A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

被引:10
|
作者
Guan, Huaiqun [1 ]
Hammoud, Rabih [2 ]
Yin, Fang-Fang [3 ]
机构
[1] Good Samaritan Hlth Syst, Dept Radiat Oncol, Kearney, NE 68848 USA
[2] Henry Ford River Ctr Oncol, Trenton, MI USA
[3] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC USA
来源
关键词
OBI; CBCT; positioning QA; 2D/2D (kV/MV); 3D/3D (planCT/CBCT); QUALITY-ASSURANCE;
D O I
10.1120/jacmp.v10i4.2954
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on-board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pre-specified accuracy); (2) that the positioning difference detected by 2D/2D (kV/MV) and 3D/3D(planCT/CBCT) matching can be reliably transferred to couch motion. A cube phantom with a 2 mm metal ball (bb) at the center was used. The bb was used to define the isocenter. Two additional bbs were placed on two phantom surfaces in order to define a spatial location of 1.5 cm anterior, 1.5 cm inferior, and 1.5 cm right from the isocenter. An axial scan of the phantom was acquired from a multislice CT simulator. The phantom was set at the linac's isocenter (lasers); either AP MV/R Lat kV images or CBCT images were taken for 2D/2D or 3D/3D matching, respectively. For 2D/2D, the accuracy of each device's isocenter was obtained by checking the distance between the central bb and the digital graticule. Then the central bb in orthogonal DRRs was manually moved to overlay to the off-axis bbs in kV/MV images. For 3D/3D, CBCT was first matched to planCT to check the isocenter difference between the two CTs. Manual shifts were then made by moving CBCT such that the point defined by the two off-axis bbs overlay to the central bb in planCT. (PlanCT can not be moved in the current version of OBI1.4.) The manual shifts were then applied to remotely move the couch. The room laser was used to check the accuracy of the couch movement. For Trilogy (or Ix-21) linacs, the coincidence of imager and linac's isocenter was better than 1 mm (or 1.5 mm). The couch shift accuracy was better than 2 mm.
引用
收藏
页码:273 / 280
页数:8
相关论文
共 50 条
  • [1] A Robust 2D/3D Image Registration System for Pediatric Patient Setup in External Beam Radiotherapy
    Wu, J.
    Su, Z.
    Li, Z.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3961 - 3961
  • [2] 2D/3D Registration Using KV-MV Image Pairs for Higher Accuracy Image Guided Radiotherapy
    Furtado, H.
    Figl, M.
    Stock, M.
    Georg, D.
    Birkfellner, W.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3673 - 3673
  • [3] 2D or 3D?
    Mills, R
    [J]. COMPUTER-AIDED ENGINEERING, 1996, 15 (08): : 4 - 4
  • [4] 2D or not 2D That is the Question, but 3D is the, answer
    Cronin, Paul
    [J]. ACADEMIC RADIOLOGY, 2007, 14 (07) : 769 - 771
  • [5] 3D IMAGE RECONSTRUCTION FROM 2D CT SLICES
    Kamencay, Patrik
    Zachariasova, Martina
    Hudec, Robert
    Benco, Miroslav
    Radil, Roman
    [J]. 2014 3DTV-CONFERENCE: THE TRUE VISION - CAPTURE, TRANSMISSION AND DISPLAY OF 3D VIDEO (3DTV-CON), 2014,
  • [6] 3D and 2D/3D holograms model
    A. A. Boriskevich
    V. K. Erohovets
    V. V. Tkachenko
    [J]. Optical Memory and Neural Networks, 2012, 21 (4) : 242 - 248
  • [7] Elastic 3D–2D Image Registration
    Paul Striewski
    Benedikt Wirth
    [J]. Journal of Mathematical Imaging and Vision, 2022, 64 : 443 - 462
  • [8] 2D/3D image registration on the GPU
    Kubias A.
    Deinzer F.
    Feldmann T.
    Paulus D.
    Schreiber B.
    Brunner Th.
    [J]. Pattern Recognition and Image Analysis, 2008, 18 (3) : 381 - 389
  • [9] 2D to 3D Medical Image Colorization
    Mathur, Aradhya Neeraj
    Khattar, Apoorv
    Sharma, Ojaswa
    [J]. 2021 IEEE WINTER CONFERENCE ON APPLICATIONS OF COMPUTER VISION WACV 2021, 2021, : 2846 - 2855
  • [10] 2D/3D image (facial) comparison using camera matching
    Goos, Mirelle I. M.
    Alberink, Ivo B.
    Ruifrok, Arnout C. C.
    [J]. FORENSIC SCIENCE INTERNATIONAL, 2006, 163 (1-2) : 10 - 17