High-accuracy comparison of lightning and switching impulse calibrators

被引:8
|
作者
Wakimoto, Takayuki [1 ]
Hallstrom, Jari K.
Chekurov, Y.
Ishii, M.
Lucas, Wolfgang
Piiroinen, J.
Shimizu, Hiroyuki
机构
[1] Nippon Inst Technol, Saitama 3458501, Japan
[2] Helsinki Univ Technol, Mittaustekniika Metrol Res Inst Teknillinen Korke, Espoo 02015, Finland
[3] Univ Tokyo, Inst Ind Sci, Tokyo 1538505, Japan
[4] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
calibration; impulse voltage; international comparison;
D O I
10.1109/TIM.2007.891060
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper summarizes the results of a comparison on two calculable impulse voltage calibrators and one commercial calibrator between Helsinki University of Technology, Finland (MIKES-TKK), Nippon Institute of Technology, Japan (NIT), and Physikalisch-Technische Bundesanstalt, Germany (PTB). The peak voltage levels of the comparison ranged from 10 to 300 V. Both lightning and switching impulse voltage calibrators were studied using both positive and negative polarity impulses. The comparison was performed by applying impulses from the calibrators to a 12-b digitizer and then comparing the reported digitizer errors. For the switching impulse peak value, the agreement between the two calculable calibrators was better than 0.02% for all the measured voltages and better than 0.4% for the time parameters. For the lightning impulse peak value, the disagreement between the same two calibrators was smaller than 0.05% for all voltages examined and smaller than 1% for the time parameters.
引用
收藏
页码:619 / 623
页数:5
相关论文
共 50 条
  • [31] High-accuracy liquid biopsies
    Beatriz Bellosillo
    Clara Montagut
    Nature Medicine, 2019, 25 : 1820 - 1821
  • [32] A high-accuracy blackbody for CLARREO
    Latvakoski, Harri
    Watson, Mike
    Topham, Shane
    Scott, Deron
    Wojcik, Mike
    Bingham, Gail
    INFRARED REMOTE SENSING AND INSTRUMENTATION XVIII, 2010, 7808
  • [33] A NOVEL PYRHELIOMETER OF HIGH-ACCURACY
    ZHANG, ZM
    GE, XS
    WANG, YF
    SOLAR ENERGY, 1987, 39 (05) : 371 - 377
  • [34] High-Accuracy Gaging with SPC
    不详
    MANUFACTURING ENGINEERING, 2009, 143 (02): : 27 - 27
  • [35] HIGH-ACCURACY BRIDGE AMPLIFIER
    SHELL, D
    ELECTRONICS WORLD & WIRELESS WORLD, 1990, 96 (1653): : 582 - 583
  • [36] High-accuracy linear scale
    不详
    MANUFACTURING ENGINEERING, 2007, 138 (05): : 56 - +
  • [37] A high-accuracy switching loss model of SiC MOSFETs in a motor drive for electric vehicles
    Ding, Xiaofeng
    Lu, Peng
    Shan, Zhenyu
    APPLIED ENERGY, 2021, 291
  • [38] HIGH-ACCURACY SAMPLING WATTMETER
    STENBAKKEN, GN
    DOLEV, A
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1992, 41 (06) : 974 - 978
  • [39] HIGH-ACCURACY INTERPOLATION OF STRESSES
    BERES, E
    ACTA TECHNICA ACADEMIAE SCIENTIARUM HUNGARICAE, 1980, 91 (3-4): : 257 - 263
  • [40] High-Accuracy Heading Determination
    El-Osery, Aly
    Bruder, Stephen
    Laughlin, Darren
    2013 8TH INTERNATIONAL CONFERENCE ON SYSTEM OF SYSTEMS ENGINEERING (SOSE), 2013, : 308 - 313