Diagnostic performance of a flat-panel detector at low tube voltage in chest radiography -: A phantom study

被引:12
|
作者
Bernhardt, TM
Rapp-Bernhardt, U
Lenzen, H
Röhl, FW
Diederich, S
Papke, K
Ludwig, K
Heindel, W
机构
[1] Univ Munster, Dept Clin Radiol, D-48129 Munster, Germany
[2] Otto Von Guericke Univ, Dept Biomed & Med Informat, Magdeburg, Germany
[3] Klinikum Duisburg, Dept Radiol & Neuroradiol, Duisburg, Germany
关键词
flat-panel detector; digital radiography; image quality; experimental investigations;
D O I
10.1097/01.rli.0000110566.23154.81
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To evaluate a large area, cesium iodide amorphous silicon flat-panel detector (CsI/a-Si) at 3 tube voltages to detect simulated interstitial lung disease, nodules, and catheters. Methods: Simulated interstitial lung disease, nodules, and catheters were superimposed over a chest phantom. Images were generated at 125 kVp, 90 kVp, and 70 kVp at the same surface dose and reduced effective dose equivalent for 90 kVp and 70 kVp and printed on hard copies. Fifty-four thousand observations were analyzed by receiver operating characteristic (ROC). Results: Detectability of linear, miliary, reticular pattern, and nodules over lucent lung as well as of catheters and nodules over obscured chest areas increased at 90 and/or 70 kVp with higher Az values; however, only it was statistically significant for reticular pattern at 70 kVp and nodules at 90 kVp compared with 125 kVp (P <0.05). The detection of ground-glass pattern was worse at lower kVp (P >0.05). Conclusion: For most simulated patterns, differences in diagnostic performance at 70 kVp/90 kVp and 125 kVp were not significant, except for reticular pattern and nodules over lucent lung.
引用
收藏
页码:97 / 103
页数:7
相关论文
共 50 条
  • [1] Diagnostic performance of a flat panel detector at low tube voltage in chest radiography
    Rapp-Bernhardt, U
    Bernhardt, TM
    Lenzen, H
    Diederich, S
    Papke, K
    Heindel, WL
    [J]. RADIOLOGY, 2002, 225 : 382 - 383
  • [2] Dependence of low contrast detail on exposure dose and tube voltage in digital flat-panel detector radiography-a pre-clinical phantom study
    Notohamiprodjo, S.
    Verstreepen, L.
    Wanninger, F.
    Hoberg, B.
    Roeper, K. M.
    Mueck, F. G.
    Treitl, K. M.
    Maxien, D.
    Wirth, S.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (02):
  • [3] Flat-panel-detector chest radiography: Effect of tube voltage on image quality
    Uffmann, M
    Neitzel, U
    Prokop, M
    Kabalan, N
    Weber, M
    Herold, CJ
    Schaefer-Prokop, C
    [J]. RADIOLOGY, 2005, 235 (02) : 642 - 650
  • [4] Dynamic chest radiography with a flat-panel detector (FPD): Ventilation-perfusion study
    Tanaka, R.
    Sanada, S.
    Fujimura, M.
    Yasui, M.
    Tsuji, S.
    Hayashi, N.
    Okamoto, H.
    Nanbu, Y.
    Matsui, O.
    [J]. MEDICAL IMAGING 2011: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2011, 7965
  • [5] A-Se Flat-Panel (FP) Detector Based Scan Equalization Digital Radiography (SEDR) System: A Chest Phantom Study
    Liu, X.
    Lai, C.
    Chen, L.
    Han, T.
    Shen, Y.
    Zhong, Y.
    Zhang, H.
    Yi, Y.
    Ge, S.
    Wang, T.
    Shaw, C.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [6] Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography
    Floyd, CE
    Warp, RJ
    Dobbins, JT
    Chotas, HG
    Baydush, AH
    Vargas-Voracek, R
    Ravin, CE
    [J]. RADIOLOGY, 2001, 218 (03) : 683 - 688
  • [7] Optimum radiographic technique for chest radiography using a flat-panel digital detector
    Samei, E
    Dobbins, JT
    Chotas, HG
    Baydush, AH
    Floyd, CE
    Ravin, CE
    [J]. RADIOLOGY, 2002, 225 : 642 - 643
  • [8] Digital radiography versus conventional radiography in chest imaging:: Diagnostic performance of a large-area silicon flat-panel detector in a clinical CT-controlled study
    Garmer, M
    Hennigs, SP
    Jäger, HJ
    Schrick, F
    van de Loo, T
    Jacobs, A
    Hanusch, A
    Christmann, A
    Mathias, K
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (01) : 75 - 80
  • [9] Radiation dose reduction in chest radiography using a flat-panel amorphous silicon detector
    Hosch, WP
    Fink, C
    Radeleff, B
    Kampschulte, A
    Kauffmann, GW
    Hansmann, J
    [J]. CLINICAL RADIOLOGY, 2002, 57 (10) : 902 - 907
  • [10] Breathing chest radiography using a dynamic flat-panel detector combined with computer analysis
    Tanaka, R
    Sanada, S
    Suzuki, M
    Kobayashi, T
    Matsui, T
    Inoue, H
    Yoshihisa, N
    [J]. MEDICAL PHYSICS, 2004, 31 (08) : 2254 - 2262