Establishment of regional diagnostic reference levels for digital mammography in Western Province of Sri Lanka

被引:4
|
作者
Niroshani, H. S. [1 ,2 ,3 ]
Jeyasugiththan, Jeyasingam [1 ]
Senanayake, G. [4 ]
Negishi, Toru [3 ]
机构
[1] Univ Colombo, Dept Nucl Sci, Colombo, Sri Lanka
[2] Gen Sir John Kotelawala Def Univ, Dept Radiog & Radiotherapy, Werahera, Sri Lanka
[3] Tokyo Metropolitan Univ, Grad Sch Human Hlth Sci, Dept Radiol Sci, Tokyo, Japan
[4] Gen Sir John Kotelawala Def Univ, Fac Med, Dept Clin Sci, Werahera, Sri Lanka
关键词
mammography; mean glandular dose (MGD); BREAST SCREENING-PROGRAM; DOSE SURVEY; MOLYBDENUM; RHODIUM;
D O I
10.1088/1361-6498/abd075
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The radiation dose to the breasts should be kept to a minimum as breast tissues are highly sensitive to radiation. In mammography, the mean glandular dose (MGD) is used to specify the patient dose. In this study, data on the MGD during diagnostic mammographic examinations was collected using the database from six digital mammography facilities available in the Western Province in Sri Lanka. Examinations involving breast pathology, breast implants, or compressed breast thicknesses (CBT) outside the range of 20-110 mm were excluded in this study. The mean MGD per breast was 3.50 mGy, with a mean CBT of 57 mm. The mean MGD per facility varies from 1.58 to 2.27 mGy, with overall 75th and 95th percentiles of 2.15 and 2.82 mGy, respectively. The 75th and 95th percentile MGD per image, for the average CBT of 57 +/- 12 mm, were 2.00 and 2.65 mGy respectively. The 75th percentile value of the MGD is suggested for the Western Province and it depends on the specific CBT.
引用
收藏
页码:79 / 96
页数:18
相关论文
共 50 条
  • [1] Mammography diagnostic reference levels in Western Australia
    Storm, Cameron
    Harvey, Alicia
    Djukelic, Mario
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (03) : 1125 - 1129
  • [2] Mammography diagnostic reference levels in Western Australia
    Cameron Storm
    Alicia Harvey
    Mario Djukelic
    [J]. Physical and Engineering Sciences in Medicine, 2020, 43 : 1125 - 1129
  • [3] Establishment of national diagnostic reference levels for full field digital mammography and digital breast tomosynthesis in Lebanon
    Nassar, J.
    Rizk, C.
    Farah, J.
    Fares, G.
    [J]. RADIOPROTECTION, 2023, 58 (04) : 289 - 298
  • [4] Establishment of institutional diagnostic reference level for coronary angiography procedures in Sri Lanka
    Ramanathan, Vijitha
    Senarathna, Hasalanka S.
    Gunaratna, Hasun U.
    Bandara, Praneeth M.
    Horadigala, Chathura J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2023, 199 (19) : 2311 - 2317
  • [5] Establishment of national diagnostic reference levels for computed tomography procedures in Sri Lanka: first nationwide dose survey
    Amalaraj, T.
    Satharasinghe, Duminda
    Pallewatte, Aruna
    Jeyasugiththan, Jeyasingam
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2022, 42 (02)
  • [6] Diagnostic reference levels in digital mammography: a systematic review
    Suleiman, Moayyad E.
    Brennan, Patrick C.
    McEntee, Mark F.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 167 (04) : 608 - 619
  • [7] Diagnostic reference levels in digital mammography: a systematic review
    Liu, Qiumei
    Suleiman, Moayyad E.
    McEntee, Mark F.
    Soh, BaoLin P.
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2022, 42 (01)
  • [8] National diagnostic reference levels for digital diagnostic and screening mammography in Uganda
    Odongo, Denish
    Musisi, Alen
    Okello, Richard Omara
    Bongomin, Felix
    Erem, Geoffrey
    [J]. PLOS ONE, 2024, 19 (08):
  • [9] Establishment of CT Diagnostic Reference Levels in Province Nova Scotia
    Tonkopi, E.
    Abdolell, M.
    Duffy, S.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3249 - 3249
  • [10] Diagnostic reference levels for digital mammography in New South Wales
    Suleiman, Moayyad E.
    McEntee, Mark F.
    Cartwright, Lucy
    Diffey, Jennifer
    Brennan, Patrick C.
    [J]. JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2017, 61 (01) : 48 - 57