A multicenter comparison study of the Humphrey Field Analyzer I and the Humphrey Field Analyzer II

被引:8
|
作者
Johnson, CA
Cioffi, GA
Drance, SM
Gaasterland, D
Mills, RP
Ashburn, F
Hnik, P
VanCoevorden, RE
机构
[1] LEGACY HOSP SYST, DEVERS EYE INST, PORTLAND, OR USA
[2] UNIV BRITISH COLUMBIA, DEPT OPHTHALMOL, VANCOUVER, BC V5Z 1M9, CANADA
[3] UNIV OPHTHAL CONSULTANTS WASHINGTON DC, WASHINGTON, DC USA
[4] UNIV WASHINGTON, DEPT OPHTHALMOL, SEATTLE, WA 98195 USA
[5] EYE ASSOCIATES WASHINGTON DC, WASHINGTON, DC USA
关键词
D O I
10.1016/S0161-6420(97)30008-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The purpose of the study is to determine the comparability of the 30-2 full-threshold program in the original Humphrey Field Analyzer (HFA) I to the same lest procedure in the new Humphrey Field Analyzer II, Methods: At each of five clinical centers, one eye of patients with ocular hypertension and normal visual fields, patients with early glaucomatous visual field loss, and patients with more advanced visual field loss were tested with the two instruments plus a retest on a separate HFA I. All participants had undergone at least one prior visual field examination. To minimize the influence of any residual learning or fatigue effects, the order of testing for the three visual field examinations was counterbalanced across subjects. A total of 250 patients were tested (81 patients with ocular hypertension, 81 patients with early glaucomatous visual field loss, and 88 patients with more advanced glaucomatous visual field loss), Results: No statistically significant differences were observed between thresholds, visual field indices, or reliability indices obtained with the HFA I and the HFA II. The small differences between the two instruments were equivalent to the variation observed for test-retest measures using only the HFA I, These results were consistent across the range of visual field characteristics shown by the ocular hypertensive, early glaucoma, and moderate glaucoma patient groups. Conclusions: The authors' results indicate that there are no differences in the visual field results obtained with the HFA I and the HFA [I. These findings suggest that with careful attention to test protocols, the HFA I and HFA II may be used interchangeably to observe patients, even within the context of multicenter clinical trials.
引用
收藏
页码:1910 / 1917
页数:8
相关论文
共 50 条
  • [31] Comparative studies in Humphrey field analyzer, FDT screener and Humphrey Matrix for the detection of normal tension glaucoma
    Kimura, T
    Kawabata, K
    Ohyama, A
    Hasegawa, S
    Fujimaki, T
    Murakami, A
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [32] Evaluating equivalence between the new Humphrey Field Analyzer (HFA3) and HFA II-i
    Callan, Thomas
    Sha, Patricia
    Luu, Jennifer Yen
    Yu, Sophia
    Chang, Aileen
    Severin, Todd
    Duh, Y-Jing
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [33] QUANTITATIVE LAYER-BY-LAYER PERIMETRY ADAPTED TO THE HUMPHREY FIELD ANALYZER
    BEK, T
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1992, 33 (04) : 971 - 971
  • [34] EVALUATION OF FASTPAC - A NEW STRATEGY FOR THRESHOLD ESTIMATION WITH THE HUMPHREY FIELD ANALYZER
    FLANAGAN, JG
    MOSS, ID
    WILD, JM
    HUDSON, C
    PROKOPICH, L
    WHITAKER, D
    ONEILL, EC
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1993, 231 (08) : 465 - 469
  • [35] Comparing the Performance of Compass Perimetry With Humphrey Field Analyzer in Eyes With Glaucoma
    Rao, Harsha L.
    Raveendran, Sharanya
    James, Varsha
    Dasari, Srilakshmi
    Palakurthy, Meena
    Reddy, Hemanth B.
    Pradhan, Zia S.
    Rao, Dhanaraj A. S.
    Puttaiah, Narendra K.
    Devi, Sathi
    JOURNAL OF GLAUCOMA, 2017, 26 (03) : 292 - 297
  • [36] A CLINICAL COMPARISON OF VISUAL-FIELD TESTING WITH A NEW AUTOMATED PERIMETER, THE HUMPHREY FIELD ANALYZER, AND THE GOLDMANN PERIMETER
    BECK, RW
    BERGSTROM, TJ
    LICHTER, PR
    OPHTHALMOLOGY, 1985, 92 (01) : 77 - 82
  • [37] Comparison of Visual Field Measurement with Heidelberg Edge Perimeter and Humphrey Visual Field Analyzer in Patients with Ocular Hypertension
    Kaczorowski, Kamil
    Mulak, Malgorzata
    Szumny, Dorota
    Baranowska, Marta
    Jakubaszko-Jablonska, Joanna
    Misiuk-Hojlo, Marta
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2016, 25 (05): : 937 - 944
  • [38] Comparison of Humphrey Matrix 24-2 standard perimetry and Humphrey Field Analyzer 24-2 SITA standard perimetry
    Bartlett, JD
    Shaikh, A
    Semes, L
    Xie, A
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [39] Comparison among humphrey field analyzer and nidek MP-1 in retinitis pigmentosa patients
    Rigoni, Erika
    Carnevale, Carmela
    Domanico, Daniela
    Vingolo, Enzo M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [40] Comparison of the threshold value of SITA with that of conventional perimetry by Humphrey field analyzer in glaucoma patients.
    Iwase, A
    Kitazawa, Y
    Ohno, Y
    Ansari, I
    Chauhan, B
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S89 - S89