Evaluation and Reduction of Calibration Residual Uncertainty in Load-Pull Measurements at Millimeter-Wave Frequencies

被引:14
|
作者
Teppati, Valeria [1 ]
Bolognesi, C. R. [2 ]
机构
[1] Politecn Torino, Dept Elect, I-10129 Turin, Italy
[2] ETH, Millimeter Wave Elect Grp, CH-8092 Zurich, Switzerland
基金
瑞士国家科学基金会;
关键词
Calibration; load-pull; microwave measurements; millimeter-wave transistors; uncertainty; ERROR-CORRECTION;
D O I
10.1109/TIM.2011.2172119
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, a thorough evaluation of calibration residual uncertainty of on-wafer load-pull systems at millimeter-wave frequencies, with actual comparisons between real-time and non-real-time load-pull systems, is reported for the first time. Two figures of merit for uncertainty evaluation are taken into account, showing the differences between the two methodologies. In the case of non-real-time systems, based on a simulation tool developed for the purpose, typical values of uncertainties to be expected at millimeter-wave frequencies are shown. Finally, a methodology to reduce calibration residual uncertainty of non-real-time load-pull measurements, based on the optimization of a thru load-pull map, is for the first time introduced, and its effects on actual measurements of microwave HBTs at 40 GHz are shown.
引用
收藏
页码:817 / 822
页数:6
相关论文
共 50 条
  • [1] Calibration Error Reduction in Millimeter-Wave Load-Pull Systems Measuring Highly Reflective Loads
    Baddeley, Alexander
    Quaglia, Roberto
    Tasker, Paul J.
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2024,
  • [2] Microwave & millimeter-wave amplifier design via load-pull techniques
    Itoh, Y
    GAAS IC SYMPOSIUM - 22ND ANNUAL, TECHNICAL DIGEST 2000, 2000, : 43 - 46
  • [3] Transistor Vector Load-Pull Characterization for Millimeter-Wave Power Amplifier Design
    Vadala, Valeria
    Raffo, Antonio
    Bosi, Gianni
    Crupi, Giovanni
    Vannini, Giorgio
    2012 79TH ARFTG MICROWAVE MEASUREMENT CONFERENCE (ARFTG), 2012,
  • [4] SOURCE-PULL LOAD-PULL OSCILLATOR MEASUREMENTS AT MICROWAVE MM WAVE FREQUENCIES
    GHANNOUCHI, FM
    BOSISIO, RG
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1992, 41 (01) : 32 - 35
  • [5] Millimeter Wave and Sub-THz Power Sweep and Active Load-Pull Measurements
    Dudkiewicz, Steve
    Galatro, Luca
    Rowland, Cliff
    MICROWAVE JOURNAL, 2024, 67 (03) : 60 - 70
  • [6] AN AUTOMATED MILLIMETER-WAVE ACTIVE LOAD-PULL MEASUREMENT SYSTEM BASED ON 6-PORT TECHNIQUES
    GHANNOUCHI, FM
    BOSISIO, RG
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1992, 41 (06) : 957 - 962
  • [7] A calibration procedure for on-wafer differential load-pull measurements
    Spirito, M
    van der Heijden, MP
    de Kok, M
    de Vreede, LCN
    ARFTG: AUTOMATIC RF TECHNIQUES GROUP, CONFERENCE DIGEST, SPRING 2003, 2003, : 1 - 4
  • [8] HIGH-POWER MILLIMETER-WAVE MMIC AMPLIFIER DESIGN USING IMPROVED LOAD-PULL METHOD
    NAGATOMO, K
    KOIKE, S
    OKUBO, N
    SHIGAKI, M
    IEICE TRANSACTIONS ON ELECTRONICS, 1992, E75C (06) : 663 - 668
  • [9] Accuracy evaluation of on-wafer load-pull measurements
    Ferrero, A
    Teppati, V
    Carullo, A
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2001, 49 (01) : 39 - 43
  • [10] A Novel Load-Pull Setup with Envelope Calibration for Bias Modulated Measurements
    Bengtsson, Olof
    Vestling, Lars
    Olsson, Jorgen
    2009 EUROPEAN MICROWAVE CONFERENCE, VOLS 1-3, 2009, : 942 - +