Swept source optical coherence tomography with external clocking using voltage controlled oscillator

被引:1
|
作者
Lee, Eung Je [1 ]
Kim, Yong Pyung [1 ]
机构
[1] Kyung Hee Univ, Coll Elect & Informat, Inst Laser Engn, Yongin 446701, Gyeonggi Do, South Korea
关键词
wavelength-swept source; frequency domain optical coherence tomography; Fourier domain mode-locked laser; recalibration of beat frequency broadening; DOMAIN MODE-LOCKING; SEMICONDUCTOR-LASER; LOCKED LASER; NM; FDML;
D O I
10.1117/1.3577705
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In this study, a beat signal recalibration method was developed for optical coherence tomography (OCT) with a high-speed wavelength-swept source. By adopting a voltage-controlled oscillator (VCO) modulated by a sinusoidal waveform as a trigger for the OCT system, the broadening of the beat frequency due to laser tuning rate variations was recalibrated. For this study, OCT based on a Fourier domain mode-locked (FDML) laser at a sweep rate of 60.9 kHz was demonstrated. OCT images of 1500x409 pixels were obtained with the sensitivity of 100 dB. Temporal frequency variations in the FDML laser and OCT images obtained with the proposed technique were also described. When compared to a conventional recalibration method using optical components, swept source optical coherence tomography operated with a VCO clock exhibited superior performance and high stability. From the experimental results, it was demonstrated that the proposed method is sufficient to recalibrate the time-frequency variations in interferometry with a high-speed wavelength-swept source. (C) 2011 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.3577705]
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Swept source optical coherence tomography artifacts in glaucoma patients
    Lee, S. Y.
    Seo, S. J.
    Lee, Y. H.
    Bae, H. Y.
    Kim, C. Y.
    [J]. ACTA OPHTHALMOLOGICA, 2014, 92
  • [32] High axial resolution swept source for optical coherence tomography
    Cordes, A. H.
    Xavier, G. B.
    de Faria, G. Vilela
    von der Weid, J. P.
    [J]. ELECTRONICS LETTERS, 2010, 46 (01) : 27 - 28
  • [33] Reproducibility of the measurements taken with swept source optical coherence tomography
    Orduna Hospital, E.
    Vilades Palomar, E.
    Cipres, M.
    Obis, J.
    Rodrigo, M. J.
    Satue, M.
    Garcia-Martin, E.
    [J]. ACTA OPHTHALMOLOGICA, 2017, 95
  • [34] SPECTRAL SIGNAL PROCESSING IN SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY
    Chang, Shoude
    Mao, Youxin
    Flueraru, Costel
    [J]. PHOTONICS NORTH 2010, 2010, 7750
  • [35] Widefield Swept-Source Optical Coherence Tomography Angiography
    Shahidzadeh, Anoush
    Kim, Alice Yon
    Aziz, Hassan A.
    Durbin, Mary K.
    Puliafito, Carmen A.
    Kashani, Amir H.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [36] Swept source optical coherence tomography angiography in choroidal melanoma
    Pellegrini, Marco
    Corvi, Federico
    Ravera, Vittoria
    Staurenghi, Giovanni
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [37] Efficient sweep buffering in swept source optical coherence tomography using a fast optical switch
    [J]. Izatt, J. A. (jizatt@duke.edu), 1600, OSA - The Optical Society (03):
  • [38] Efficient sweep buffering in swept source optical coherence tomography using a fast optical switch
    Dhalla, Al-Hafeez
    Shia, Kevin
    Izatt, Joseph A.
    [J]. BIOMEDICAL OPTICS EXPRESS, 2012, 3 (12):
  • [39] External line-cavity wavelength-swept source at 850 nm for optical coherence tomography
    Lee, Sang-Won
    Kim, Chang-Seok
    Kim, Beop-Min
    [J]. IEEE PHOTONICS TECHNOLOGY LETTERS, 2007, 19 (2-4) : 176 - 178
  • [40] Choroidal Analysis in Healthy Eyes Using Swept-Source Optical Coherence Tomography Compared to Spectral Domain Optical Coherence Tomography
    Adhi, Mehreen
    Liu, Jonathan J.
    Qavi, Ahmed H.
    Grulkowski, Ireneusz
    Lu, Chen D.
    Mohler, Kathrin J.
    Ferrara, Daniela
    Kraus, Martin F.
    Baumal, Caroline R.
    Witkin, Andre J.
    Waheed, Nadia K.
    Hornegger, Joachim
    Fujimoto, James G.
    Duker, Jay S.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2014, 157 (06) : 1272 - 1281