Delayed migration of a screw into the gastrointestinal tract after anterior cervical spine plating

被引:49
|
作者
Gazzeri, Roberto [1 ]
Tamorri, Marco [2 ]
Faiola, Andrea [2 ]
Gazzeri, Giovanni [2 ]
机构
[1] San Giovanni Addolorata Hosp, Dept Neurosurg, Rome, Italy
[2] San Filippo Neri Hosp, Rome, Italy
关键词
anterior fusion; instrumentation failure; cervical spine; cervical plating; complication; screw migration;
D O I
10.1097/BRS.0b013e31816b8831
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Case report and clinical discussion. Objective. To describe a rare case of spinal cervical screw migration into the gastrointestinal tract 11 years after surgery. Summary of Background Data. Anterior cervical spine fusion and stabilization with plating is a well-established procedure for cervical myelopathy, cervical spinal trauma, and spinal infectious disease. Esophageal injury has been related to screw or plate extrusion. Methods. We present a 45-year-old white man suffering from severe quadriparesis and neck pain. Cervical spine magnetic resonance images showed spondylodiscitis and spinal fracture of C4 and C5 vertebral bodies. He underwent anterior surgical decompression consisting of C4 and C5 corpectomies and fusion with fibular allograft fixated with a cervical plate. The patient's neurologic examination gradually improved during his follow-up. Results. The patient returned 11 years after cervical spine graft and plating, complaining of severe dysphagia and high fever. Radiographs of the cervical spine showed a screw back-out. Three days later, a new radiograph of the cervical spine revealed a progression of the screw's extrusion. In a new cervical spine radiograph, obtained immediately before surgical intervention, the displaced screw was no more visible. Abdominal radiograph showed the missed screw in the right lower abdominal quadrant. Barium meal swallowing test showed no leakage on the cervical area. At 6 months follow-up, the patient is in good condition and symptom free. Conclusion. In this case, the delayed esophageal perforation occurred 11 years after initial surgery: the rapid progression of the screw extrusion after initial pull-out (6 days) was documented by several radiographs and spontaneous closure of the fistula was achieved.
引用
收藏
页码:E268 / E271
页数:4
相关论文
共 50 条
  • [21] Esophageal perforation more than 10 years after anterior cervical spine plating
    Lucas, J.
    Smith, E.
    Eskander, M.
    McPhee, J.
    Lapinsky, A.
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2013, 115 (09) : 1842 - 1844
  • [22] Investigation of stabilizing potential of two anterior cervical spine plating systems incorporating unicortical locking screw designs
    Goel, VK
    Clausen, JD
    Sawin, PD
    Ryken, TC
    Traynelis, VC
    NEURO-ORTHOPEDICS, 1999, 25 (1-2): : 67 - 73
  • [23] Delayed pharyngo-oesophageal perforation after anterior cervical spine surgery
    Poissonnet, V.
    Lubrano, V.
    Sadeler, A.
    Chabrillac, E.
    NEUROCHIRURGIE, 2020, 66 (05) : 414 - 416
  • [24] Intrathoracic migration of a screw in the cervical spine: a case report
    Kalaycioglu, Seray
    Findik, Gokturk
    Aydogdu, Koray
    Dagli, Gulfidan
    Kaya, Sadi
    TURK GOGUS KALP DAMAR CERRAHISI DERGISI-TURKISH JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 18 (04): : 338 - 340
  • [25] Delayed esophageal perforation 17 years after anterior cervical fusion surgery with plating
    Ikuma, Hisanori
    Ishibashi, Katsuhiko
    Yamashita, Kazuta
    Joko, Ryoji
    Makihara, Seiichiro
    Kobayashi, Masahiko
    Yokoyama, Yoshiki
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2017, 22 (03) : 566 - 570
  • [26] Range of motion in the cervical spine after odontoid fracture treated with anterior screw fixation
    Wolan-Nieroda, Andzelina
    Maciejczak, Andrzej
    Guzik, Agnieszka
    Przysada, Grzegorz
    Szeliga, Ewa
    Druzbicki, Mariusz
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (1)
  • [27] Range of motion in the cervical spine after odontoid fracture treated with anterior screw fixation
    Andżelina Wolan-Nieroda
    Andrzej Maciejczak
    Agnieszka Guzik
    Grzegorz Przysada
    Ewa Szeliga
    Mariusz Drużbicki
    Journal of Orthopaedic Surgery and Research, 14
  • [28] Fibular allograft and anterior plating for dislocations/fractures of the cervical spine
    Ranmarain, A.
    Govender, S.
    INDIAN JOURNAL OF ORTHOPAEDICS, 2008, 42 (01) : 83 - 86
  • [29] Cement Augmentation of the Cervical Spine - a Technique Enhancing Stability of Anterior Cervical Plating
    Bayley, E.
    Klezl, Z., Jr.
    Bommiredy, R.
    Klezl, Z., Sr.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2017, 84 (06) : 448 - 452
  • [30] Investigation of stabilizing potential of two anterior cervical spine plating systems incorporating unicortical locking screw designs - Comment
    Haid, R
    NEURO-ORTHOPEDICS, 1999, 25 (1-2): : 73 - 73