A novel thin-film temperature and heat-flux microsensor for heat transfer measurements in microchannels

被引:15
|
作者
Hamadi, David [1 ]
Garnier, Bertrand [1 ]
Willaime, Herve [2 ]
Monti, Fabrice [2 ]
Peerhossaini, Hassan [1 ]
机构
[1] Univ Nantes, Ecole Polytech, UMR CNRS 6607, Lab Thermocinet, F-44306 Nantes, France
[2] CNRS ESPCI, Lab Microfluid MEMS & Nanostruct, F-75005 Paris, France
关键词
FORCED-CONVECTION; FLOW; WATER; MICROTUBES;
D O I
10.1039/c2lc20919e
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Temperature and heat-flux measurement at the microscale for convective heat-transfer studies requires highly precise, minimally intrusive sensors. For this purpose, a new generic temperature and heat-flux sensor was designed, calibrated and tested. The sensor allows measurement of temperature and heat flux distributions along the direction of flow. It is composed of forty gold thermoresistances, 85 nm thick, deposited on both sides of a borosilicate substrate. Their sensitivities are about 37.8 mu VK-1, close to those of a K-type wire thermocouple. Using a thermoelectrical model, temperature biases due to the Joule effect were calculated using the current crossing each thermoresistance and the heat-transfer coefficient. Finally, heat-transfer measurements were performed with deionized water flowing in a straight PDMS microchannel for various Reynolds numbers. The Nusselt number was obtained for microchannels of 50 to 10 mu m span. The results were found to be in good agreement with classical Nu-Re macroscopic correlations.
引用
收藏
页码:652 / 658
页数:7
相关论文
共 50 条
  • [21] RECTANGULAR HEAT-SOURCE ON A SEMI-INFINITE SOLID - AN ANALYSIS FOR A THIN-FILM HEAT-FLUX GAUGE CALIBRATION
    KELTNER, NR
    BAINBRIDGE, BL
    BECK, JV
    JOURNAL OF HEAT TRANSFER-TRANSACTIONS OF THE ASME, 1988, 110 (01): : 42 - 48
  • [23] EFFECT OF SYSTEM PRESSURE ON FILM-BOILING HEAT-TRANSFER, MINIMUM HEAT-FLUX, AND MINIMUM TEMPERATURE
    SAKURAI, A
    SHIOTSU, M
    HATA, K
    NUCLEAR SCIENCE AND ENGINEERING, 1984, 88 (03) : 321 - 330
  • [24] DEVELOPMENT OF THIN-FILM THERMOMETER TO MEASURE TRANSIENT PLASMA HEAT-FLUX IN DIVERTER ASSEMBLY (DIVA)
    SEKI, M
    KAWAMURA, H
    UEDA, N
    MAEDA, H
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 1977, 14 (07) : 534 - 536
  • [25] ERRORS IN HEAT-FLUX MEASUREMENTS DUE TO THE THERMAL-RESISTANCE OF HEAT-FLUX DISKS
    DUCHARME, MB
    FRIM, J
    TIKUISIS, P
    JOURNAL OF APPLIED PHYSIOLOGY, 1990, 69 (02) : 776 - 784
  • [26] Augmentation of nucleate boiling heat transfer and critical heat flux using nanoparticle thin-film coatings
    Forrest, Eric
    Williamson, Erik
    Buongiorno, Jacopo
    Hu, Lin-Wen
    Rubner, Michael
    Cohen, Robert
    INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2010, 53 (1-3) : 58 - 67
  • [27] HEAT-TRANSFER AT COORDINATED HEAT-FLUX OSCILLATIONS
    SOZIN, YA
    IZVESTIYA VYSSHIKH UCHEBNYKH ZAVEDENII AVIATSIONAYA TEKHNIKA, 1981, (01): : 100 - 102
  • [28] MICROSENSORS FOR HIGH HEAT-FLUX MEASUREMENTS
    HAGER, JM
    LANGLEY, LW
    ONISHI, S
    DILLER, TE
    JOURNAL OF THERMOPHYSICS AND HEAT TRANSFER, 1993, 7 (03) : 531 - 534
  • [29] DYNAMIC MEASUREMENTS BY HEAT-FLUX CALORIMETRY
    BODDINGTON, T
    CHIA, HA
    HALFORDMAW, P
    FENG, HT
    LAYE, PG
    THERMOCHIMICA ACTA, 1992, 195 : 365 - 372
  • [30] TEMPERATURE AND HEAT-FLUX MEASUREMENTS IN NOISELESS FILM BOILING IN SUPERFLUID-HELIUM WITH A FLAT HEATER
    MAZA, J
    JEBALI, F
    FRANCOIS, MX
    VIDAL, F
    CRYOGENICS, 1989, 29 (03) : 200 - 202