Equivalence testing and equivalence limits of metered-dose inhalers and dry powder inhalers measured by in vitro impaction

被引:4
|
作者
Weda, M
Geuns, ERM
Vermeer, RCR
Buiten, NRA
Hendriks-de Jong, K
Bult, A
Zanen, P
Barends, DM
机构
[1] Natl Inst Publ Hlth & Environm, Lab Qual Control Med, NL-3720 BA Bilthoven, Netherlands
[2] Univ Utrecht, Fac Pharm, Utrecht, Netherlands
[3] Hart Lung Ctr Utrecht, Utrecht, Netherlands
关键词
equivalence; dry powder inhalers; metered-dose inhalers; impaction; twin impinger;
D O I
10.1016/S0939-6411(00)00064-3
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
In this study, criteria for the acceptability of comparative in vitro equivalence testing are proposed. Furthermore, the following equivalence limits for in vitro impaction methods are postulated: the 90% confidence interval (CI) of the in vitro deposition ratio of the test product and the reference product should lie within 0.80-1.20. The aim of this study was to challenge these limits by applying them to in vitro impaction results of several groups of pressurized metered-dose inhalers and dry powder inhalers containing salbutamol and beclomethasone dipropionate. The deposition results were obtained with the Twin Impinger. All products had a marketing authorization in the Netherlands and were considered therapeutically equivalent within each group. The postulated equivalence limits/group were challenged by fictitiously assigning a preparation as a test product or reference product and calculating the 90% CI of the deposition ratio of the test and reference products. All possible combinations of products within a group were tested. The products were considered equivalent if the 90% CI of the quotient lay within 0.80-1.20. In most cases, the quotient of the test product and reference product remains within 0.80-1.20, but due to a high variability in the deposition results of several products, the 90% CI of the quotient sometimes falls outside the proposed equivalence limits. It is concluded that the equivalence limits postulated are rather conservative, with respect to accepting equivalence. The limits can therefore serve as a prudent predictor of equivalence within the acceptability criteria proposed, but have to be further validated. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:295 / 302
页数:8
相关论文
共 50 条
  • [41] Inability of physicians to use metered-dose inhalers
    Tsang, KWT
    Lam, WK
    Ip, M
    Kou, M
    Yam, L
    Lam, B
    Cheung, MT
    Lauder, IJ
    Kumana, CR
    JOURNAL OF ASTHMA, 1997, 34 (06) : 493 - 498
  • [42] IN-VITRO TESTING OF DRY POWDER INHALERS
    VANOORT, M
    AEROSOL SCIENCE AND TECHNOLOGY, 1995, 22 (04) : 364 - 373
  • [43] Evaluation of Impaction Force of Nasal Sprays and Metered-Dose Inhalers Using the Texture Analyser
    Guo, Changning
    Ye, Wei
    Kauffman, John
    Doub, William H.
    JOURNAL OF PHARMACEUTICAL SCIENCES, 2009, 98 (08) : 2799 - 2806
  • [44] Pharmacodynamic equivalence study of CFC-free and CFC-containing procaterol hydrochloride metered-dose inhalers
    Kawai, M
    Nakashima, M
    Takaori, S
    Nakamura, Y
    Akamatsu, K
    Nakashima, M
    Miyamoto, T
    METHODS AND FINDINGS IN EXPERIMENTAL AND CLINICAL PHARMACOLOGY, 2005, 27 (08): : 555 - 558
  • [45] SINGLE-SHOT UNIT SPRAY TESTING OF METERED-DOSE INHALERS
    HARDY, JG
    HAIGH, RA
    BELL, JH
    JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1994, 7 (02): : 209 - 212
  • [46] In vitro evaluation of nonconventional accessory devices for pressurized metered-dose inhalers
    Sheth, Poonam
    Bertsch, Matthew D.
    Knapp, Carrie L.
    Myrdal, Paul B.
    ANNALS OF ALLERGY ASTHMA & IMMUNOLOGY, 2014, 113 (01) : 55 - 62
  • [47] Handling of inhaler devices in actual pulmonary practice: Metered-dose inhaler versus dry powder inhalers
    Khassawneh, Basheer Y.
    Al-Ali, Musa K.
    Alzoubi, Karem H.
    Batarseh, Morad Z.
    Al-Safi, Saafan A.
    Sharara, Abdelmonem M.
    Alnasr, Hatem M.
    RESPIRATORY CARE, 2008, 53 (03) : 324 - 328
  • [48] MEDICAL PERSONNELS KNOWLEDGE OF AND ABILITY TO USE INHALING DEVICES - METERED-DOSE INHALERS, SPACING CHAMBERS AND BREATH-ACTUATED DRY POWDER INHALERS
    HANANIA, NA
    WITTMAN, R
    KESTEN, S
    CHAPMAN, KR
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 147 (04): : A983 - A983
  • [49] DISPERSION AND SIZE DISTRIBUTION OF AMILORIDE BY METERED DOSE AND DRY POWDER INHALERS
    BOUNDY, MG
    LEITH, D
    KNOWLES, MR
    JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1990, 3 (04): : 233 - 241
  • [50] MEDICAL PERSONNELS KNOWLEDGE OF AND ABILITY TO USE INHALING DEVICES - METERED-DOSE INHALERS, SPACING CHAMBERS, AND BREATH-ACTUATED DRY POWDER INHALERS
    HANANIA, NA
    WITTMAN, R
    KESTEN, S
    CHAPMAN, KR
    CHEST, 1994, 105 (01) : 111 - 116