Biomechanical Comparison of Polyaxial-Type Locking Plates and a Fixed-Angle Locking Plate for Internal Fixation of Distal Femur Fractures

被引:68
|
作者
Otto, Randall J. [1 ]
Moed, Berton R. [1 ]
Bledsoe, J. Gary [2 ]
机构
[1] St Louis Univ, Sch Med, Dept Orthopaed Surg, St Louis, MO 63110 USA
[2] St Louis Univ, Pk Coll Engn Aviat & Technol, Dept Biomed Engn, St Louis, MO 63110 USA
关键词
polyaxial; distal femur; locking plate; fracture fixation; biomechanics; INVASIVE STABILIZATION SYSTEM; BLADE PLATE; SCREW;
D O I
10.1097/BOT.0b013e3181a567c8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To test the stability to axial loading of 2 new polyaxial locking screw-plate designs and analyze different angles of screw insertion. The noncontact bridging (NCB) polyaxial locking plate (Zimmer) and the POLYAX plate (DePuy) were compared with a fixed-angle less invasive stabilization system (LISS; Synthes). Methods: Twenty-five synthetic femurs were divided into 5 groups and assigned fixation with the LISS plate (group I), POLYAX plate (groups IIA and IIB), or NCB plate (groups IIIA and IIIB). The polyaxial constructs were divided into parallel and crossed distal condylar screw configurations. Each construct was tested under axial loading and stressed to failure at a displacement rate of 5 mm/min with a preload of 100 N. Outcome measurements included stiffness, load to failure, peak force, and mode of failure, Results: All LISS and POLYAX constructs failed by plastic deformation of the plate, whereas 9 of 10 NCB constructs failed by an intraarticular lateral condyle fracture. No failures occurred at the screwplate interface in either polyaxial constructs. Load to failure of the LISS was 33% greater than the parallel POLYAX (P < 0.01) and 24% greater than the crossed POLYAX (P < 0.01). Load to failure of NCB (parallel and crossed) were 24% greater than the parallel POLYAX (P < 0.01 and P < 0.01, respectively) and 1501,, greater than the crossed POLYAX (P < 0.01 and P = 0.02, respectively). The POLYAX also had significantly lower stiffness and peak force compared with the LISS and NCB. There was no difference between the LISS and NCB with regard to stiffness, load to failure, and peak force. Parallel and crossed polyaxial constructs showed no difference in stiffness or failure loads. Conclusions: There were no failures of either polyaxial screw-plate interface despite large forces and screw angle did not affect the overall strength of these constructs, supporting the biomechanical soundness of both polyaxial device designs under axial loading. However, the POLYAX supported smaller loads compared with the LISS and NCB while under axial loading. In addition, the mode of failure of the NCB plate, creating an intra-articular fracture propagating from the distal posterior screw hole, may be of some concern. Additional testing is needed to determine the clinical importance of the demonstrated differences among these plate designs.
引用
收藏
页码:645 / 652
页数:8
相关论文
共 50 条
  • [1] Polyaxial locking plate fixation in distal femur fractures: A biomechanical comparison
    Wilkens, Kenneth J.
    Curtiss, Shane
    Lee, Mark A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (09) : 624 - 628
  • [2] THE CHETTINAD EXPERIENCE WITH FIXED-ANGLE LOCKING PLATE FIXATION FOR FRACTURES OF DISTAL TIBIA
    Chander, S. G.
    Kanagasarthy, K.
    Govindarajan, Balaji
    Reddy, M. Narayana
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2015, 4 (25): : 4308 - 4316
  • [3] Comparative study between fixed-angle and polyaxial screws in distal radius fixation with two volar locking plates
    Toro-Aguilera, Alvaro
    Martinez-Galarza, Pablo
    Camacho-Carrasco, Pilar
    Caballero, Miguel
    Segur, Josep M.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2021, 107 (05)
  • [4] Biomechanical comparison of a lateral polyaxial locking plate with a posterolateral polyaxial locking plate applied to the distal fibula
    Hallbauer, Jakob
    Klos, Kajetan
    Rausch, Sascha
    Graefenstein, Andreas
    Wipf, Felix
    Beimel, Claudia
    Hofmann, Gunther
    Mueckley, Thomas
    FOOT AND ANKLE SURGERY, 2014, 20 (03) : 180 - 185
  • [5] Results in treatment of distal femur fractures using polyaxial locking plate
    Pascarella, R.
    Bettuzzi, C.
    Bosco, G.
    Leonetti, D.
    Dessi, S.
    Forte, P.
    Amendola, L.
    STRATEGIES IN TRAUMA AND LIMB RECONSTRUCTION, 2014, 9 (01): : 13 - 18
  • [6] Mono- versus polyaxial locking plates in distal femur fractures – a biomechanical comparison of the Non-Contact-Bridging- (NCB) and the PERILOC-plate
    Bilal Farouk El-Zayat
    Turgay Efe
    Steffen Ruchholtz
    Salim Khatib
    Nina Timmesfeld
    Antonio Krüger
    Ralph Zettl
    BMC Musculoskeletal Disorders, 15
  • [7] Mono- versus polyaxial locking plates in distal femur fractures - a biomechanical comparison of the Non-Contact-Bridging- (NCB) and the PERILOC-plate
    El-Zayat, Bilal Farouk
    Efe, Turgay
    Ruchholtz, Steffen
    Khatib, Salim
    Timmesfeld, Nina
    Krueger, Antonio
    Zettl, Ralph
    BMC MUSCULOSKELETAL DISORDERS, 2014, 15
  • [8] Comparison Between External Fixators and Fixed-Angle Volar-Locking Plates in the Treatment of Distal Radius Fractures
    Agustin Jorge-Mora, Alberto
    Cecilia-Lopez, David
    Rodriguez-Vega, Victor
    Suarez-Arias, Laura
    Andres-Esteban, Eva
    Porras-Moreno, Miguel Angel
    Resines-Erasun, Carlos
    JOURNAL OF HAND AND MICROSURGERY, 2012, 4 (02) : 50 - 54
  • [9] Distal radius measurements and efficacy of fixed-angle locking volar plates
    Agir, Ismail
    Aytekin, Mahmut Nedim
    Kucukdurmaz, Fatih
    Basci, Onur
    Tetik, Cihangir
    TURKISH JOURNAL OF MEDICAL SCIENCES, 2014, 44 (01) : 36 - 41
  • [10] Complication Rate Comparing Variable Angle Distal Locking Plate to Fixed Angle Plate Fixation of Distal Radius Fractures
    Mehrzad, Raman
    Kim, David C.
    ANNALS OF PLASTIC SURGERY, 2016, 77 (06) : 623 - 625