Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations

被引:154
|
作者
Jarry, Genevieve [1 ]
Graham, Sean A.
Moseley, Douglas J.
Jaffray, David J.
Siewerdsen, Jeffrey H.
Verhaegen, Frank
机构
[1] McGill Univ, Montreal, PQ H3G 1A4, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[3] Princess Margaret Hosp, Univ Hlth Network, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[4] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
关键词
cone-beam CT; scatter; Monte Carlo; scatter correction;
D O I
10.1118/1.2358324
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Kilovoltage (kV) cone beam computed tomography (CBCT) images suffer from a substantial scatter contribution. In this study, Monte Carlo (MC) simulations are used to evaluate the scattered radiation present in projection images. These predicted scatter distributions are also used as a scatter correction technique. Images were acquired using a kV CBCT bench top system. The EGSnrc MC code was used to model the flat panel imager, the phantoms, and the x-ray source. The x-ray source model was validated using first and second half-value layers (HVL) and profile measurements. The HVLs and the profile were found to agree within 3% and 6%, respectively. MC simulated and measured projection images for a cylindrical water phantom and for an anthropomorphic head phantom agreed within 8% and 10%. A modified version of the DOSXYZnrc MC code was used to score phase space files with identified scattered and primary particles behind the phantoms. The cone angle, the source-to-detector distance, the phantom geometry, and the energy were varied to determine their effect on the scattered radiation distribution. A scatter correction technique was developed in which the MC predicted scatter distribution is subtracted from the projections prior to reconstruction. Preliminary testing of the procedure was done with an anthropomorphic head phantom and a contrast phantom. Contrast and profile measurements were obtained for the scatter corrected and noncorrected images. An improvement of 3% for contrast between solid water and a liver insert and 11% between solid water and a Teflon insert were obtained and a significant reduction in cupping and streaking artifacts was observed. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:4320 / 4329
页数:10
相关论文
共 50 条
  • [21] Monte Carlo simulations of scattered power from irradiated optical elements
    Secco, Eleonora
    del Rio, Manuel Sanchez
    ADVANCES IN COMPUTATIONAL METHODS FOR X-RAY OPTICS II, 2011, 8141
  • [22] Estimation of scattered photons in gamma ray transmission CT using Monte Carlo simulations.
    Ogawa, K
    Kawamura, Y
    Kubo, A
    Ichihara, T
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) : 1225 - 1230
  • [24] Efficient Monte Carlo calculations of CBCT scatter using egs_cbct
    Mainegra-Hing, E.
    Thing, R. S.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S265 - S266
  • [25] Monte Carlo simulations of edge artefacts in MULSAM images
    El-Gomati, MM
    Assa'd, AMD
    MIKROCHIMICA ACTA, 1998, : 333 - 339
  • [26] Scatter correction for kilovoltage cone-beam computed tomography (CBCT) images using Monte Carlo simulations - art. no. 614254
    Jarry, G.
    Graham, S. A.
    Jaffray, D. A.
    Moseley, D. J.
    Verhaegen, F.
    Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : 14254 - 14254
  • [27] Characterization and Monte Carlo simulations for a CLYC detector
    Borella, Alessandro
    Boogers, Eric
    Rossa, Riccardo
    Schillebeeckx, Peter
    PROCEEDINGS OF THE 2017 INTERNATIONAL CONFERENCE ON APPLICATIONS OF NUCLEAR TECHNIQUES (CRETE17), 2018, 48
  • [28] Hierarchical characterization of aggregates for Monte Carlo simulations
    Briesen, Heiko
    AICHE JOURNAL, 2006, 52 (07) : 2436 - 2446
  • [29] CBCT dose distribution calculation with GATE Monte Carlo simulations for lymphoma patients
    Riveira-Martin, M.
    Sanchez-Diaz, I.
    Dorado Dorado, R.
    Salvador Gomez, F. J.
    Andrade Alvarez, B.
    Salgado Fernandez, M.
    Munoz Garzon, V.
    Ochagavia Galilea, V.
    Nieto Regueira, I.
    Myronakis, M.
    Lopez Medina, A.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1542 - S1544
  • [30] A method of CBCT imaging scatter correction based on the EGSnrc monte carlo simulations
    Meng, H. P.
    Dong, H. J.
    Wang, K. Q.
    Feng, Y. M.
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2018, 123 : 21 - 22