In this paper, we evaluate the performance of recently released avalanche photodiode arrays from Advanced Photonix, Inc., for scintillator matrices readout. The large-area avalanche photodiode (LAAPD) quadrant device is a monolithic 2 x 2 pixel structure with an active area of 5.3 mm(2) per pixel. The device allows stable operation with high gains up to 200 and detection efficiency of 73similar to+/- 10% for 420-nm photons. It is furthermore characterized by a pixel linear fill factor of 90% and low noise equal to 16 electrons equivalent noise charge measured at room temperature. We have measured pixel-to-pixel gain nommiformity smaller than 1.5% and interpixel crosstalk of 1.7% in pulse-mode operation with a device gain of 50. An energy resolution of 12.3 +/- 0.5% was achieved for the 511-keV photopeak from a Na-22 source placed on top of a 2 x 2 x 10 mm(3) LSO crystal coupled to one pixel of a quadrant LAAPD. An array of LSO crystals was later coupled to the device allowing individual crystal identification. A coincidence timing resolution of 1.9 +/- 0.1 ns full-width at half-maximum has been obtained for quadrant pixel with LSO scintillator for the 511-keV peak from a Na-22 source. Finally, we compared the characteristics and readout performance of the LAAPD array with published results of some commercially available avalanche photodiode arrays.
引用
收藏
页码:1693 / 1698
页数:6
相关论文
共 50 条
  • [31] Performance of borated scintillator screens for high-resolution neutron imaging
    Burkhard Schillinger
    William Chuirazzi
    Aaron Craft
    Steven Cool
    Alessandro Tengattini
    Journal of Radioanalytical and Nuclear Chemistry, 2022, 331 : 5287 - 5295
  • [32] Modular 64x64 CdZnTe arrays with multiplexer readout for high-resolution nuclear medicine imaging
    Woolfenden, JM
    Barber, HB
    Barrett, HH
    Dereniak, EL
    Eskin, JD
    Marks, DG
    Matherson, KJ
    Young, ET
    Augustine, FL
    INFRARED APPLICATIONS OF SEMICONDUCTORS II, 1998, 484 : 267 - 271
  • [33] Modular 64x64 CdZnTe arrays with multiplexer readout for high-resolution nuclear medicine imaging
    Woolfenden, JM
    Barber, HB
    Barrett, HH
    Dereniak, EL
    Eskin, JD
    Marks, DG
    Matherson, KJ
    Young, ET
    Augustine, FL
    SEMICONDUCTORS FOR ROOM-TEMPERATURE RADIATION DETECTOR APPLICATIONS II, 1997, 487 : 239 - 243
  • [34] High-resolution beam forming for ultrasonic arrays
    Webb, P
    Wykes, C
    IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 1996, 12 (01): : 138 - 146
  • [35] Evaluation of a modular all-in-one high-resolution PET detector and readout electronics setup
    Schmidt, F. P.
    Kraemer, J. C.
    Parl, C.
    Schmand, M.
    Pichler, B. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (11):
  • [36] OPTICAL READOUT OF A HIGH-RESOLUTION PARALLEL PLATE AVALANCHE CHAMBER
    POTTER, DM
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 228 (01): : 56 - 61
  • [37] A HIGH-RESOLUTION DELAY-LINE READOUT FOR MICROCHANNEL PLATES
    SIEGMUND, OHW
    LAMPTON, ML
    RAFFANTI, R
    EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY AND ATOMIC PHYSICS, 1989, 1159 : 476 - 485
  • [38] HIGH-RESOLUTION TWO-DIMENSIONAL READOUT OF DRIFT CHAMBERS
    MAESHIMA, K
    MALCHOW, RL
    SPARKS, KB
    THIEL, TG
    WILLIAMS, MCS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 252 (2-3): : 612 - 615
  • [39] A new photodiode readout scheme for high resolution scintillation crystal arrays
    Levin, CS
    Hoffman, EJ
    1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1223 - 1227
  • [40] Evaluation of High-Resolution Mass Spectrometry for the Quantitative Analysis of Mycotoxins in Complex Feed Matrices
    Jensen, Tolke
    de Boevre, Marthe
    Preusske, Nils
    de Saeger, Sarah
    Birr, Tim
    Verreet, Joseph-Alexander
    Soennichsen, Frank D.
    TOXINS, 2019, 11 (09)