Analytic and quality control validation and assessment of field performance of a point-of-care chemistry analyzer for use in the White rhinoceros

被引:8
|
作者
Hooijberg, Emma H. [1 ]
Steenkamp, Gerhard [1 ]
du Preez, Jacques P. [1 ]
Goddard, Amelia [1 ]
机构
[1] Univ Pretoria, Fac Vet Sci, Dept Compan Anim Clin Studies, Pretoria, South Africa
关键词
Ceratotherium simum; dry chemistry; total allowable error; VetTest; wildlife; PLASMA BIOCHEMISTRY; VETERINARY-MEDICINE; GUIDELINES; ASSURANCE; ANTICOAGULANTS; SERUM;
D O I
10.1111/vcp.12456
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
BackgroundA chemistry point-of-care analyzer would be useful for evaluating injured wildlife, particularly White rhinoceros (Ceratotherium simum) that survive poaching attempts. The IDEXX VetTest could be suitable, but species-specific validation, development of a statistical quality control (QC) strategy, and evaluation under field conditions are necessary. ObjectivesThe objectives were to (1) validate the VetTest for the White rhinoceros, (2) perform QC validation on the VetTest and generate a statistical QC strategy, and (3) apply this QC strategy to monitor performance under typical field conditions. MethodsDifferences between White rhinoceros heparin plasma and serum, short-term imprecision, and reportable range using rhinoceros plasma and long-term imprecision using commercial quality control material (QCM) were assessed against prescribed total allowable error (TEa) for up to 15 analytes. Quality control validation was performed using data from the long-term imprecision study and TEa. A QC strategy using QCM was developed and used to monitor performance under field conditions. ResultsImprecision was acceptable for all analytes except for ALP, ALT, and AST at low activities. The reportable range for AST and LDH differed from the manufacturer's specifications. Eleven analytes were suitable for statistical QC using the 1(3s) rule, 3 using the 2(s) rule; ALP was not suitable. In the field, observed error was < TEa for all 15 analytes and the sigma metric was > 3.0 for 12 analytes. ConclusionsThe VetTest is suitable for use in the White rhinoceros. Statistical QC is possible for most analytes and useful for evaluation of field performance.
引用
收藏
页码:100 / 110
页数:11
相关论文
共 34 条
  • [31] Innovative methodology for point-of-care circulating cathodic antigen with rapid urine concentration for use in the field for detecting low Schistosoma mansoni infection and for control of cure with high accuracy
    Grenfell, Rafaella F. Q.
    Taboada, Diana
    Coutinho, Luclia A.
    Pedrosa, Maria Luysa C.
    Assis, Jessica V.
    Oliveira, Matheus S. P.
    Cruz, Renata R.
    Almeida, Aureo
    Silva-Moraes, Vanessa
    Katz, Naftale
    Coelho, Paulo Marcos Z.
    TRANSACTIONS OF THE ROYAL SOCIETY OF TROPICAL MEDICINE AND HYGIENE, 2018, 112 (01) : 1 - 7
  • [32] "Point-of-Care Blood Glucose Testing: Post-Market Performance Assessment of the Accu-Chek Inform II Hospital-Use Glucose Meter" (vol 95, 12, 2023)
    Petryaykina, Elena E.
    Mayanskiy, Nikolay A.
    Demina, Elena S.
    Karamysheva, Irina V.
    Gorst, Kseniya A.
    Timofeev, Alexei V.
    TERAPEVTICHESKII ARKHIV, 2024, 96 (03) : 312 - 314
  • [33] COMPARISON OF A UNIT-USE BLOOD GAS ANALYZER (BGA) WITH A STANDARD LAB BGA DEVICE AS TWO OPTIONS FOR POINT-OF-CARE TECHNIQUE (POCT) DURING ROUTINE PRACTICE ON A SURGICAL ICU: DATA QUALITY, DEVIATIONS, FEASIBILITY, AND IMPACT FACTORS
    Toussaint, S.
    von Wachsmann, D.
    Gerlach, H.
    INTENSIVE CARE MEDICINE, 2012, 38 : S286 - S286
  • [34] Updated Requirements for Measurement Quality and Quality Assurance of Point-Of-Care Testing (POCT) - Blood Glucose Measurement Systems with Unit-Use Reagents Suitable for the Initial Diagnosis of Diabetes Manifested inPregnancy or Gestational Diabetes Mellitus (GDM) According to the GDM Guideline of the German Diabetes Association (DDG): Consensus Recommendation by the German United Society for Clinical Chemistry and Laboratory Medicine (DGKL) and the German Diabetes Association 2015
    Neumaier, M.
    Luppa, P. B.
    Koschinsky, T.
    Siegel, E.
    Freckmann, G.
    Heinemann, L.
    DIABETOLOGIE UND STOFFWECHSEL, 2015, 10 (04) : E1 - E3