Study Design. Case report and clinical discussion. Objective. To describe a rare case of spinal cervical screw migration into the gastrointestinal tract 11 years after surgery. Summary of Background Data. Anterior cervical spine fusion and stabilization with plating is a well-established procedure for cervical myelopathy, cervical spinal trauma, and spinal infectious disease. Esophageal injury has been related to screw or plate extrusion. Methods. We present a 45-year-old white man suffering from severe quadriparesis and neck pain. Cervical spine magnetic resonance images showed spondylodiscitis and spinal fracture of C4 and C5 vertebral bodies. He underwent anterior surgical decompression consisting of C4 and C5 corpectomies and fusion with fibular allograft fixated with a cervical plate. The patient's neurologic examination gradually improved during his follow-up. Results. The patient returned 11 years after cervical spine graft and plating, complaining of severe dysphagia and high fever. Radiographs of the cervical spine showed a screw back-out. Three days later, a new radiograph of the cervical spine revealed a progression of the screw's extrusion. In a new cervical spine radiograph, obtained immediately before surgical intervention, the displaced screw was no more visible. Abdominal radiograph showed the missed screw in the right lower abdominal quadrant. Barium meal swallowing test showed no leakage on the cervical area. At 6 months follow-up, the patient is in good condition and symptom free. Conclusion. In this case, the delayed esophageal perforation occurred 11 years after initial surgery: the rapid progression of the screw extrusion after initial pull-out (6 days) was documented by several radiographs and spontaneous closure of the fistula was achieved.
引用
收藏
页码:E268 / E271
页数:4
相关论文
共 50 条
  • [1] Extrusion of a screw into the gastrointestinal tract after anterior cervical spine plating
    Fountas, Kostas N.
    Kapsalaki, Eftychia Z.
    Machinis, Theofilos
    Robinson, Joe Sam
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (03): : 199 - 203
  • [2] Oral extrusion of a screw after anterior cervical spine plating
    Geyer, TE
    Foy, MA
    SPINE, 2001, 26 (16) : 1814 - 1816
  • [3] Screw Migration into Colon after Anterior Cervical Plating - An Unusual Complication: A Case Report
    Sath, S.
    MALAYSIAN ORTHOPAEDIC JOURNAL, 2020, 14 (03) : 184 - 187
  • [4] THE ANTERIOR PLATING OF THE CERVICAL-SPINE WITH THE TITANIUM HOLLOW-SCREW PLATING SYSTEM
    MORSCHER, E
    SUTTER, F
    JENNY, H
    OLERUD, S
    CHIRURG, 1986, 57 (11): : 702 - 707
  • [5] Biomechanics of variable angle screw fixation in anterior plating of the cervical spine
    Janssen, M.
    Sims, M.
    Lam, C.
    Journal of Biomechanics, 2001, 34 (SUPPL. 1)
  • [6] Delayed Anterior Cervical Screw Migration Causing Swallowing Difficulty
    Maurer, Timothy
    Maurer, Robert
    Reiter, George T.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (07)
  • [7] Delayed Screw Migration Following Anterior Cervical Discectomy and Fusion
    Carr, Matthew T.
    Steinberger, Jeremy
    Houten, John K.
    WORLD NEUROSURGERY, 2025, 194
  • [8] Successful surgical management of a delayed pharyngo-esophageal perforation after anterior cervical spine plating
    Solerio, Dino
    Ruffini, Enrico
    Gargiulo, Giosue
    Camandona, Michele
    Raggio, Eleonora
    Solini, Antonio
    Poli, Marcello Dei
    EUROPEAN SPINE JOURNAL, 2008, 17 (Suppl 2) : S280 - S284
  • [9] Successful surgical management of a delayed pharyngo-esophageal perforation after anterior cervical spine plating
    Dino Solerio
    Enrico Ruffini
    Giosué Gargiulo
    Michele Camandona
    Eleonora Raggio
    Antonio Solini
    Marcello Dei Poli
    European Spine Journal, 2008, 17 : 280 - 284
  • [10] A Rare Case of Hypopharyngeal Screw Migration after Spine Stabilization with Plating
    Salis, G.
    Pittore, B.
    Balata, G.
    Bozzo, C.
    CASE REPORTS IN OTOLARYNGOLOGY, 2013, 2013