Validation of a SIMIND Monte Carlo modelled gamma camera for Iodine-123 and Iodine-131 imaging

被引:11
|
作者
Morphis, Michaella [1 ]
van Staden, Johan A. [1 ]
du Raan, Hanlie [1 ]
Ljungberg, Michael [2 ]
机构
[1] Univ Free State, Fac Hlth Sci, Dept Med Phys, Bloemfontein, South Africa
[2] Lund Univ, Med Radiat Phys, Lund, Sweden
基金
新加坡国家研究基金会; 英国医学研究理事会;
关键词
Validation; Monte Carlo simulations; SIMIND; I-123; I-131; Static planar; WB planar; SPECT; ACTIVITY QUANTIFICATION; RADIONUCLIDE THERAPY; PROCEDURE GUIDELINES; DOSIMETRY; SPECT; SIMULATION; RECONSTRUCTION; GATE; CT;
D O I
10.1016/j.heliyon.2021.e07196
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: Monte Carlo (MC) modelling techniques can assess the quantitative accuracy of both planar and SPECT Nuclear Medicine images. It is essential to validate the MC code's capabilities in modelling a specific clinical gamma camera, for radionuclides of interest, before its use as a clinical image simulator. This study aimed to determine if the SIMIND MC code accurately simulates emission images measured with a Siemens Symbia (TM) T16 SPECT/CT system for I-123 with a LEHR and a ME collimator and for I-131 with a HE collimator. Methods: The static and WB planar validation tests included extrinsic system energy pulse-height distributions (EPHDs), system sensitivity and system spatial resolution in air as well as a scatter medium. The SPECT validation test comprised the sensitivity from a simple geometry of a sphere in a cylindrical water-filled phantom. Results: The system EPHDs compared well, with differences between measured and simulated primary photopeak FWHM values not exceeding 4.6 keV. Measured and simulated planar system sensitivity values displayed percentage differences less than 6.9% and 6.3% for static and WB planar images, respectively. Measured and simulated planar system spatial resolution values in air showed percentage differences not exceeding 6.4% (FWHM) and 10.0% (FWTM), and 5.1% (FWHM) and 5.4% (FWTM) for static and WB planar images, respectively. For static planar system spatial resolution measured and simulated in a scatter medium, percentage differences of FWHM and FWTM values were less than 5.8% and 12.6%, respectively. The maximum percentage difference between the measured and simulated SPECT validation results was 3.6%. Conclusion: The measured and simulated validation results compared well for all isotope-collimator combinations and showed that the SIMIND MC code could be used to accurately simulate static and WB planar and SPECT projection images of the Siemens Symbia (TM) T16 SPECT/CT for both I-123 and I-131 with their respective collimators.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Evaluation of Iodine-123 and Iodine-131 SPECT activity quantification: a Monte Carlo study
    Morphis, Michaella
    van Staden, Johan A.
    du Raan, Hanlie
    Ljungberg, Michael
    [J]. EJNMMI PHYSICS, 2021, 8 (01)
  • [2] Evaluation of Iodine-123 and Iodine-131 SPECT activity quantification: a Monte Carlo study
    Michaella Morphis
    Johan A. van Staden
    Hanlie du Raan
    Michael Ljungberg
    [J]. EJNMMI Physics, 8
  • [3] Is iodine-123 as sensitive as iodine-131 for whole body imaging in differentiated thyroid cancer?
    Sarkar, SD
    Palestro, CJ
    [J]. RADIOLOGY, 2000, 217 : 454 - 454
  • [4] Monte Carlo modeling of penetration effect for iodine-131 pinhole imaging
    Wang, HL
    Jaszczak, RJ
    Coleman, RE
    [J]. 1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1558 - 1562
  • [5] Monte carlo modeling of penetration effect for iodine-131 pinhole imaging
    Wang, HL
    Jaszczak, RJ
    Coleman, RE
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (06) : 3272 - 3277
  • [6] Comparison of iodine-123 and iodine-131 as a scanning agent for the detection of metastatic thyroid cancer.
    Bautovich, GJ
    Towson, JE
    Eberl, S
    Turtle, J
    Chua, E
    Fawdry, R
    MacKinnon, I
    McHarg, D
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 611 - 611
  • [7] Superiority of iodine-123 compared with iodine-131 scanning for thyroid remnants in patients with differentiated thyroid cancer
    Mandel, SJ
    Shankar, LK
    Benard, F
    Yamamoto, A
    Alavi, A
    [J]. CLINICAL NUCLEAR MEDICINE, 2001, 26 (01) : 6 - 9
  • [9] Validation of a Monte Carlo modelled gamma camera for Lutetium-177 imaging
    Ramonaheng, K.
    van Staden, J. A.
    du Raan, H.
    [J]. APPLIED RADIATION AND ISOTOPES, 2020, 163