Portal dose image verification:: formalism and application of the collapsed cone superposition method

被引:9
|
作者
Dahlgren, CV
Ahnesjö, A
Montelius, A
Rikner, G
机构
[1] MDS Nordion AB, SE-75147 Uppsala, Sweden
[2] Univ Hosp, Akad Sjukhuset, Dept Oncol Radiol & Clin Immunol, Dept Hosp Phys, SE-75185 Uppsala, Sweden
[3] Univ Hosp, Akad Sjukhuset, Dept Biomed Radiat Sci, SE-75185 Uppsala, Sweden
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2002年 / 47卷 / 24期
关键词
D O I
10.1088/0031-9155/47/24/305
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A formalism tailored for portal dose image verification is proposed to facilitate the comparison of calculated and measured portal dose distributions. Each portal image is converted into a dose proportional image and normalized to the reference beam calibration dose per monitor unit. The calculated or measured dose to a detector phantom is accordingly normalized so as to enable direct comparison. The collapsed cone kernel superposition method is adapted and evaluated for calculation of portal dose distributions in a water-equivalent detector phantom through comparisons with Monte Carlo calculations and with measurements. The deviation compared with Monte Carlo calculations for 6 and 15 MV was between +0.9% (the 0.9 quantile) and -2.1% (the 0.1 quantile) for a range of investigated geometries. Collapsed cone calculations compared with measurements for clinical fields agreed within [-1.9%, +2.4%] for 15 MV and [-0.9%, +3.2%] for 6 MV for the 0.1 and 0.9 quantiles, respectively. Hence, the absolute portal dose to a detector phantom could be calculated and verified well within the present accuracy requirements for clinical dose calculations.
引用
收藏
页码:4371 / 4387
页数:17
相关论文
共 50 条
  • [1] Portal dose image verification:: the collapsed cone superposition method applied with different electronic portal imaging devices
    Dahlgren, CV
    Eilertsen, K
    Jorgensen, TD
    Ahnesjö, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (02): : 335 - 349
  • [2] Use of collapsed cone convolution for an initial investigation of portal dose image prediction
    Dahlgren, CV
    Ahnesjö, A
    [J]. USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 224 - 226
  • [3] Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method
    Butson, MJ
    Elferink, R
    Cheung, T
    Yu, PKN
    Stokes, M
    Quach, KY
    Metcalfe, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11): : N143 - N149
  • [4] The collapsed cone superposition algorithm applied to scatter dose calculations in brachytherapy
    Carlsson, ÅK
    Ahnesjö, A
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2320 - 2332
  • [5] Comparison of Collapsed Cone Convolution Superposition and Monte Carlo Dose Calculations for Spine SBRT Dose
    Parenica, H.
    Rembish, J.
    Saenz, D.
    Rasmussen, K.
    Kirby, N.
    Shi, Z.
    Li, Y.
    Fakhreddine, M.
    Crownover, R.
    Ha, C.
    Papanikolaou, N.
    Stathakis, S.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E163 - E164
  • [6] A portal dosimetry dose prediction method based on collapsed cone algorithm using the clinical beam model
    Martinez Ortega, J.
    Gomez Gonzalez, N.
    Castro Tejero, P.
    Pinto Monedero, M.
    Tolani, N. B.
    Nunez Martin, L.
    Sanchez Montero, R.
    [J]. MEDICAL PHYSICS, 2017, 44 (01) : 333 - 341
  • [7] A collapsed-cone convolution/superposition dose computation algorithm for sliding window beams
    Moreau, M
    Gibbons, J
    Bzdusek, K
    McNutt, T
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2161 - 2161
  • [8] Accelerating Collapsed Cone Convolution/Superposition Dose Calculation On GPU Using Spatial Decomposition
    Xiao, K.
    Zhou, B.
    Hu, X. S.
    Chen, D. Z.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [9] Accounting for heterogeneities and shields in brachytherapy dose calculation bg use of collapsed cone superposition
    Carlsson, ÅK
    Ahnesjö, A
    [J]. USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 474 - 476
  • [10] Calculation of portal dose using the convolution/superposition method
    McNutt, TR
    Mackie, TR
    Reckwerdt, P
    Papanikolaou, N
    Paliwal, BR
    [J]. MEDICAL PHYSICS, 1996, 23 (04) : 527 - 535