Fabrication, calibration and evaluation of a phosphate ion-selective microelectrode

被引:13
|
作者
Wang, John J. [1 ]
Bishop, Paul L. [2 ]
机构
[1] Univ Cincinnati, Clarksburg, MD 20871 USA
[2] Univ Cincinnati, Dept Civil & Environm Engn, Cincinnati, OH 45221 USA
关键词
Phosphate; Ion-selective microelectrode; Enhanced biological phosphorus removal (EBPR); Floc; Micro-environment study; INORGANIC-PHOSPHATE; ELECTRODE; SENSOR;
D O I
10.1016/j.envpol.2010.08.007
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
To conduct the micro-environment study of flocs in an enhanced biological phosphorus removal (EBPR) process, a phosphate ion-selective microelectrode was developed. The cobalt-based microelectrodes have tip diameters of 5-20 mu m and respond to all the three forms of phosphate ions, namely, H(2)PO(4)(-), HPO(4)(2-), and PO(4)(3-). The calibration curve at pH 7.5 had a slope of 31.5 mV per decade change of concentration and a R(2) value of 0.99. Other characteristics of this microelectrode, such as response time, interferences from pH, ion strength, DO and other anions were also evaluated. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:3612 / 3617
页数:6
相关论文
共 50 条
  • [1] Thermodynamic Simulation of Phosphate Precipitation based on Ion-Selective Microelectrode Measurements
    Platt, Gustavo M.
    Bastos, Ivan N.
    de Andrade, Monica C.
    Taryba, Marina
    Lamaka, Sviatlana V.
    Simoes, Alda
    Soares, Gloria D.
    [J]. JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2013, 24 (06) : 1064 - 1071
  • [2] CALIBRATION OF ION-SELECTIVE MICROELECTRODES
    CROXTON, TL
    ARMSTRONG, WM
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 262 (05): : C1324 - C1334
  • [3] Ion-selective microelectrode sensitive to neurotransmitter dopamine
    Sun, YH
    Zheng, XX
    [J]. IEEE-EMBS ASIA PACIFIC CONFERENCE ON BIOMEDICAL ENGINEERING - PROCEEDINGS, PTS 1 & 2, 2000, : 411 - 412
  • [4] AN ION-SELECTIVE MICROELECTRODE FOR BILE-SALTS
    KUNZE, LJ
    LINDENBAUM, S
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1989, 7 (08) : 947 - 954
  • [5] MEASUREMENT OF NITRATE GRADIENTS WITH AN ION-SELECTIVE MICROELECTRODE
    DEBEER, D
    SWEERTS, JPR
    [J]. ANALYTICA CHIMICA ACTA, 1989, 219 (02) : 351 - 356
  • [6] PHOSPHATE ION-SELECTIVE ELECTRODE
    TAKAISHVILI, OG
    MOTSONELIDZE, EP
    KARACHENTSEVA, YM
    DAVITAYA, PI
    [J]. SOVIET ELECTROCHEMISTRY, 1976, 12 (02): : 288 - 290
  • [7] PHOSPHATE ION-SELECTIVE ELECTRODE
    MERRICK, DD
    SNYDER, WJ
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1975, (169): : 23 - 24
  • [8] EVALUATION OF CALCIUM ION-SELECTIVE ELECTRODES BASED ON DI(N-ALKYLPHENYL)PHOSPHATE SENSORS AND THEIR CALIBRATION WITH ION BUFFERS
    CRAGGS, A
    MOODY, GJ
    THOMAS, JDR
    [J]. ANALYST, 1979, 104 (1238) : 412 - 418
  • [9] CALIBRATION OF BROMIDE ION-SELECTIVE ELECTRODE
    GYENGE, R
    TOTH, K
    KOROS, E
    PUNGOR, E
    [J]. MAGYAR KEMIAI FOLYOIRAT, 1977, 83 (03): : 135 - 138
  • [10] CALIBRATION OF BROMIDE ION-SELECTIVE ELECTRODES
    GYENGE, R
    TOTH, K
    PUNGOR, E
    KOROS, E
    [J]. ANALYTICA CHIMICA ACTA, 1977, 94 (01) : 111 - 117