that hyperthermia with the proposed four-lens applicator of moderate LHM losses could be effective in achieving desired power deposition in a heterogeneous breast model. Temperature distribution obtained by solving the bio-heat transfer equation demonstrates that temperature above 43 degrees C can be maintained in the tumour volume for specific periods of time. Flat slab LHM lenses offer a feasible alternative to traditional mechanically scanned lens applicator and electronically scanned phased-array applicators.
引用
收藏
页码:434 / 445
页数:12
相关论文
共 50 条
  • [21] Optimal design for microwave hyperthermia applicator
    Morega, Mihaela
    Mogos, Laura
    Neagu, M.
    Morega, Al.
    PROCEEDINGS OF THE 10TH INTERNATIONAL CONFERENCE ON OPTIMIZATION OF ELECTRICAL AND ELECTRONIC EQUIPMENT, VOL I: ELECTROTECHNICS, 2006, : 219 - 224
  • [22] Microwave Applicator for Soft Tissue Hyperthermia
    Morega, Mihaela
    Morega, Alexandru
    Machedon, Alina
    NONLINEAR OPTICS QUANTUM OPTICS-CONCEPTS IN MODERN OPTICS, 2009, 39 (2-3): : 151 - 163
  • [23] Microwave applicator for soft tissue hyperthermia
    Morega, Mihaela
    Morega, Alexandru
    MacHedon, Alina
    Nonlinear Optics Quantum Optics, 2009, 39 (2-3): : 151 - 163
  • [24] Slotted applicator for microwave local hyperthermia
    Safarik, Jakub
    Vrba, Jan
    PROCEEDINGS OF THE 14TH CONFERENCE ON MICROWAVE TECHNIQUES: COMITE 2008, 2008, : 275 - 278
  • [25] Analysis and Design of Microwave Dielectric Heating with Curved Plate Applicator for Deep Hyperthermia in Breast Cancer Treatment
    Kotchapradit, Supawat
    Thongsopa, Chanchai
    Thosdeekoraphat, Thanaset
    RADIOENGINEERING, 2019, 28 (04) : 703 - 713
  • [26] A wearable applicator for microwave hyperthermia of breast cancer: performance evaluation with patient-specific anatomic models
    Curto, S.
    Ruvio, G.
    Ammann, M. J.
    Prakash, P.
    PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON ELECTROMAGNETICS IN ADVANCED APPLICATIONS (ICEAA), 2015, : 1159 - 1162
  • [27] Development of a metamaterial-based compact applicator for hyperthermia
    Sharma, Nitika
    Singh, Hari Shankar
    Khanna, Rajesh
    Kaur, Amanpreet
    Agarwal, Mayank
    MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2023, 65 (02) : 525 - 533
  • [28] Left-Handed Metamaterial Lens Applicator with Built-in Cooling Feature for Superficial Tumor Hyperthermia
    Tao, Yonghui
    Yang, Erfu
    Wang, Gang
    APPLIED COMPUTATIONAL ELECTROMAGNETICS SOCIETY JOURNAL, 2017, 32 (11): : 1029 - 1034
  • [29] STUDY ON SEMICYLINDRICAL MICROSTRIP APPLICATOR FOR MICROWAVE HYPERTHERMIA
    SHIMOTORI, T
    NIKAWA, Y
    MORI, S
    IEICE TRANSACTIONS ON ELECTRONICS, 1994, E77C (06) : 942 - 948
  • [30] Compact Applicator Antenna for Superficial Microwave Hyperthermia
    Duque, Jose L.
    Gonzalez, Juan F.
    Araque, Javier L.
    2024 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND INC/USNCURSI RADIO SCIENCE MEETING, AP-S/INC-USNC-URSI 2024, 2024, : 2643 - 2644