A drop tower for controlled impact testing of biological tissues

被引:40
|
作者
Burgin, Leanne V. [1 ]
Aspden, Richard M. [1 ]
机构
[1] Univ Aberdeen, Dept Orthopaed Surg, Aberdeen, Scotland
关键词
impact; articular cartilage; mechanical properties; drop tower; mechanical testing;
D O I
10.1016/j.medengphy.2006.06.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Impact damage, in particular to tissues such as articular cartilage, is a recognised source of morbidity. To understand better the clinical outcomes, it is important to know the mechanics of the damage sustained and the biological response of cells to rapidly applied forces and subsequent tissue disruption. An instrumented drop tower has been designed to enable controlled impact loads to be applied to small samples of biological materials. Impact severity can be controlled by using impactors of different masses and various drop heights. Force and deceleration at impact are recorded at 50,000 samples s(-1) by a force transducer under the sample and an accelerometer on the impactor. Repeatability was tested on rubber washers and coefficients of variation were found to be better than 8% for dynamic stiffness, 3.4% for stress and 4.3% for strain. Initial tests on isolated biopsies of articular cartilage showed that at an initial strain rate of 916 s(-1), the peak dynamic modulus of human femoral head cartilage was 59 MPa, and for a bovine biopsy the initial strain rate and corresponding peak dynamic modulus were 3380 s(-1) and 130 Wa, respectively. The equipment described is capable of applying an impact load to small biopsies of tissue with a defined energy and velocity and measuring deformation and load at high rates of loading. (c) 2006 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:525 / 530
页数:6
相关论文
共 50 条
  • [31] Impact studies on structural components using a free-flight drop tower
    S. Bhattacharya
    K. C. Krishnamurthy
    R. Rajendran
    K. Prem sai
    S. Basu
    Experimental Techniques, 2006, 30 : 52 - 58
  • [32] Method of testing very soft biological tissues in compression
    Miller, K
    JOURNAL OF BIOMECHANICS, 2005, 38 (01) : 153 - 158
  • [33] Impact protection of borosilicate glass plates with elastomeric coatings in drop tower tests
    Henderson, Christine N.
    Monteith, John
    Solis-Ramos, Euripides
    Godard, Rachel
    Predecki, Paul
    Kumosa, Maciej
    INTERNATIONAL JOURNAL OF IMPACT ENGINEERING, 2020, 137
  • [34] Impact studies on structural components using a free-flight drop tower
    Bhattacharya, S
    Krishnamurthy, KC
    Rajendran, R
    Sai, KP
    Basu, S
    EXPERIMENTAL TECHNIQUES, 2006, 30 (02) : 52 - 58
  • [35] Correlating Drop Impact Simulations With Drop Impact Testing Using High-Speed Camera Measurements
    Zaal, J. J. M.
    van Driel, W. D.
    Kessels, F. J. H. G.
    Zhang, G. Q.
    JOURNAL OF ELECTRONIC PACKAGING, 2009, 131 (01) : 0110071 - 0110079
  • [36] IMPACT TESTING OF CONCRETE USING A DROP-WEIGHT IMPACT MACHINE
    BANTHIA, N
    MINDESS, S
    BENTUR, A
    PIGEON, M
    EXPERIMENTAL MECHANICS, 1989, 29 (01) : 63 - 69
  • [37] DROP-WEIGHT IMPACT SENSITIVITY TESTING OF EXPLOSIVES
    BOYARS, C
    LEVINE, D
    PYRODYNAMIC, 1968, 6 (01): : 53 - &
  • [38] Quantitative Testing of Robustness on Superomniphobic Surfaces by Drop Impact
    Nguyen, Thi Phuong Nhung
    Brunet, Philippe
    Coffinier, Yannick
    Boukherroub, Rabah
    LANGMUIR, 2010, 26 (23) : 18369 - 18373
  • [39] TRANSITION RANGE DROP TOWER J-R CURVE TESTING OF A106 STEEL
    JOYCE, JA
    HACKETT, EM
    EXPERIMENTAL MECHANICS, 1989, 29 (03) : 274 - 278
  • [40] Impact testing of structural biological materials
    Lee, Steve
    Novitskaya, Ekaterina E.
    Reynante, Brandon
    Vasquez, Joshua
    Urbaniak, Robert
    Takahashi, Tsukasa
    Woolley, Evan
    Tombolato, Luca
    Chen, Po-Yu
    McKittrick, Joanna
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2011, 31 (04): : 730 - 739