A simple ensemble strategy of uninformative variable elimination and partial least-squares for near-infrared spectroscopic calibration of pharmaceutical products

被引:22
|
作者
Tan, Chao [1 ,2 ]
Wu, Tong [1 ]
Xu, Zehong [1 ]
Li, Weiyi [1 ]
Zhang, Kaishi [1 ]
机构
[1] Yibin Univ, Dept Chem & Chem Engn, Yibin 644007, Sichuan, Peoples R China
[2] Yibin Univ, Computat Phys Key Lab Sichuan Prov, Yibin 644007, Sichuan, Peoples R China
关键词
Near-infrared; Spectroscopy; Ensemble; Calibration; Uninformative variable elimination; SUBSPACE REGRESSION ENSEMBLE; REFLECTANCE SPECTROSCOPY; WAVELET TRANSFORM; FEASIBILITY; TOBACCO;
D O I
10.1016/j.vibspec.2011.09.011
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Based on uninformative variable elimination (UVE) and partial least squares (PLS), a simple ensemble strategy, named EUVEPLS, is proposed for multivariate calibration of near-infrared spectroscopy. In such a strategy, different calibration sets are first formed by randomly selecting a fixed number of objects from the available calibration data, with replacement, in this manner, different PLS models can be obtained, as in the jackknife procedure. Then, the UVE is used to shrink the original variable space into a specific Subspace, in which a PLS model with the optimal number of latent variables is determined by cross-validation. This process is repeated until a fixed number of candidate models, and therefore the same number of subspaces, are obtained. Finally, those models with better performance are used to produce an ensemble model by simple/weighted average. Two NIR spectral datasets concerning the pharmaceutical products are used to verify the accuracy and robustness of the proposed EUVEPLS. The results confirmed the superiority of EUVEPLS to two single model reference methods (UVEPLS and full-spectrum PLS). (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:44 / 49
页数:6
相关论文
共 50 条
  • [1] Variable selection for partial least-squares calibration of near-infrared data from orthogonally designed experiments
    Setarehdan, SK
    Soraghan, JJ
    Littlejohn, D
    Sadler, DA
    [J]. APPLIED SPECTROSCOPY, 2002, 56 (03) : 337 - 345
  • [2] Elimination of the Uninformative Calibration Sample Subset in the Modified UVE(Uninformative Variable Elimination)–PLS (Partial Least Squares) Method
    Jun Koshoubu
    Tetsuo Iwata
    Shigeo Minami
    [J]. Analytical Sciences, 2001, 17 : 319 - 322
  • [3] Elimination of the uninformative calibration sample subset in the modified UVE (Uninformative Variable Elimination)-PLS (Partial Least Squares) method
    Koshoubu, J
    Iwata, T
    Minami, S
    [J]. ANALYTICAL SCIENCES, 2001, 17 (02) : 319 - 322
  • [4] Assessment of partial least-squares calibration and wavelength selection for complex near-infrared spectra
    McShane, MJ
    Cote, GL
    Spiegelman, CH
    [J]. APPLIED SPECTROSCOPY, 1998, 52 (06) : 878 - 884
  • [5] Variable space boosting partial least squares for multivariate calibration of near-infrared spectroscopy
    Bian, Xihui
    Li, Shujuan
    Shao, Xueguang
    Liu, Peng
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2016, 158 : 174 - 179
  • [6] A variable selection method based on uninformative variable elimination for multivariate calibration of near-infrared spectra
    Cai, Wensheng
    Li, Yankun
    Shao, Xueguang
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2008, 90 (02) : 188 - 194
  • [7] An improved ensemble partial least squares for analysis of near-infrared spectra
    Hu, Yong
    Peng, Silong
    Peng, Jiangtao
    Wei, Jiping
    [J]. TALANTA, 2012, 94 : 301 - 307
  • [8] Nondestructive Discrimination of Pharmaceutical Preparations Using Near-Infrared Spectroscopy and Partial Least-Squares Discriminant Analysis
    Chen, Hui
    Lin, Zan
    Tan, Chao
    [J]. ANALYTICAL LETTERS, 2018, 51 (04) : 564 - 574
  • [9] Retention prediction of peptides based on uninformative variable elimination by partial least squares
    Put, R.
    Daszykowski, M.
    Baczek, T.
    Vander Heyden, Y.
    [J]. JOURNAL OF PROTEOME RESEARCH, 2006, 5 (07) : 1618 - 1625
  • [10] Application of partial least-squares regression to near-infrared reflectance spectroscopic determination of shive content in flax
    Sohn, M
    Barton, FE
    Morrison, WH
    Archibald, DD