Distributed Pressure Sensors for a Urethral Catheter

被引:0
|
作者
Ahmadi, Mahdi [1 ]
Rajamani, Rajesh [1 ]
Timm, Gerald [2 ]
Sezen, A. S. [1 ]
机构
[1] Univ Minnesota, Dept Mech Engn, St Paul, MN 55455 USA
[2] Univ Minnesota, Dept Urol, St Paul, MN 55455 USA
来源
2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2015年
关键词
URINARY-INCONTINENCE; PREVALENCE; EXERCISE;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A flexible strip that incorporates multiple pressure sensors and is capable of being fixed to a urethral catheter is developed. The urethral catheter thus instrumented will be useful for measurement of pressure in a human urethra during urodynamic testing in a clinic. This would help diagnose the causes of urinary incontinence in patients. Capacitive pressure sensors are fabricated on a flexible polyimide-copper substrate using surface micromachining processes and alignment/assembly of the top and bottom portions of the sensor strip. The developed sensor strip is experimentally evaluated in an in vitro test rig using a pressure chamber. The sensor strip is shown to have adequate sensitivity and repeatability. While the calibration factors for the sensors on the strip vary from one sensor to another, even the least sensitive sensor has a resolution better than 0.1 psi.
引用
收藏
页码:7610 / 7613
页数:4
相关论文
共 50 条
  • [11] CATHETER WITHDRAWING APPARATUS FOR CLINICAL URETHRAL PRESSURE PROFILE STUDIES
    LOWE, PJ
    SAUNDERS, GAB
    DOWNIE, JW
    AWAD, SA
    JOURNAL OF UROLOGY, 1976, 116 (05): : 626 - 628
  • [12] Distributed optical fiber pressure sensors
    Schenato, Luca
    Galtarossa, Andrea
    Pasuto, Alessandro
    Palmieri, Luca
    OPTICAL FIBER TECHNOLOGY, 2020, 58
  • [13] High definition urethral pressure profilometry: Evaluating a novel microtip catheter
    Kluender, Mario
    Amend, Bastian
    Vaegler, Martin
    Kelp, Alexandra
    Feuer, Ronny
    Sievert, Karl-Dietrich
    Stenzl, Arnulf
    Sawodny, Oliver
    Ederer, Michael
    NEUROUROLOGY AND URODYNAMICS, 2016, 35 (08) : 888 - 894
  • [14] RELATIVE URETHRAL LEAKAGE PRESSURE VERSUS MAXIMUM URETHRAL CLOSURE PRESSURE - THE RELIABILITY OF THE MEASUREMENT OF URETHRAL COMPETENCE WITH THE NEW TUBE-FOIL SLEEVE CATHETER IN PATIENTS
    VANVENROOIJ, GEPM
    BLOK, C
    VANRIEL, MPJM
    COOLSAET, BLRA
    JOURNAL OF UROLOGY, 1985, 134 (03): : 592 - 595
  • [15] INDWELLING URETHRAL CATHETER
    MAROSSZEKY, JE
    MEDICAL JOURNAL OF AUSTRALIA, 1973, 2 (11) : 565 - 565
  • [16] Knotting of a urethral catheter
    Minami, Yousuke
    Tanaka, Hideaki
    Masumoto, Kouji
    PEDIATRICS INTERNATIONAL, 2017, 59 (08) : 932 - 933
  • [17] A NEW URETHRAL CATHETER
    BROCKLEHURST, JC
    HICKEY, DS
    DAVIES, I
    KENNEDY, AP
    MORRIS, JA
    BMJ-BRITISH MEDICAL JOURNAL, 1988, 296 (6638): : 1691 - 1693
  • [18] A PLEA FOR THE URETHRAL CATHETER
    QUENU, L
    NOUVELLE PRESSE MEDICALE, 1981, 10 (39): : 3250 - 3250
  • [19] RETENTION URETHRAL CATHETER
    EBERHART, CA
    JOURNAL OF UROLOGY, 1946, 55 (06): : 682 - 683
  • [20] URETHRAL CATHETER STRICTURES
    BURKITT, DS
    BARWELL, NJ
    WILSON, AGM
    LANCET, 1986, 1 (8482): : 688 - 688