Analytical Low-Dose CBCT Reconstruction Using Non-local Total Variation Regularization for Image Guided Radiation Therapy

被引:4
|
作者
Sohn, James J. [1 ]
Kim, Changsoo [2 ]
Kim, Dong Hyun [2 ]
Lee, Seu-Ran [3 ]
Zhou, Jun [1 ]
Yang, Xiaofeng [1 ]
Liu, Tian [1 ]
机构
[1] Emory Univ, Dept Radiat Oncol, Atlanta, GA 30322 USA
[2] Catholic Univ Pusan, Dept Radiol Sci, Busan, South Korea
[3] Catholic Univ Pusan, Dept Biomed Engn, Seoul, South Korea
来源
FRONTIERS IN ONCOLOGY | 2020年 / 10卷
关键词
low-dose CBCT; non-local total variation; compressed sensing; image reconstruction; image-guided radiation therapy (IGRT); BEAM CT RECONSTRUCTION; MGIR;
D O I
10.3389/fonc.2020.00242
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Conventional iterative low-dose CBCT reconstruction techniques are slow and tend to over-smooth edges through uniform weighting of the image penalty gradient. In this study, we present a non-iterative analytical low-dose CBCT reconstruction technique by restoring the noisy low-dose CBCT projection with the non-local total variation (NLTV) method. Methods: We modeled the low-dose CBCT reconstruction as recovering high quality, high-dose CBCT x-ray projections (100 kVp, 1.6 mAs) from low-dose, noisy CBCT x-ray projections (100 kVp, 0.1 mAs). The restoration of CBCT projections was performed using the NLTV regularization method. In NLTV, the x-ray image is optimized by minimizing an energy function that penalizes gray-level difference between pair of pixels between noisy x-ray projection and denoising x-ray projection. After the noisy projection is restored by NLTV regularization, the standard FDK method was applied to generate the final reconstruction output. Results: Significant noise reduction was achieved comparing to original, noisy inputs while maintaining the image quality comparable to the high-dose CBCT projections. The experimental validations show the proposed NLTV algorithm can robustly restore the noise level of x-ray projection images while significantly improving the overall image quality. The improvement in normalized mean square error (NMSE) and peak signal-to-noise ratio (PSNR) measured from the non-local total variation-gradient projection (NLTV-GPSR) algorithm is noticeable compared to that of uncorrected low-dose CBCT images. Moreover, the difference of CNRs from the gains from the proposed algorithm is noticeable and comparable to high-dose CBCT. Conclusion: The proposed method successfully restores noise degraded, low-dose CBCT projections to high-dose projection quality. Such an outcome is a considerable improvement to the reconstruction result compared to the FDK-based method. In addition, a significant reduction in reconstruction time makes the proposed algorithm more attractive. This demonstrates the potential use of the proposed algorithm for clinical practice in radiotherapy.
引用
下载
收藏
页数:7
相关论文
共 50 条
  • [21] Non-local total variation regularization approach for image restoration under a Poisson degradation
    Kayyar, Shivarama Holla
    Jidesh, P.
    JOURNAL OF MODERN OPTICS, 2018, 65 (19) : 2231 - 2242
  • [22] Poissonian image deblurring method by non-local total variation and framelet regularization constraint
    Shi, Yu
    Song, Jie
    Hua, Xia
    COMPUTERS & ELECTRICAL ENGINEERING, 2017, 62 : 319 - 329
  • [23] IMAGE RESTORATION VIA MULTI-SCALE NON-LOCAL TOTAL VARIATION REGULARIZATION
    Mu, Jing
    Xiong, Ruiqin
    Fan, Xiaopeng
    Ma, Siwei
    2017 IEEE INTERNATIONAL CONFERENCE ON MULTIMEDIA AND EXPO (ICME), 2017, : 205 - 210
  • [24] Single Image Dehazing With Depth-Aware Non-Local Total Variation Regularization
    Liu, Qi
    Gao, Xinbo
    He, Lihuo
    Lu, Wen
    IEEE TRANSACTIONS ON IMAGE PROCESSING, 2018, 27 (10) : 5178 - 5191
  • [25] Low Dose PET Reconstruction with Total Variation Regularization
    Wang, Chenye
    Hu, Zhenghui
    Shi, Pengcheng
    Liu, Huafeng
    2014 36TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2014, : 1917 - 1920
  • [26] Noise Reduction in PET Sinograms Using Non-Local Total Variation Regularization
    Mejia, Jose
    Mederos, Boris
    Cabrera, Sergio D.
    Ochoa Dominguez, Humberto
    Vergara Villegas, Osslan O.
    2014 IEEE SOUTHWEST SYMPOSIUM ON IMAGE ANALYSIS AND INTERPRETATION (SSIAI 2014), 2014, : 17 - 20
  • [27] Low-dose CT reconstruction via edge-preserving total variation regularization
    Tian, Zhen
    Jia, Xun
    Yuan, Kehong
    Pan, Tinsu
    Jiang, Steve B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (18): : 5949 - 5967
  • [28] Noninvasive electrocardiographic imaging with low-rank and non-local total variation regularization
    Mu, Lide
    Liu, Huafeng
    PATTERN RECOGNITION LETTERS, 2020, 138 : 106 - 114
  • [29] VECTORIAL NON-LOCAL TOTAL VARIATION REGULARIZATION FOR CALIBRATION-FREE PARALLEL MRI RECONSTRUCTION
    Lefkimmiatis, Stamatios
    Saucedo, Andres
    Osher, Stanley
    Sung, Kyunghyun
    2015 IEEE 12TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2015, : 1256 - 1259
  • [30] Improvement of Image Registration Using Total-Variation Based Noise Reduction Algorithms for Low-Dose CBCT
    Mukherjee, S.
    Farr, J.
    Merchant, T.
    Yao, W.
    MEDICAL PHYSICS, 2016, 43 (06) : 3388 - 3388